Measurement and analysis of channel attenuation characteristics for an implantable galvanic coupling human-body communication

被引:2
|
作者
Zhang, Shuang [1 ,2 ,3 ,4 ]
Pun, Sio Hang [1 ,2 ]
Mak, Peng Un [1 ,2 ]
Qin, Yu-Ping [3 ]
Liu, Yi-He [4 ]
Vai, Mang I. [1 ,2 ]
机构
[1] Univ Macau, State Key Lab Analog & Mixed Signal VLSI, Macau, Peoples R China
[2] Univ Macau, Dept Elect & Comp Engn, Fac Sci & Technol, Macau, Peoples R China
[3] Chengdu Univ Technol, Engn & Tech Coll, Leshan, Sichuan, Peoples R China
[4] Neijiang Normal Univ, Coll Comp Sci, Neijiang, Sichuan, Peoples R China
关键词
Implantable; galvanic coupling human-body communication; attenuation; measurement;
D O I
10.3233/THC-161229
中图分类号
R19 [保健组织与事业(卫生事业管理)];
学科分类号
摘要
In this study, an experiment was designed to verify the low power consumption of galvanic coupling human-body communication. A silver electrode (silver content: 99%) is placed in a pig leg and a sine wave signal with the power of 0 dBm is input. Compared with radio frequency communication and antenna transmission communication, attenuation is reduced by approximately 10 to 15 dB, so channel characteristics are highly improved.
引用
收藏
页码:821 / 826
页数:6
相关论文
共 9 条
  • [1] Path loss modeling and verification of implantable human-body communication based on a point source field
    Yan, Dan-Dan
    Zhang, Tao
    Zhang, Shuang
    TECHNOLOGY AND HEALTH CARE, 2020, 28 (03) : 283 - 292
  • [2] Noise Measurement and Analysis in a Power Line Communication Channel
    Rajkumarsingh, B.
    Sokappadu, B. N.
    EMERGING TRENDS IN ELECTRICAL, ELECTRONIC AND COMMUNICATIONS ENGINEERING, ELECOM 2016, 2017, 416 : 81 - 93
  • [3] Bluetooth Signal Attenuation Analysis in Human Body Tissue Analogues
    Christoe, Michael J.
    Yuan, Jinhong
    Michael, Aron
    Kalantar-Zadeh, Kourosh
    IEEE ACCESS, 2021, 9 : 85144 - 85150
  • [4] Biometric Identity Based on Intra-Body Communication Channel Characteristics and Machine Learning
    Khorshid, Ahmed E.
    Alquaydheb, Ibrahim N.
    Kurdahi, Fadi
    Jover, Roger Piqueras
    Eltawil, Ahmed
    SENSORS, 2020, 20 (05)
  • [5] MHz-Band RF Signal Propagation Characteristics on Human Body for Intra-body communication
    Ishida, Masaki
    Nakamura, Tomonori
    Nozawa, Mami
    Watanabe, Naoto
    Kado, Yuichi
    Shinagawa, Mitsuru
    2014 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE (I2MTC) PROCEEDINGS, 2014, : 797 - 801
  • [6] A Visible Human Data-Based Whole-Body Model for Investigating the Transmission Attenuation of Intrabody Communication
    Wei, Ziliang
    Gao, Yueming
    Chen, Zhizhang
    Pun, Sio Hang
    Vai, Mang, I
    Du, Min
    IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2022, 70 (08) : 3827 - 3837
  • [7] Coupling characteristics analysis of micro-emulsion cutting fluid for thickness on-machine measurement using ultrasonic
    Liu, Haibo
    Lyu, Jiarui
    Liu, Baoliang
    Lian, Meng
    Li, Te
    Wang, Yongqing
    ULTRASONICS, 2021, 115
  • [8] Measurement and analysis on human torque/angular velocity/ angle position characteristics using isoviscous loading
    Multimedia Center, Shiga University of Medical Science, Seta Tsukinowa-cho, Otsu-shi, Shiga, 520-2192, Japan
    Nihon Kikai Gakkai Ronbunshu C, 2009, 750 (343-348): : 343 - 348