Evaluation of the Elekta Symmetry™ 4D IGRT System by Using a Moving Lung Phantom

被引:2
|
作者
Shin, Hun-Joo [1 ]
Kim, Shin-Wook [1 ]
Kay, Chul Seung [1 ]
Seo, Jae-Hyuk [2 ]
Lee, Gi-Woong [3 ]
Kang, Ki-Mun [4 ]
Jang, Hong Seok [5 ]
Kang, Young-nam [5 ]
机构
[1] Catholic Univ Korea, Coll Med, Incheon St Marys Hosp, Dept Radiat Oncol, Inchon 403720, South Korea
[2] Catholic Univ Korea, Coll Med, Bucheon St Marys Hosp, Dept Radiat Oncol, Bucheon 420717, South Korea
[3] Kyonggi Univ Korea, Sch Med, Dept Med Phys, Suwon 443760, South Korea
[4] Gyeong Sang Natl Univ Hosp, Dept Radiat Oncol, Chinju 660702, South Korea
[5] Catholic Univ Korea, Coll Med, Seoul St Marys Hosp, Dept Radiat Oncol, Seoul 135701, South Korea
关键词
Symmetry; 4D; IGRT; Registration; Moving phantom; BEAM COMPUTED-TOMOGRAPHY; MODULATED ARC THERAPY; RADIATION-THERAPY; CT; VERIFICATION; RADIOTHERAPY;
D O I
10.3938/jkps.67.260
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
Purpose: 4D cone-beam computed tomography (CBCT) is a beneficial tool for the treatment of movable tumors because it can help us to understand where the tumors are actually located and it has a precise treatment plan. However, general CBCT images have a limitation in that they cannot perfectly perform a sophisticated registration. On the other hand, the Symmetry (TM) 4D image-guided radiation therapy (IGRT) system of Elekta offers a 4D CBCT registration option. In this study, we evaluated the usefulness of Symmetry (TM). Method and Materials: Planning CT images of the CIRS moving lung phantom were acquired 4D multi-detector CT (MDCT), and the images were sorted as 10 phases from 0% phase to 90% phase. The thickness of the CT images was 1 mm. Acquired MDCT images were transferred to the contouring software, and a virtual target was generated. A one-arc volumetric-modulated arc therapy (VMAT) plan was performed by using the treatment planning system on the virtual target. Finally, the movement of the phantom was verified by using the XVI Symmetry (TM) system. Results: The physical movement of the CIRS moving lung phantom was +/- 10.0 mm in the superior-inferior direction, +/- 1.0 mm in the lateral direction, and +/- 2.5 mm in the anterior-posterior direction. The movement of the phantom was measured from the 4D MDCT registration as +/- 10.2 mm in the superior-inferior direction, +/- 0.9 mm in the lateral direction, and +/- 2.45 mm in the anterior-posterior direction. The movement of the phantom was measured from the Symmetry (TM) registration as +/- 10.1 mm in the superior-inferior direction, +/- 0.9 mm in the lateral direction, and +/- 2.4 mm in the anterior-posterior direction. Conclusion: We confirmed that 4D CBCT is a beneficial tool for the treatment of movable tumors, and that the 4D registration of SymmetryTM can increase the precision of the registration when a movable tumor is the target of radiation treatment.
引用
收藏
页码:260 / 263
页数:4
相关论文
共 50 条
  • [11] Four-dimensional dose reconstruction system for lung cancer VMAT treatment: a 4D phantom study
    Hashimoto, M.
    Ito, Y.
    Tanaka, Y.
    Nakano, Masahiro
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2023, 21 (03): : 467 - 474
  • [12] Establishing a framework to implement 4D XCAT Phantom for 4D radiotherapy research
    Panta, Raj K.
    Segars, Paul
    Yin, Fang-Fang
    Cai, Jing
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2012, 8 (04) : 565 - 570
  • [13] 4D CT lung ventilation images are affected by the 4D CT sorting method
    Yamamoto, Tokihiro
    Kabus, Sven
    Lorenz, Cristian
    Johnston, Eric
    Maxim, Peter G.
    Diehn, Maximilian
    Eclov, Neville
    Barquero, Cristian
    Loo, Billy W., Jr.
    Keall, Paul J.
    MEDICAL PHYSICS, 2013, 40 (10)
  • [14] Evaluation of a novel 4D in vivo dosimetry system
    Cherpak, A.
    Ding, W.
    Hallil, A.
    Cygler, J. E.
    MEDICAL PHYSICS, 2009, 36 (05) : 1672 - 1679
  • [15] Evaluation and commissioning of a surface based system for respiratory sensing in 4D CT
    Spadea, Maria Francesca
    Baroni, Guido
    Gierga, David P.
    Turcotte, Julie C.
    Chen, George T. Y.
    Sharp, Gregory C.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2011, 12 (01): : 162 - 169
  • [16] Dosimetric evaluation of 4D-CBCT reconstructed by Simultaneous Motion Estimation and Image Reconstruction (SMEIR) for carbon ion therapy of lung cancer
    Shrestha, Deepak
    Tsai, Min-Yu
    Qin, Nan
    Zhang, You
    Jia, Xun
    Wang, Jing
    MEDICAL PHYSICS, 2019, 46 (09) : 4087 - 4094
  • [17] 4D XCAT phantom for multimodality imaging research
    Segars, W. P.
    Sturgeon, G.
    Mendonca, S.
    Grimes, Jason
    Tsui, B. M. W.
    MEDICAL PHYSICS, 2010, 37 (09) : 4902 - 4915
  • [18] Towards clinical implementation of ultrafast combined kV-MV CBCT for IGRT of lung cancer Evaluation of registration accuracy based on phantom study
    Arns, Anna
    Blessing, Manuel
    Fleckenstein, Jens
    Stsepankou, Dzmitry
    Boda-Heggemann, Judit
    Simeonova-Chergou, Anna
    Hesser, Juergen
    Lohr, Frank
    Wenz, Frederik
    Wertz, Hansjoerg
    STRAHLENTHERAPIE UND ONKOLOGIE, 2016, 192 (05) : 312 - 321
  • [19] ADAM phantom to test 4D medical imaging and dose delivery devices
    Pallotta, Stefania
    Calusi, Silvia
    Masi, Laura
    Talamonti, Cinzia
    Marrazzo, Livia
    Foggi, Leonardo
    Casati, Marta
    Livihs, Lorenzo
    Sirnontacchis, Gabriele
    Desideribs, Isacco
    Lisci, Riccardo
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (10)
  • [20] Simulation of spatiotemporal CT data sets using a 4D MRI-based lung motion model
    Marx, Mirko
    Ehrhardt, Jan
    Werner, Rene
    Schlemmer, Heinz-Peter
    Handels, Heinz
    INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2014, 9 (03) : 401 - 409