Design and characterization of flattening filter for high dose rate 192Ir and 60Co Leipzig applicators used in skin cancer brachytherapy: A Monte Carlo study

被引:4
作者
Ghobadi, Parvin [1 ]
Farhood, Bagher [1 ]
Ghorbani, Mahdi [2 ]
Mohseni, Mehran [1 ]
机构
[1] Kashan Univ Med Sci, Fac Paramed Sci, Dept Med Phys & Radiol, Kashan, Iran
[2] Shahid Beheshti Univ Med Sci, Sch Med, Biomed Engn & Med Phys Dept, Tehran, Iran
关键词
Skin cancer; High dose rate brachytherapy; Ir-192; Co-60; Leipzig applicator; Flattening filter; VALENCIA; PHANTOM;
D O I
10.1016/j.compbiomed.2020.103878
中图分类号
Q [生物科学];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose: This study aimed to design optimal flattening filters for high dose rate (HDR) Ir-192 and Co-60 Leipzig applicators which are used to treat skin cancer. Materials and methods: MCNPX Monte Carlo code was used to design flattening filters for Leipzig applicators with inner diameters of 1, 2 and 3 cm. Then, their dosimetric characterizations such as dose distribution, dose profile, percentage depth dose, flatness, symmetry and homogeneity were evaluated in a 20 x 20 x 20 cm(3) water phantom and compared with those without the flattening filter. Results: The flattening filter thickness varied from 0 mm (at the edge) to the maximum values of 0.30, 1.18, and 2.41 mm for the Ir-192 Leipzig applicators of H1, H2, and H3 type, respectively. This quantity has maximum values of 0.96, 6.27, and 12.31 mm for the Co-60 double wall applicators of D1, D2, and D3 type, respectively. The dose profile flatness values for the H1, H2, and H3 Ir-192 Leipzig applicators with the optimal flattening filters were 0.76, 1.26, and 1.85%, respectively. Furthermore, the dose profile flatness values for the D1, D2, and D3 Co-60 double wall applicators with the optimal flattening filters were 1.11, 2.10 and 3.12%, respectively. The dose profile symmetry values obtained from various source-applicator combinations were less than 1.02. Compared to the applicators without flattening filter, the homogeneity values for the H1, H2, and H3 Ir-192 Leipzig applicators with the optimal flattening filters were improved 1.68, 6.51, and 13.17 times, respectively, and for the D1, D2, and D3 Co-60 double wall applicators were improved 1.23, 6.21 and 9.54 times, respectively. Conclusion: The findings revealed that the inhomogeneous dose distribution resulted from the Leipzig applicators without the optimal flattening filter at the treatment surface could be improved by insertion of optimal lead flattening filters between the sources and treatment surface.
引用
收藏
页数:12
相关论文
共 29 条
  • [21] Intraoperative high-dose-rate 192Ir radical implant in early breast cancer:: A quality assurance and dosimetry study
    Narayanan, SS
    Goel, V
    Sarin, R
    Jalali, R
    Shrivastava, SK
    Deshpande, DD
    Dinshaw, KA
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 56 (03): : 690 - 696
  • [22] Direction modulated brachytherapy (DMBT) for treatment of cervical cancer: A planning study with 192 Ir, 60Co, and 169Yb HDR sources
    Safigholi, Habib
    Han, Dae Yup
    Mashouf, Shahram
    Soliman, Abraam
    Meigooni, Ali S.
    Owrangi, Amir
    Song, William Y.
    MEDICAL PHYSICS, 2017, 44 (12) : 6538 - 6547
  • [23] Treatment planning study of the 3D dosimetric differences between Co-60 and Ir-192 sources in high dose rate (HDR) brachytherapy
    Palmer, Antony
    Hayman, Orla
    Muscat, Sarah
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2012, 4 (01) : 52 - 59
  • [24] Determination of variation in dosimetric parameters of treatment planning with Co-60 and Ir-192 sources in high dose rate brachytherapy of cervical carcinoma
    Srivastava, Shraddha
    Singh, Navin
    Varghese, Misha
    RADIATION PHYSICS AND CHEMISTRY, 2022, 196
  • [25] The dosimetric impact of heterogeneity corrections in high-dose-rate 192Ir brachytherapy for cervical cancer: Investigation of both conventional Point-A and volume-optimized plans
    Hyer, Daniel E.
    Sheybani, Arshin
    Jacobson, Geraldine M.
    Kim, Yusung
    BRACHYTHERAPY, 2012, 11 (06) : 515 - 520
  • [26] Monte carlo study on dose distributions around 192Ir, 169Yb, and 125I brachytherapy sources using EGSnrc-based egs_brachy user-code
    Mishra, Subhalaxmi
    Mishra, Bibekananda
    Selvam, T. Palani
    Deshpande, Sudesh
    Pathan, Munir Shabbir
    Kumar, Rajesh
    JOURNAL OF MEDICAL PHYSICS, 2022, 47 (03) : 270 - 278
  • [27] Consequences of removing the flattening filter from linear accelerators in generating high dose rate photon beams for clinical applications: A Monte Carlo study verified by measurement
    Parsai, E. Ishmael
    Pearson, David
    Kvale, Thomas
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2007, 261 (1-2) : 755 - 759
  • [28] Technical note: Dosimetric study for the new BEBIG 60Co HDR source used in brachytherapy in water and different media using Monte Carlo N-Particle eXtended code
    Elboukhari, Said
    Yamni, Khalid
    Ouabi, Hmad
    Bouassa, Taoufiq
    Mlouk, Lahcen Ait
    APPLIED RADIATION AND ISOTOPES, 2020, 159
  • [29] Second primary malignancies after high-dose-rate 60Co photon or 252Cf neutron brachytherapy in conjunction with external-beam radiotherapy for endometrial cancer
    Janulionis, Ernestas
    Samerdokiene, Vitalija
    Valuckas, Konstantinas Povilas
    Atkocius, Vydmantas
    Rivard, Mark J.
    BRACHYTHERAPY, 2018, 17 (05) : 768 - 774