Repeatability of Arterial Spin Labeling MRI in Measuring Blood Perfusion in the Human Eye

被引:16
作者
Khanal, Safal [1 ]
Turnbull, Philip R. K. [1 ]
Vaghefi, Ehsan [1 ]
Phillips, John R. [1 ,2 ]
机构
[1] Univ Auckland, Sch Optometry & Vis Sci, 85 Pk Rd, Auckland 1071, New Zealand
[2] Asia Univ, Dept Optometry, Taichung, Taiwan
关键词
arterial spin labeling; chorio-retinal perfusion; ischemic retinal diseases; magnetic resonance imaging; CHOROIDAL THICKNESS; FLOW QUANTIFICATION; RELIABILITY; CIRCULATION; INVERSION; RETINA; COIL;
D O I
10.1002/jmri.26323
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background Quantifying blood perfusion in ocular tissues is challenging, partly because the majority of the blood is carried by the choroid, which is difficult to visualize because it is located between the retina and sclera. Purpose/Hypothesis To evaluate the intra- and interday repeatability of MRI measures of chorio-retinal blood perfusion. Study Type Prospective, cross-sectional, observational study. Population Twenty young healthy adults (six male, age: 25 +/- 5 years) scanned twice within a single session repeated at the same time of day on 2 days. Field Strength/Sequence Arterial spin labeling (ASL) MRI at 3.0T using pseudocontinuous ASL (PCASL) labeling scheme and a 3D turbo-gradient-spin-echo (TGSE) acquisition, including axial T-2-weighted structural images using a 2D turbo-spin-echo (TSE) sequence. Assessments Region-of-interest analysis for assessment of chorio-retinal blood perfusion. Statistical Tests Intra- and interday repeatability of measures analyzed using intraclass correlation coefficients (ICC), Pearson's correlation analysis, paired t-tests, and Bland-Altman plots. Results The mean chorio-retinal perfusion was 77.86 (standard deviation [SD] = 29.80) ml/100ml/min. Perfusion measurements correlated strongly within a single session (r = 0.95, 95% confidence interval [CI] [0.880-0.980], P < 0.001) and between the two sessions based on a single run (r = 0.80 [0.582-0.913], P < 0.001), and two runs (r = 0.80 [0.479-0.918], P < 0.001). There were mean differences of 2.69 [16.85 to -22.23] ml/100ml/min for intraday measures, -7.44 [27.45 to -42.32] ml/100ml/min for single-run interday measures, and 5.73 [28.71 to -40.17] ml/100ml/min for two-run interday measures, but none were significant (all P > 0.05). Data Conclusion Quantitative ASL-MRI measurements of chorio-retinal blood perfusion showed high intra- and interday repeatability. The ASL-MRI technique provides reliable measures of chorio-retinal perfusion in vivo.
引用
收藏
页码:966 / 974
页数:9
相关论文
共 45 条
  • [1] Recommended Implementation of Arterial Spin-Labeled Perfusion MRI for Clinical Applications: A Consensus of the ISMRM Perfusion Study Group and the European Consortium for ASL in Dementia
    Alsop, David C.
    Detre, John A.
    Golay, Xavier
    Guenther, Matthias
    Hendrikse, Jeroen
    Hernandez-Garcia, Luis
    Lu, Hanzhang
    MacIntosh, Bradley J.
    Parkes, Laura M.
    Smits, Marion
    van Osch, Matthias J. P.
    Wang, Danny J. J.
    Wong, Eric C.
    Zaharchuk, Greg
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2015, 73 (01) : 102 - 116
  • [2] Measurement of Choroidal Thickness Following Caffeine Intake in Healthy Subjects
    Altinkaynak, Hasan
    Ceylan, Erdinc
    Kartal, Baki
    Keles, Sadullah
    Ekinci, Metin
    Olcaysu, Osman Okan
    [J]. CURRENT EYE RESEARCH, 2016, 41 (05) : 708 - 714
  • [3] A general kinetic model for quantitative perfusion imaging with arterial spin labeling
    Buxton, RB
    Frank, LR
    Wong, EC
    Siewert, B
    Warach, S
    Edelman, RR
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1998, 40 (03) : 383 - 396
  • [4] Test-Retest Reliability of Arterial Spin Labeling With Common Labeling Strategies
    Chen, Yufen
    Wang, Danny J. J.
    Detre, John A.
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2011, 33 (04) : 940 - 949
  • [5] Effect of retinal image defocus on the thickness of the human choroid
    Chiang, Samuel T-H.
    Phillips, John R.
    Backhouse, Simon
    [J]. OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2015, 35 (04) : 405 - 413
  • [6] COLLIGNON A, 1995, COMP IMAG VIS, V3, P263
  • [7] Continuous Flow-Driven Inversion for Arterial Spin Labeling Using Pulsed Radio Frequency and Gradient Fields
    Dai, Weiying
    Garcia, Dairon
    de Bazelaire, Cedric
    Alsop, David C.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2008, 60 (06) : 1488 - 1497
  • [8] de Carlo Talisa E, 2015, Int J Retina Vitreous, V1, P5
  • [9] Applications of arterial spin labeled MRI in the brain
    Detre, John A.
    Rao, Hengyi
    Wang, Danny J. J.
    Chen, Yu Fen
    Wang, Ze
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2012, 35 (05) : 1026 - 1037
  • [10] Structural Correlation-based Outlier Rejection (SCORE) algorithm for arterial spin labeling time series
    Dolui, Sudipto
    Wang, Ze
    Shinohara, Russell T.
    Wolk, David A.
    Detre, John A.
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2017, 45 (06) : 1786 - 1797