Measurement and analysis of channel attenuation characteristics for an implantable galvanic coupling human-body communication

被引:2
|
作者
Zhang, Shuang [1 ,2 ,3 ,4 ]
Pun, Sio Hang [1 ,2 ]
Mak, Peng Un [1 ,2 ]
Qin, Yu-Ping [3 ]
Liu, Yi-He [4 ]
Vai, Mang I. [1 ,2 ]
机构
[1] Univ Macau, State Key Lab Analog & Mixed Signal VLSI, Macau, Peoples R China
[2] Univ Macau, Dept Elect & Comp Engn, Fac Sci & Technol, Macau, Peoples R China
[3] Chengdu Univ Technol, Engn & Tech Coll, Leshan, Sichuan, Peoples R China
[4] Neijiang Normal Univ, Coll Comp Sci, Neijiang, Sichuan, Peoples R China
关键词
Implantable; galvanic coupling human-body communication; attenuation; measurement;
D O I
10.3233/THC-161229
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
In this study, an experiment was designed to verify the low power consumption of galvanic coupling human-body communication. A silver electrode (silver content: 99%) is placed in a pig leg and a sine wave signal with the power of 0 dBm is input. Compared with radio frequency communication and antenna transmission communication, attenuation is reduced by approximately 10 to 15 dB, so channel characteristics are highly improved.
引用
收藏
页码:821 / 826
页数:6
相关论文
共 50 条
  • [21] MEASUREMENT OF ELECTROSTATIC VOLTAGE, CAPACITANCE AND ENERGY-STORAGE CHARACTERISTICS OF HUMAN-BODY
    MARBLE, AE
    MACDONALD, AC
    MCVICAR, D
    ROBERTS, A
    PHYSICS IN MEDICINE AND BIOLOGY, 1977, 22 (02): : 365 - 367
  • [22] Comparable Investigation of Characteristics for Implant Intra-Body Communication Based on Galvanic and Capacitive Coupling
    Li, Maoyuan
    Song, Yong
    Hou, Yongtao
    Li, Ning
    Jiang, Yurong
    Sulaman, Muhammad
    Hao, Qun
    IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2019, 13 (06) : 1747 - 1758
  • [23] Analysis of Wearable Ungrounded Antennas for UHF RFIDs with respect to the coupling with Human-Body
    Casula, G. A.
    Montisci, G.
    Michel, A.
    Nepa, P.
    2016 IEEE INTERNATIONAL CONFERENCE ON RFID TECHNOLOGY AND APPLICATIONS (RFID-TA), 2016, : 6 - 9
  • [24] CHARACTERISTICS OF THE TEMPERATURE REGULATION SYSTEM IN THE HUMAN-BODY
    KAWASHIMA, Y
    JOURNAL OF THERMAL BIOLOGY, 1993, 18 (5-6) : 307 - 323
  • [25] CHARACTERISTICS OF ACETABULAR CUPS WORN IN HUMAN-BODY
    DOWLING, JM
    ATKINSON, JR
    DOWSON, D
    CHARNLEY, J
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1978, 60 (03): : 375 - 382
  • [26] IMMUNOCHEMICAL AND PHYSICOCHEMICAL CHARACTERISTICS OF HUMAN-BODY LACTOFERRIN
    NIKOLAEV, AA
    ANSHAKOVA, NI
    VOPROSY MEDITSINSKOI KHIMII, 1985, 31 (03): : 128 - 132
  • [27] A Finite-Element Simulation of Galvanic Coupling Intra-Body Communication Based on the Whole Human Body
    Song, Yong
    Zhang, Kai
    Hao, Qun
    Hu, Lanxin
    Wang, Jingwen
    Shang, Fuzhou
    SENSORS, 2012, 12 (10): : 13567 - 13582
  • [28] Signal Path Loss Simulation of Human Arm for Galvanic Coupling Intra-body Communication
    Ito, Kenichi
    Hotta, Yu
    JOURNAL OF ADVANCED SIMULATION IN SCIENCE AND ENGINEERING, 2016, 3 (01): : 29 - 46
  • [29] Impact of Electrical Properties Variations of Human-Body Phantom on Implantable Antenna Performance During In-Vitro Measurement
    Ghosh, Soham
    Chatterjee, Sanjana
    Ghosh, Satyaki
    Gupta, Bhaskar
    IETE JOURNAL OF RESEARCH, 2025,
  • [30] Signal Transmission by Galvanic Coupling Through the Human Body
    Wegmueller, Marc Simon
    Oberle, Michael
    Felber, Norbert
    Kuster, Niels
    Fichtner, Wolfgang
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2010, 59 (04) : 963 - 969