Optimal Level of Continuous Positive Airway Pressure: Auto-adjusting Titration Versus Titration With a Predictive Equation

被引:3
|
作者
Choi, Ji Ho [1 ]
Jun, Young Joon [1 ]
Oh, Jeong In [1 ]
Jung, Jong Yoon [1 ]
Hwang, Gyu Ho [1 ]
Kwon, Soon Young [1 ]
Lee, Heung Man [1 ]
Kim, Tae Hoon [1 ]
Lee, Sang Hag [1 ]
Lee, Seung Hoon [1 ]
机构
[1] Korea Univ, Coll Med, Dept Otorhinolaryngol Head & Neck Surg, Seoul 136705, South Korea
关键词
adult; body mass index; continuous positive airway pressure; obstructive sleep apnea syndrome; polysomnography; OBSTRUCTIVE SLEEP-APNEA; PRACTICE PARAMETERS; THERAPY;
D O I
10.1177/000348941312200509
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
Objectives: The aims of the present study were twofold. We sought to compare two methods of titrating the level of continuous positive airway pressure (CPAP) - auto-adjusting titration and titration using a predictive equation with full-night manual titration used as the benchmark. We also investigated the reliability of the two methods in patients with obstructive sleep apnea syndrome (OSAS). Methods: Twenty consecutive adult patients with OSAS who had successful, full-night manual and auto-adjusting CPAP titration participated in this study. The titration pressure level was calculated with a previously developed predictive equation based on body mass index and apnea-hypopnea index. Results: The mean titration pressure levels obtained with the manual, auto-adjusting, and predictive equation methods were 9.0 +/- 3.6, 9.4 +/- 3.0, and 8.1 +/- 1.6 cm H2O, respectively. There was a significant difference in the concordance within the range of +/- 2 cm H2O (p = 0.019) between both the auto-adjusting titration and the titration using the predictive equation compared to the full-night manual titration. However, there was no significant difference in the concordance within the range of 1 cm H2O (p > 0.999). Conclusions: When compared to full-night manual titration as the standard method, auto-adjusting titration appears to be more reliable than using a predictive equation for determining the optimal CPAP level in patients with OSAS.
引用
收藏
页码:339 / 343
页数:5
相关论文
共 50 条
  • [1] Nasal versus oronasal mask in patients under auto-adjusting continuous positive airway pressure titration: a real-life study
    Duarte, Ricardo L. M.
    Mendes, Bruno A.
    Oliveira-e-Sa, Tiago S.
    Magalhaes-da-Silveira, Flavio J.
    Gozal, David
    EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2020, 277 (12) : 3507 - 3512
  • [2] Auto-adjusting and Advanced Positive Airway Pressure Therapeutic Modalities
    Ryden, Armand
    Bando, Joanne M.
    Aysola, Ravi S.
    SEMINARS IN RESPIRATORY AND CRITICAL CARE MEDICINE, 2014, 35 (05) : 593 - 603
  • [3] Accuracy of positive airway pressure titration through telemonitoring of auto-adjusting positive airway pressure device connected to a pulse oximetry in patients with obstructive sleep apnea
    Foresi, Antonio
    Vitale, Tommaso
    Prestigiacomo, Rosaria
    Ranieri, Piera
    Bosi, Marcello
    CLINICAL RESPIRATORY JOURNAL, 2023, 17 (08) : 740 - 747
  • [4] Comparison of manual versus automatic continuous positive airway pressure titration and the development of a predictive equation for therapeutic continuous positive airway pressure in Chinese patients with obstructive sleep apnoea
    Luo, Jiaying
    Xiao, Sichang
    Qiu, Zhihui
    Song, Ning
    Luo, Yuanming
    RESPIROLOGY, 2013, 18 (03) : 528 - 533
  • [5] Flexible Positive Airway Pressure Improves Treatment Adherence Compared with Auto-adjusting PAP
    Chihara, Yuichi
    Tsuboi, Tomomasa
    Hitomi, Takefumi
    Azuma, Masanori
    Murase, Kimihiko
    Toyama, Yoshiro
    Harada, Yuka
    Aihara, Kensaku
    Tanizawa, Kiminobu
    Handa, Tomohiro
    Yoshimura, Chikara
    Oga, Toru
    Yamamoto, Kazuhiko
    Mishima, Michiaki
    Chin, Kazuo
    SLEEP, 2013, 36 (02) : 229 - 236
  • [6] Continuous positive airway pressure titration for obstructive sleep apnoea: automatic versus manual titration
    McArdle, Nigel
    Singh, Bhajan
    Murphy, Michelle
    Gain, Kevin R.
    Maguire, Christine
    Mutch, Sarah
    Hillman, David R.
    THORAX, 2010, 65 (07) : 606 - 611
  • [7] Auto-adjusting Positive Airway Pressure Treatment for Sleep Apnea Diagnosed by Home Sleep Testing
    Berry, Richard B.
    Sriram, Peruvemba
    JOURNAL OF CLINICAL SLEEP MEDICINE, 2014, 10 (12): : 1269 - 1275
  • [8] Titration with automatic continuous positive airway pressure in obstructive steep apnea
    Dias, C.
    Sousa, L.
    Batata, L.
    Reis, R.
    Teixeira, F.
    Moita, J.
    Moutinho dos Santos, J.
    REVISTA PORTUGUESA DE PNEUMOLOGIA, 2017, 23 (04) : 203 - 207
  • [9] NightBalance Sleep Position Treatment Device Versus Auto-Adjusting Positive Airway Pressure for Treatment of Positional Obstructive Sleep Apnea
    Berry, Richard B.
    Uhles, Matthew L.
    Abaluck, Brian K.
    Winslow, David H.
    Schweitzer, Paula K.
    Gaskins, Raymond A., Jr.
    Doekel, Robert C., Jr.
    Emsellem, Helene A.
    JOURNAL OF CLINICAL SLEEP MEDICINE, 2019, 15 (07): : 947 - 956
  • [10] The effects of heated humidifier in continuous positive airway pressure titration
    Yu, Chung-Chieh
    Luo, Cheng-Ming
    Liu, Yu-Chih
    Wu, Huang-Pin
    SLEEP AND BREATHING, 2013, 17 (01) : 133 - 138