Improving the efficiency of quantitative 1H NMR: An innovative external standard-internal reference approach

被引:18
|
作者
Huang, Yande [1 ]
Su, Bao-Ning [1 ]
Ye, Qingmei [1 ]
Palaniswamy, Venkatapuram A. [1 ]
Bolgar, Mark S. [1 ]
Raglione, Thomas V. [1 ]
机构
[1] Bristol Myers Squibb Co, Analyt & Bioanalyt Dev, New Brunswick, NJ 08903 USA
关键词
Quantitative NMR; External standard; Internal reference substance; qNMR method; Internal standard; NATURAL-PRODUCTS; NMR-SPECTROSCOPY; ERETIC METHOD; VALIDATION;
D O I
10.1016/j.jpba.2013.07.043
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The classical internal standard quantitative NMR (qNMR) method determines the purity of an analyte by the determination of a solution containing the analyte and a standard. Therefore, the standard must meet the requirements of chemical compatibility and lack of resonance interference with the analyte as well as a known purity. The identification of such a standard can be time consuming and must be repeated for each analyte. In contrast, the external standard qNMR method utilizes a standard with a known purity to calibrate the NMR instrument. The external standard and the analyte are measured separately, thereby eliminating the matter of chemical compatibility and resonance interference between the standard and the analyte. However, the instrumental factors, including the quality of NMR tubes, must be kept the same. Any deviations will compromise the accuracy of the results. An innovative qNMR method reported herein utilizes an internal reference substance along with an external standard to assume the role of the standard used in the traditional internal standard qNMR method. In this new method, the internal reference substance must only be chemically compatible and be free of resonance-interference with the analyte or external standard whereas the external standard must only be of a known purity. The exact purity or concentration of the internal reference substance is not required as long as the same quantity is added to the external standard and the analyte. The new method reduces the burden of searching for an appropriate standard for each analyte significantly. Therefore the efficiency of the qNMR purity assay increases while the precision of the internal standard method is retained. (C) 2013 Elsevier B.V. All rights reserved.
引用
收藏
页码:1 / 6
页数:6
相关论文
共 50 条
  • [11] Quantitative analysis of iobitridol in an injectable preparation by 1H NMR spectroscopy
    Borioni, Anna
    Gostoli, Gianluca
    Bossu, Elena
    Sestili, Isabella
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2014, 94 : 19 - 22
  • [12] Quantification of acesulfame potassium in processed foods by quantitative 1H NMR
    Ohtsuki, Takashi
    Sato, Kyoko
    Abe, Yutaka
    Sugimoto, Naoki
    Akiyama, Hiroshi
    TALANTA, 2015, 131 : 712 - 718
  • [13] VALIDATION OF QUANTITATIVE 1H NMR METHOD FOR THE ANALYSIS OF PHARMACEUTICAL FORMULATIONS
    Santos, Maiara da S.
    Colnago, Luiz Alberto
    QUIMICA NOVA, 2013, 36 (02): : 324 - 330
  • [14] Optimized slice-selective 1H NMR experiments combined with highly accurate quantitative 13C NMR using an internal reference method
    Jezequel, Tangi
    Silvestre, Virginie
    Dinis, Katy
    Giraudeau, Patrick
    Akoka, Serge
    JOURNAL OF MAGNETIC RESONANCE, 2018, 289 : 18 - 25
  • [15] Determination of the iodine value of biodiesel using 1H NMR with 1,4-dioxane as an internal standard
    Oromi-Farrus, Mireia
    Villorbina, Gemma
    Eras, Jordi
    Gatius, Ferran
    Torres, Merce
    Canela, Ramon
    FUEL, 2010, 89 (11) : 3489 - 3492
  • [16] The application of absolute quantitative 1H NMR spectroscopy in drug discovery and development
    Singh, Suruchi
    Roy, Raja
    EXPERT OPINION ON DRUG DISCOVERY, 2016, 11 (07) : 695 - 706
  • [17] Evaluation for Standardization of 1H Quantitative NMR Spectroscopy by Round Robin Test
    Miura, Toru
    Sugimoto, Naoki
    Watanabe, Ryuichi
    Suematsu, Takako
    Takayanagi, Yoichiro
    Ito, Yoshinobu
    Saito, Naoki
    Sawa, Ryuichi
    Kato, Tsuyoshi
    Fujimine, Yoshinori
    Koike, Ryo
    Ohfuku, Yuko
    Yamada, Yuko
    Utsumi, Hiroaki
    Suzuki, Toshiyuki
    YAKUGAKU ZASSHI-JOURNAL OF THE PHARMACEUTICAL SOCIETY OF JAPAN, 2017, 137 (12): : 1543 - 1553
  • [18] Application of the Multi-standard Methodology for Calculating 1H NMR Chemical Shifts
    Sarotti, Ariel M.
    Pellegrinet, Silvina C.
    JOURNAL OF ORGANIC CHEMISTRY, 2012, 77 (14) : 6059 - 6065
  • [19] Quantitative determination of azathioprine in tablets by 1H NMR spectroscopy
    Göger, NG
    Parlatan, HK
    Basan, H
    Berkkan, A
    Özden, T
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1999, 21 (03) : 685 - 689
  • [20] Quantitative Analysis of Porous Silicon Nanoparticles Functionalization by 1H NMR
    Cheng, Ruoyu
    Wang, Shiqi
    Moslova, Karina
    Maekilae, Ermei
    Salonen, Jarno
    Li, Jiachen
    Hirvonen, Jouni
    Xia, Bing
    Santos, Helder A.
    ACS BIOMATERIALS SCIENCE & ENGINEERING, 2022, 8 (10) : 4132 - 4139