Dosimetric investigation of LDR brachytherapy 192Ir wires by Monte Carlo and TPS calculations

被引:3
作者
Bozkurt, Ahmet [1 ]
Acun, Hediye [2 ]
Kemikler, Gonul [3 ]
机构
[1] Harran Univ, Dept Phys, Fac Arts & Sci, TR-63300 Sanliurfa, Turkey
[2] Harran Univ, Dept Biophys, Fac Med, TR-63300 Sanliurfa, Turkey
[3] Istanbul Univ, Dept Med Phys, Inst Oncol, TR-34390 Istanbul, Turkey
关键词
Monte Carlo; LDR brachytherapy; Ir-192; wires; Dose rate; INTERSTITIAL BRACHYTHERAPY; DOSE DISTRIBUTION; DISTRIBUTIONS;
D O I
10.1007/s11604-012-0140-1
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose The aim of this study was to investigate the dose rate distribution around Ir-192 wires used as radioactive sources in low-dose-rate brachytherapy applications. Materials and methods Monte Carlo modeling of a 0.3-mm diameter source and its surrounding water medium was performed for five different wire lengths (1-5 cm) using the MCNP software package. Results The computed dose rates per unit of air kerma at distances from 0.1 up to 10 cm away from the source were first verified with literature data sets. Then, the simulation results were compared with the calculations from the XiO CMS commercial treatment planning system. Conclusion The study results were found to be in concordance with the treatment planning system calculations except for the shorter wires at close distances.
引用
收藏
页码:24 / 30
页数:7
相关论文
共 14 条
  • [1] Experimental determination of dosimetry functions of Ir-192 sources
    Anctil, JC
    Clark, BG
    Arsenault, CJ
    [J]. MEDICAL PHYSICS, 1998, 25 (12) : 2279 - 2287
  • [2] MONTE-CARLO TECHNIQUES IN MEDICAL RADIATION PHYSICS
    ANDREO, P
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (07) : 861 - 920
  • [3] Assessment of the relative dose distribution around an 192Ir line source using a liquid ionization chamber
    Bahar-Gogani, J
    Wickman, G
    Johansson, L
    Johansson, BE
    [J]. MEDICAL PHYSICS, 1999, 26 (09) : 1932 - 1942
  • [4] Monte Carlo calculation of dose rate distributions around Ir-192 wires
    Ballester, F
    Hernandez, C
    PerezCalatayud, J
    Lliso, F
    [J]. MEDICAL PHYSICS, 1997, 24 (08) : 1221 - 1228
  • [5] Briesmeister J.F., 1997, ALAMOS NATL LAB REPO
  • [6] The dose distribution close to an Ir-192 wire source: EGS4 Monte Carlo calculations
    Cheung, YC
    Yu, PKN
    Young, ECM
    Chan, CL
    Ng, MF
    Tang, F
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (02) : 401 - 406
  • [7] Dutreix A, 1987, MODERN BRACHYTHERAPY, P17
  • [8] Dosimetric characteristics of 192Ir sources used in interstitial brachytherapy
    Ghiassi-Nejad, M
    Jafarizadeh, M
    Ahmadian-Pour, MR
    Ghahramani, AR
    [J]. APPLIED RADIATION AND ISOTOPES, 2001, 55 (02) : 189 - 195
  • [9] COMPARISON OF MEASURED AND CALCULATED DOSE DISTRIBUTIONS AROUND AN IR-192 WIRE
    GILLIN, MT
    LOPEZ, F
    KLINE, RW
    GRIMM, DF
    NIROOMANDRAD, A
    [J]. MEDICAL PHYSICS, 1988, 15 (06) : 915 - 918
  • [10] Dosimetry of 192Ir wires for LDR interstitial brachytherapy following the AAPM TG-43 dosimetric formalism
    Karaiskos, P
    Papagiannis, P
    Angelopoulos, A
    Sakelliou, L
    Baltas, D
    Sandilos, P
    Vlachos, L
    [J]. MEDICAL PHYSICS, 2001, 28 (02) : 156 - 166