The Aer-O-Scope:: Proof of the concept of a pneumatic, skill-independent, self-propelling, self-navigating colonoscope in a pig model

被引:42
作者
Pfeffer, J
Grinshpon, R
Rex, D
Levin, B
Rösch, T
Arber, N
Halpern, Z
机构
[1] Tel Aviv Sourasky Med Ctr, Integrated Canc Prevent Ctr, IL-64239 Tel Aviv, Israel
[2] Tel Aviv Med Ctr & Sch Med, Dept Gastroenterol, Tel Aviv, Israel
[3] Indiana Univ, Med Ctr, Div Gastroenterol & Hepatol, Dept Med, Indianapolis, IN 46204 USA
[4] Univ Texas, MD Anderson Canc Ctr, Houston, TX 77030 USA
[5] Charite Med Ctr, Ctr Interdisciplinary Endoscopy Unit, Berlin, Germany
[6] Tel Aviv Univ, Dept Internal Med, Sackler Fac Med, IL-69978 Tel Aviv, Israel
关键词
D O I
10.1055/s-2006-925089
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
Background and Study Aims: Considerable training is needed to enable endoscopists to use the currently available commercial colonoscopes and sigmoidoscopes effectively and safely. The aim of this study was to evaluate the safety and efficacy of the propulsion mechanism incorporated into the Aer-O-Scope (R) (GI View Ltd., Ramat Gan, Israel) - a novel self-propelled, self-navigating colonic endoscope for diagnostic purposes. Materials and Methods: Twenty young female pigs underwent complete bowel preparation followed by a sedated examination using the new device. Ten pigs underwent two consecutive procedures, with a wash-out period of 7 days between each procedure. The total number of procedures was 30. Two prototypes of the Aer-O-Scope, with different cable lengths and vehicle balloon sizes (n = 20 and n = 10 for prototypes I and II, respectively) were used. Each examination was followed by a standard colonoscopy for safety evaluation. The insertion length of the Aer-O-Scope was determined by fluoroscopy images. Results: The colon was adequately clean in 25 procedures. Maximum insertion was achieved in 21 procedures (84%) - 80% with prototype I (n = 15) and 90% with prototype 11 (n = 10). The time to maximum insertion averaged 8.9 +/- 4.4 min (10 +/- 4.6 and 6.6 +/- 2.9 min for prototypes I and II; P < 0.05), and the withdrawal time averaged 3.4 +/- 2.1 and 4.2 +/- 3.4 min, respectively. The driving pressures for prototypes I and II averaged 46.3 and 34.5 mbar, respectively. The follow-up conventional colonoscopy identified no mucosal tears or perforations. Minor mucosal petechiae were noted in 43% of the cases. No adverse events were noted up to 7 days after the procedure. Conclusions: The propulsion mechanism used in this novel self-propelled, self-navigating colonoscope was effective and safe in pigs.
引用
收藏
页码:144 / 148
页数:5
相关论文
共 15 条
  • [1] Quality improvement programme to achieve acceptable colonoscopy completion rates: prospective before and after study
    Ball, JE
    Osbourne, J
    Jowett, S
    Pellen, M
    Welfare, MR
    [J]. BMJ-BRITISH MEDICAL JOURNAL, 2004, 329 (7467): : 665 - 667
  • [2] A prospective study of colonoscopy practice in the UK today: are we adequately prepared for national colorectal cancer screening tomorrow?
    Bowles, CJA
    Leicester, R
    Romaya, C
    Swarbrick, E
    Williams, CB
    Epstein, O
    [J]. GUT, 2004, 53 (02) : 277 - 283
  • [3] Randomized, controlled trial of carbon dioxide insufflation during colonoscopy
    Church, J
    Delaney, C
    [J]. DISEASES OF THE COLON & RECTUM, 2003, 46 (03) : 322 - 326
  • [4] COTTON PB, 2003, GASTROINTEST ENDOSC, V57, P198
  • [5] Risk of perforation after colonoseopy and sigmoidoscopy: A population-based study
    Gatto, NM
    Frucht, H
    Sundararajan, V
    Jacobson, JS
    Grann, VR
    Neugut, AI
    [J]. JOURNAL OF THE NATIONAL CANCER INSTITUTE, 2003, 95 (03) : 230 - 236
  • [6] Levin TR, 2004, GASTROENTEROLOGY, V126, pA25
  • [7] MANCIASSI A, 2002, P 2002 IEEE RSJ INT
  • [8] Ng W. S., 2000, US Patent, Patent No. [6,162,171, 6162171]
  • [9] Colorectal cancer prevention 2000: Screening recommendations of the American College of Gastroenterology
    Rex, DK
    Johnson, DA
    Lieberman, DA
    Burt, RW
    Sonnenberg, A
    [J]. AMERICAN JOURNAL OF GASTROENTEROLOGY, 2000, 95 (04) : 868 - 877
  • [10] Colonography using multislice CT
    Rogalla, P
    Meiri, N
    Rückert, JC
    Hamm, B
    [J]. EUROPEAN JOURNAL OF RADIOLOGY, 2000, 36 (02) : 81 - 85