Application of temporal subtraction for detection of interval changes on chest radiographs: Improvement of subtraction images using automated initial image matching

被引:56
作者
Ishida, T
Ashizawa, K
Engelmann, R
Katsuragawa, S
MacMahon, H
Doi, K
机构
[1] Univ Chicago, Dept Radiol, Chicago, IL 60637 USA
[2] Univ Chicago, Kurt Rossmann Labs Radiol Image Res, Chicago, IL 60637 USA
关键词
computer-aided diagnosis; digital image subtraction; image matching; interval change; chest radiograph;
D O I
10.1007/BF03168846
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The authors developed a temporal subtraction scheme based on a nonlinear geometric warping technique to assist radiologists in the detection of interval changes in chest radiographs obtained on different occasions. The performance of the current temporal subtraction scheme is reasonably good; however, severe misregistration can occur in some cases. The authors evaluated the quality of 100 chest temporal subtraction images selected from their clinical image database. Severe misregistration was mainly attributable to initial incorrect global matching. Therefore, they attempted to improve the qualify of the subtraction images by applying a new initial image matching technique to determine the global shift value between the current and the previous chest images. A crosscorrelation method was employed for the initial image matching by use of blurred low-resolution chest images. Nineteen Cases (40.4%) among 47 poor registered subtraction images were improved, These results show that the new initial image matching technique is very effective for improving the quality of chest temporal subtraction images, which can greatly enhance subtle changes in chest radiographs. Copyright (C) 1999 by W.B. Saunders Company.
引用
收藏
页码:77 / 86
页数:10
相关论文
共 6 条
[1]   Digital chest radiography: Effect of temporal subtraction images on detection accuracy [J].
Difazio, MC ;
MacMahon, H ;
Xu, XW ;
Tsai, P ;
Shiraishi, J ;
Armato, SG ;
Doi, K .
RADIOLOGY, 1997, 202 (02) :447-452
[2]   DIGITAL IMAGE SUBTRACTION OF TEMPORALLY SEQUENTIAL CHEST IMAGES FOR DETECTION OF INTERVAL CHANGE [J].
KANO, A ;
DOI, K ;
MACMAHON, H ;
HASSELL, DD ;
GIGER, ML .
MEDICAL PHYSICS, 1994, 21 (03) :453-461
[3]   COMPUTED RADIOGRAPHY UTILIZING SCANNING LASER STIMULATED LUMINESCENCE [J].
SONODA, M ;
TAKANO, M ;
MIYAHARA, J ;
KATO, H .
RADIOLOGY, 1983, 148 (03) :833-838
[4]  
TAKEO H, 1994, P SOC PHOTO-OPT INS, V2163, P98, DOI 10.1117/12.174245
[5]   IMAGE FEATURE ANALYSIS FOR COMPUTER-AIDED DIAGNOSIS - ACCURATE DETERMINATION OF RIBCAGE BOUNDARY IN CHEST RADIOGRAPHS [J].
XU, XW ;
DOI, K .
MEDICAL PHYSICS, 1995, 22 (05) :617-626
[6]  
YOSHIMURA H, 1993, MED PHYS, V20, P179