Development and evaluation of endurance test system for ventricular assist devices

被引:9
作者
Sumikura, Hirohito [1 ]
Homma, Akihiko [2 ]
Ohnuma, Kentaro [1 ]
Taenaka, Yoshiyuki [1 ]
Takewa, Yoshiaki [1 ]
Mukaibayashi, Hiroshi [3 ]
Katano, Kazuo [3 ]
Tatsumi, Eisuke [1 ]
机构
[1] Natl Cerebral & Cardiovasc Ctr, Res Inst, Dept Artificial Organs, Suita, Osaka 5658565, Japan
[2] Tokyo Denki Univ, Sch Sci & Engn, Saitama, Japan
[3] IWAKI Co Ltd, Saitama, Japan
关键词
Endurance test; Circulation circuit; Ventricular assist devices (VADs); Pulse duplicator; Field: artificial heart (basic); CENTRIFUGAL BLOOD PUMP; CLINICAL-EXPERIENCE; MOCK CIRCULATION; CHILDREN;
D O I
10.1007/s10047-013-0687-3
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We developed a novel endurance test system that can arbitrarily set various circulatory conditions and has durability and stability for long-term continuous evaluation of ventricular assist devices (VADs), and we evaluated its fundamental performance and prolonged durability and stability. The circulation circuit of the present endurance test system consisted of a pulsatile pump with a small closed chamber (SCC), a closed chamber, a reservoir and an electromagnetic proportional valve. Two duckbill valves were mounted in the inlet and outlet of the pulsatile pump. The features of the circulation circuit are as follows: (1) the components of the circulation circuit consist of optimized industrial devices, giving durability; (2) the pulsatile pump can change the heart rate and stroke length (SL), as well as its compliance using the SCC. Therefore, the endurance test system can quantitatively reproduce various circulatory conditions. The range of reproducible circulatory conditions in the endurance test circuit was examined in terms of fundamental performance. Additionally, continuous operation for 6 months was performed in order to evaluate the durability and stability. The circulation circuit was able to set up a wide range of pressure and total flow conditions using the SCC and adjusting the pulsatile pump SL. The long-term continuous operation test demonstrated that stable, continuous operation for 6 months was possible without leakage or industrial device failure. The newly developed endurance test system demonstrated a wide range of reproducible circulatory conditions, durability and stability, and is a promising approach for evaluating the basic characteristics of VADs.
引用
收藏
页码:138 / 148
页数:11
相关论文
共 23 条
  • [1] Blood pumps: Technologies and markets in transformation
    Anderson, DW
    [J]. ARTIFICIAL ORGANS, 2001, 25 (05) : 406 - 410
  • [2] Normal ranges of heart rate and respiratory rate in children from birth to 18 years of age: a systematic review of observational studies
    Fleming, Susannah
    Thompson, Matthew
    Stevens, Richard
    Heneghan, Carl
    Plueddemann, Annette
    Maconochie, Ian
    Tarassenko, Lionel
    Mant, David
    [J]. LANCET, 2011, 377 (9770) : 1011 - 1018
  • [3] Frazier OH, 2004, TEX HEART I J, V31, P157
  • [4] Initial clinical experience with the Jarvik 2000 implantable axial-flow left ventricular assist system
    Frazier, OH
    Myers, TJ
    Gregoric, ID
    Khan, T
    Delgado, R
    Croitoru, M
    Miller, K
    Jarvik, R
    Westaby, S
    [J]. CIRCULATION, 2002, 105 (24) : 2855 - 2860
  • [5] In Vivo Experience of the Child-Size Pediatric Jarvik 2000 Heart: Update
    Gibber, Marc
    Wu, Zhongjun J.
    Chang, Won-Bae
    Bianchi, Giacomo
    Hu, Jingping
    Garcia, Jose
    Jarvik, Robert
    Griffith, Bartley P.
    [J]. ASAIO JOURNAL, 2010, 56 (04) : 369 - 376
  • [6] Griffith K, 2009, PERFUSION-UK, V24, P83, DOI [10.1177/0267659109106826, 10.1177/0267659109106903]
  • [7] Kiani A, 2003, IJMS, V28, P169
  • [8] Kitao T, 2011, ARTIF ORGANS, V35, P543, DOI [10.1111/j.1525-1594.2011.01321.x, 10.1111/j.1525-1594.2010.01192.x]
  • [9] Development and initial testing of a pediatric centrifugal blood pump
    Litwak, P
    Butler, KC
    Thomas, DC
    Taylor, LP
    Macha, M
    Yamazaki, K
    Konishi, H
    Kormos, RL
    Griffith, BP
    Borovetz, HS
    [J]. ANNALS OF THORACIC SURGERY, 1996, 61 (01) : 448 - 451
  • [10] Durability testing of components for the Jarvik 2000 completely implantable axial flow left ventricular assist device
    Marlinski, E
    Jacobs, G
    Deirmengian, C
    Jarvik, R
    [J]. ASAIO JOURNAL, 1998, 44 (05) : M741 - M744