Patient doses in hysterosalpingography

被引:22
作者
Fernandez, JM [1 ]
Vano, E [1 ]
Guibelalde, E [1 ]
机构
[1] UNIV COMPLUTENSE MADRID,FAC MED,CATEDRA FIS MED,E-28040 MADRID,SPAIN
关键词
D O I
10.1259/0007-1285-69-824-751
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The results of a survey of doses imparted on 41 patients undergoing hysterosalpingography is presented. Dosimetric evaluations were carried out by measuring both the dose-area product using a transmission ionization type chamber, and the entrance surface dose using thermoluminiscent dosimeters. As a result, a local reference dose value was obtained for this examination. Correlations between the dose-area product and the entrance surface dose data were analysed and compared in order to find the most appropriate dosimetric procedure. The median dose-area product obtained was 713 cGy cm(2) (range 247 cGy cm(2)-1623 cGy cm(2)). Ovarian doses were also calculated, and a median value of 4.6 mGy was obtained for the whole examination. Effective doses were estimated with a median value of 3.1 mSv, and a range of 1.0 mSv-8.1 mSv.
引用
收藏
页码:751 / 754
页数:4
相关论文
共 13 条
[11]  
VANDERWEIDEN RMF, 1989, BRIT J OBSTET GYNAEC, V96, P471
[12]   EVOLUTION OF DIAGNOSTIC-RADIOLOGY IN A BIG HOSPITAL DURING A 5-YEAR PERIOD, AND THE DERIVED COLLECTIVE DOSE [J].
VANO, E ;
VELASCO, A ;
MORAN, P ;
GONZALEZ, L ;
PEDROSA, CSA .
BRITISH JOURNAL OF RADIOLOGY, 1993, 66 (790) :892-898
[13]  
VANO E, 1992, RADIOLOGIA, V34, P27