共 50 条
Comparison of the bioMerieux NucliSENS EasyQ HIV-1 v2.0-HIV-1 RNA quantification assay versus Abbott RealTime HIV-1 and Roche Cobas TaqMan HIV-1 v2.0 on current epidemic HIV-1 variants
被引:18
作者:
Mourez, Thomas
[1
,2
,3
]
Delaugerre, Constance
[4
,5
]
Vray, Muriel
[6
]
Lemee, Veronique
[3
]
Simon, Francois
[4
,5
]
Plantier, Jean-Christophe
[1
,2
,3
]
机构:
[1] Normandie Univ, Caen, France
[2] Univ Rouen, IRIB, Equipe Accueil 2656, F-76183 Rouen, France
[3] CHU Rouen, Lab Virol, Associe Ctr Natl Reference VIH, F-76031 Rouen, France
[4] Univ Paris 07, F-75010 Paris, France
[5] Hop St Louis, AP HP, Lab Virol, F-75010 Paris, France
[6] Inst Pasteur, Unite Epidemiol Malad Emergentes, F-75014 Paris, France
关键词:
HIV-1;
Quantification;
Genetic diversity;
HIV-O;
Viral load;
IMMUNODEFICIENCY-VIRUS TYPE-1;
GROUP-O;
AMPLICOR;
DISCREPANCIES;
DIVERSITY;
SUBTYPES;
VERSION;
IMPACT;
D O I:
10.1016/j.jcv.2015.08.007
中图分类号:
Q93 [微生物学];
学科分类号:
071005 ;
100705 ;
摘要:
Background: An improved version of the bioMerieux NucliSENS (R) EasyQ (R) HIV-1 v2.0 has been introduced to overcome the underquantification observed with previous versions, especially with non-B HIV-1 subtypes. Objectives: Comparing bioMerieux NucliSENS (R) EasyQ (R) HIV-1 v2.0 versus Roche Cobas CA/CTM v2.0 and Abbott RealTime HIV-1 assays for HIV-1 group M and non-M (N, O, P) viral load measurement. Study design: The three assays were tested in parallel on 103HIV-1 group M plasma samples, and on non-group M HIV-1 culture supernatants. Results: Values obtained for the 103HIV-1 group M plasma samples tested with bioMerieux assay showed good overall correlation compared to the 2 others. The Roche Cobas assay gave higher values than the bioMerieux assay, while Abbott and bioMerieux both displayed similar results. However, analysis showed a wider dispersion in results when comparing the bioMerieux NucliSENS (R) EasyQ (R) HIV-1 v2.0 and the other 2 techniques. All data taken into account, we observed frequent discrepancies in quantification, of the plasma samples and major differences above 1 log in 10/72 (13.8%). The quantification of non-M HIV groups in culture supernatant has shown variable results, with better quantification of HIV-O and of HIV-N respectively with the Abbott assay and the bioMerieux assay. Conclusions: The bioMerieux NucliSENS (R) EasyQ (R) HIV-1 v2.0 showed improved sensitivity to non-B HIV-M subtypes compared to previous versions. Notwithstanding, we observed frequent discrepancies and a wide dispersion in results when comparing bioMerieux NucliSENS (R) EasyQ (R) HIV-1 v2.0 and the other 2 techniques. (C) 2015 Elsevier B.V. All rights reserved.
引用
收藏
页码:76 / 81
页数:6
相关论文