A comparison of two systems of patient immobilization for prostate radiotherapy

被引:13
作者
White, Peter [1 ]
Yee, Chui Ka [1 ]
Shan, Lee Chi [1 ]
Chung, Lee Wai [1 ]
Ng Ho Man [1 ]
Cheung, Yik Shing [1 ]
机构
[1] Hong Kong Polytech Univ, Kowloon, Hong Kong, Peoples R China
来源
RADIATION ONCOLOGY | 2014年 / 9卷
关键词
Immobilization; Hipfix; Alpha cradle; Set-up errors; Treatment verification; CTV-PTV margins; Prostate cancer; 3-DIMENSIONAL CONFORMAL RADIOTHERAPY; SET-UP VERIFICATION; RADIATION-THERAPY; POSITIONING ERRORS; ORGAN MOTION; CANCER; REPRODUCIBILITY; UNCERTAINTIES; IRRADIATION; CARCINOMA;
D O I
10.1186/1748-717X-9-29
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: Reproducibility of different immobilization systems, which may affect set-up errors, remains uncertain. Immobilization systems and their corresponding set-up errors influence the clinical target volume to planning target volume (CTV-PTV) margins and thus may result in undesirable treatment outcomes. This study compared the reproducibility of patient positioning with Hipfix system and whole body alpha cradle with respect to localized prostate cancer and investigated the existing CTV-PTV margins in the clinical oncology departments of two hospitals. Methods: Forty sets of data of patients with localized T1-T3 prostate cancer were randomly selected from two regional hospitals, with 20 patients immobilized by a whole-body alpha cradle system and 20 by a thermoplastic Hipfix system. Seven sets of the anterior-posterior (AP), cranial-caudal (CC) and medial-lateral (ML) deviations were collected from each patient. The reproducibility of patient positioning within the two hospitals was compared using a total vector error (TVE) parameter. In addition, CTV-PTV margins were computed using van Herk's formula. The resulting values were compared to the current CTV-PTV margins in both hospitals. Results: The TVE values were 5.1 and 2.8 mm for the Hipfix and the whole-body alpha cradle systems respectively. TVE associated with the whole-body alpha cradle system was found to be significantly less than the Hipfix system (p < 0.05). The CC axis in the Hipfix system attained the highest frequency of large (23.6%) and serious (7.9%) set-up errors. The calculated CTV to PTV margin was 8.3, 1.9 and 2.3 mm for the Hipfix system, and 2.1, 3.4 and 1.8 mm for the whole body alpha cradle in CC, ML and AP axes respectively. All but one (CC axis using Hipfix) margin calculated did not exceed the corresponding hospital protocol. The whole body alpha cradle system was found to be significantly better than the Hipfix system in terms of reproducibility (p < 0.05), especially in the CC axis. Conclusions: The whole body alpha cradle system was more reproducible than the Hipfix system. In particular, the difference in CC axis contributed most to the results and the current CC margin for the Hipfix system might be considered as inadequate.
引用
收藏
页数:12
相关论文
共 47 条
  • [1] Portal imaging for evaluation of daily on-line setup errors and off-line organ motion during conformal irradiation of carcinoma of the prostate
    Alasti, H
    Petric, MP
    Catton, CN
    Warde, PR
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 49 (03): : 869 - 884
  • [2] The dosimetric impact of daily setup error on target volumes and surrounding normal tissue in the treatment of prostate cancer with intensity-modulated radiation therapy
    Algan, Ozer
    Jamgade, Ambarish
    Ali, Imad
    Christie, Alana
    Thompson, J. Spencer
    Thompson, David
    Ahmad, Salahuddin
    Herman, Terence
    [J]. MEDICAL DOSIMETRY, 2012, 37 (04) : 406 - 411
  • [3] [Anonymous], 1993, J ICRU
  • [4] [Anonymous], 2012, Image-guided and adaptive radiation therapy
  • [5] [Anonymous], 1999, 62 ICRU
  • [6] Analysis of prostate and seminal vesicle motion: Implications for treatment planning
    Beard, CJ
    Kijewski, P
    Bussiere, M
    Gelman, R
    Gladstone, D
    Shaffer, K
    Plunkett, M
    Costello, P
    Coleman, CN
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 34 (02): : 451 - 458
  • [7] Impact of cradle immobilization on setup reproducibility during external beam radiation therapy for lung cancer
    Bentel, GC
    Marks, LB
    Krishnamurthy, R
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (03): : 527 - 531
  • [8] THE EFFECTIVENESS OF IMMOBILIZATION DURING PROSTATE IRRADIATION
    BENTEL, GC
    MARKS, LB
    SHEROUSE, GW
    SPENCER, DP
    ANSCHER, MS
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 31 (01): : 143 - 148
  • [9] Caswell N, 2005, DETERMINATION GEOMET
  • [10] PROSTATE MOTION DURING STANDARD RADIOTHERAPY AS ASSESSED BY FIDUCIAL MARKERS
    CROOK, JM
    RAYMOND, Y
    SALHANI, D
    YANG, H
    ESCHE, B
    [J]. RADIOTHERAPY AND ONCOLOGY, 1995, 37 (01) : 35 - 42