Direct determination of glucose, lactate and triglycerides in blood serum by a tunable quantum cascade laser-based mid-IR sensor

被引:68
作者
Brandstetter, M. [1 ]
Volgger, L. [1 ]
Genner, A. [1 ]
Jungbauer, C. [2 ]
Lendl, B. [1 ]
机构
[1] Vienna Univ Technol, Inst Chem Technol & Analyt, A-1060 Vienna, Austria
[2] Austrian Red Cross, Blood Serv Vienna Lower Austria & Burgenland, A-1040 Vienna, Austria
来源
APPLIED PHYSICS B-LASERS AND OPTICS | 2013年 / 110卷 / 02期
关键词
TRANSFORM INFRARED-SPECTROSCOPY; CRITICALLY-ILL PATIENTS; WHOLE-BLOOD; INTERFERENCES; SPECTROMETRY; ICODEXTRIN; NOISE;
D O I
10.1007/s00340-012-5080-z
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
This work reports on a compact sensor for fast and reagent-free point-of-care determination of glucose, lactate and triglycerides in blood serum based on a tunable (1030-1230 cm(-1)) external-cavity quantum cascade laser (EC-QCL). For simple and robust operation a single beam set-up was designed and only thermoelectric cooling was used for the employed laser and detector. Full computer control of analysis including liquid handling and data analysis facilitated routine measurements. A high optical pathlength (> 100 mu m) is a prerequisite for robust measurements in clinical practice. Hence, the optimum optical pathlength for transmission measurements in aqueous solution was considered in theory and experiment. The experimentally determined maximum signal-to-noise ratio (SNR) was around 140 mu m for the QCL blood sensor and around 50 mu m for a standard FT-IR spectrometer employing a liquid nitrogen cooled mercury cadmium telluride (MCT) detector. A single absorption spectrum was used to calculate the analyte concentrations simultaneously by using a partial-least-squares (PLS) regression analysis. Glucose was determined in blood serum with a prediction error (RMSEP) of 6.9 mg/dl and triglycerides with an error of cross-validation (RMSECV) of 17.5 mg/dl in a set of 42 different patients. In spiked serum samples the lactate concentration could be determined with an RMSECV of 8.9 mg/dl.
引用
收藏
页码:233 / 239
页数:7
相关论文
共 30 条
[1]   Infrared spectroscopy of proteins [J].
Barth, Andreas .
BIOCHIMICA ET BIOPHYSICA ACTA-BIOENERGETICS, 2007, 1767 (09) :1073-1101
[2]   Infrared intensities of liquids .20. The intensity of the OH stretching band of liquid water revisited, and the best current values of the optical constants of H2O(1) at 25 degrees C between 15,000 and 1 cm(-1) [J].
Bertie, JE ;
Lan, ZD .
APPLIED SPECTROSCOPY, 1996, 50 (08) :1047-1057
[3]   GLUCOSE DETERMINATION IN SIMULATED BLOOD-SERUM SOLUTIONS BY FOURIER-TRANSFORM INFRARED-SPECTROSCOPY - INVESTIGATION OF SPECTRAL INTERFERENCES [J].
BHANDARE, P ;
MENDELSON, Y ;
STOHR, E ;
PEURA, RA .
VIBRATIONAL SPECTROSCOPY, 1994, 6 (03) :363-378
[4]   Tunable mid-infrared lasers in physical chemosensors towards the detection of physiologically relevant parameters in biofluids [J].
Brandstetter, Markus ;
Lendl, Bernhard .
SENSORS AND ACTUATORS B-CHEMICAL, 2012, 170 :189-195
[5]   Tunable external cavity quantum cascade laser for the simultaneous determination of glucose and lactate in aqueous phase [J].
Brandstetter, Markus ;
Genner, Andreas ;
Anic, Kresimir ;
Lendl, Bernhard .
ANALYST, 2010, 135 (12) :3260-3265
[6]   Hypertriglyceridemia [J].
Brunzell, John D. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (10) :1009-1017
[7]   OPTIMUM OPTICAL DENSITY IN SPECTROPHOTOMETRY [J].
COLE, R .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA, 1951, 41 (01) :38-40
[8]   Quantum cascade lasers in chemical physics [J].
Curl, Robert F. ;
Capasso, Federico ;
Gmachl, Claire ;
Kosterev, Anatoliy A. ;
McManus, Barry ;
Lewicki, Rafal ;
Pusharsky, Michael ;
Wysocki, Gerard ;
Tittel, Frank K. .
CHEMICAL PHYSICS LETTERS, 2010, 487 (1-3) :1-18
[9]  
de Bode C., DIABETES PROGRAMME
[10]   Applications of FT-IR spectrometry to plasma contents analysis and monitoring [J].
Déléris, G ;
Petibois, C .
VIBRATIONAL SPECTROSCOPY, 2003, 32 (01) :129-136