Blurring the boundaries between frame-based and frameless stereotaxy: feasibility study for brain biopsies performed with the use of a head-mounted robot

被引:35
作者
Grimm, Florian [1 ]
Naros, Georgios [1 ]
Gutenberg, Angelika [2 ]
Keric, Naureen [2 ]
Giese, Alf
Gharabaghi, Alireza [1 ]
机构
[1] Univ Tubingen, Dept Neurosurg, Div Funct & Restorat Neurosurg, D-72076 Tubingen, Germany
[2] Johannes Gutenberg Univ Mainz, Dept Neurosurg, D-55122 Mainz, Germany
关键词
stereotaxy; frameless brain biopsy; frame-based brain biopsy; robotics; neuronavigation; tumor; oncology; diagnostic and operative techniques; ACCURACY; ADVANTAGES; SURGERY; DEVICES; NEEDLE;
D O I
10.3171/2014.12.JNS141781
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECT Frame-based stereotactic interventions are considered the gold standard for brain biopsies, but they have limitations with regard to flexibility and patient comfort because of the bulky head ring attached to the patient. Frameless image guidance systems that use scalp fiducial markers offer more flexibility and patient comfort but provide less stability and accuracy during drilling and biopsy needle positioning. Head-mounted robot guided biopsies could provide the advantages of these 2 techniques without the downsides. The goal of this study was to evaluate the feasibility and safety of a robotic guidance device, affixed to the patient's skull through a small mounting platform, for use in brain biopsy procedures. METHODS This was a retrospective study of 37 consecutive patients who presented with supratentorial lesions and underwent brain biopsy procedures in which a surgical guidance robot was used to determine clinical outcomes and technical procedural operability. RESULTS The portable head-mounted device was well tolerated by the patients and enabled stable drilling and needle positioning during surgery. Flexible adjustments of predefined paths and selection of new trajectories were successfully performed intraoperatively without the need for manual settings and fixations. The patients experienced no permanent deficits or infections after surgery. CONCLUSIONS The head-mounted robot guided approach presented here combines the stability of a bone-mounted set-up with the flexibility and tolerability of frameless systems. By reducing human interference (i.e., manual parameter settings, calibrations, and adjustments), this technology might be particularly useful in neurosurgical interventions that necessitate multiple trajectories.
引用
收藏
页码:737 / 742
页数:6
相关论文
共 19 条
  • [1] Comparing the risks of frameless stereotactic biopsy in eloquent and noneloquent regions of the brain: a retrospective review of 284 cases Clinical article
    Air, Ellen L.
    Leach, James L.
    Warnick, Ronald E.
    McPherson, Christopher M.
    [J]. JOURNAL OF NEUROSURGERY, 2009, 111 (04) : 820 - 824
  • [2] Image-guided frameless stereotactic needle biopsy in awake patients without the use of rigid head fixation
    Amin, Devin V.
    Lozanne, Karl
    Parry, Phillip V.
    Engh, Johnathan A.
    Seelman, Kathleen
    Mintz, Arlan
    [J]. JOURNAL OF NEUROSURGERY, 2011, 114 (05) : 1414 - 1420
  • [3] Devices for Targeting the Needle
    Barbre, Carol J.
    [J]. NEUROSURGERY CLINICS OF NORTH AMERICA, 2009, 20 (02) : 187 - +
  • [4] Frameless robotically targeted stereotactic brain biopsy: feasibility, diagnostic yield, and safety
    Bekelis, Kimon
    Badwan, Tarek A.
    Desai, Atman
    Roberts, David W.
    [J]. JOURNAL OF NEUROSURGERY, 2012, 116 (05) : 1002 - 1006
  • [5] Accuracy of true frameless stereotaxy: in vivo measurement and laboratory phantom studies - Technical note
    Dorward, NL
    Alberti, O
    Palmer, JD
    Kitchen, ND
    Thomas, DGT
    [J]. JOURNAL OF NEUROSURGERY, 1999, 90 (01) : 160 - 168
  • [6] The advantages of frameless stereotactic biopsy over frame-based biopsy
    Dorward, NL
    Paleologos, TS
    Alberti, O
    Thomas, DGT
    [J]. BRITISH JOURNAL OF NEUROSURGERY, 2002, 16 (02) : 110 - 118
  • [7] Glauser D, 1995, J Image Guid Surg, V1, P266, DOI 10.3109/10929089509106332
  • [8] Kitchen Neil D., 1993, Stereotactic and Functional Neurosurgery, V61, P195, DOI 10.1159/000100637
  • [9] Robotically-Adjustable Microstereotactic Frames for Image-Guided Neurosurgery
    Kratchman, Louis B.
    Fitzpatrick, J. Michael
    [J]. MEDICAL IMAGING 2013: IMAGE-GUIDED PROCEDURES, ROBOTIC INTERVENTIONS, AND MODELING, 2013, 8671
  • [10] A ROBOT WITH IMPROVED ABSOLUTE POSITIONING ACCURACY FOR CT GUIDED STEREOTACTIC BRAIN SURGERY
    KWOH, YS
    HOU, J
    JONCKHEERE, EA
    HAYATI, S
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (02) : 153 - 160