Application of the Electronic Nose Technique to Differentiation between Model Mixtures with COPD Markers

被引:25
作者
Dymerski, Tomasz [1 ]
Gebicki, Jacek [2 ]
Wisniewska, Paulina [1 ]
Sliwinska, Magdalena [1 ]
Wardencki, Waldemar [1 ]
Namiesnik, Jacek [1 ]
机构
[1] Gdansk Univ Technol, Fac Chem, Dept Analyt Chem, PL-80233 Gdansk, Poland
[2] Gdansk Univ Technol, Fac Chem, Dept Chem & Proc Engn, PL-80233 Gdansk, Poland
关键词
electronic nose; VOC; MOS; PCA; COPD; VOLATILE ORGANIC-COMPOUNDS; LUNG-CANCER; EXHALED BREATH; DISCRIMINATION; SYSTEM; GAS;
D O I
10.3390/s130405008
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The paper presents the potential of an electronic nose technique in the field of fast diagnostics of patients suspected of Chronic Obstructive Pulmonary Disease (COPD). The investigations were performed using a simple electronic nose prototype equipped with a set of six semiconductor sensors manufactured by FIGARO Co. They were aimed at verification of a possibility of differentiation between model reference mixtures with potential COPD markers (N,N-dimethylformamide and N,N-dimethylacetamide). These mixtures contained volatile organic compounds (VOCs) such as acetone, isoprene, carbon disulphide, propan-2-ol, formamide, benzene, toluene, acetonitrile, acetic acid, dimethyl ether, dimethyl sulphide, acrolein, furan, propanol and pyridine, recognized as the components of exhaled air. The model reference mixtures were prepared at three concentration levels-10 ppb, 25 ppb, 50 ppb v/v-of each component, except for the COPD markers. Concentration of the COPD markers in the mixtures was from 0 ppb to 100 ppb v/v. Interpretation of the obtained data employed principal component analysis (PCA). The investigations revealed the usefulness of the electronic device only in the case when the concentration of the COPD markers was twice as high as the concentration of the remaining components of the mixture and for a limited number of basic mixture components.
引用
收藏
页码:5008 / 5027
页数:20
相关论文
共 45 条
[1]  
Amico A.D., 1999, SENS SER, V22C2, P747
[2]   The electronic nose applied to dairy products: a review [J].
Ampuero, S ;
Bosset, JO .
SENSORS AND ACTUATORS B-CHEMICAL, 2003, 94 (01) :1-12
[3]   Differentiation between cerebrospinal fluid and serum with electronic nose [J].
Aronzon, A ;
Hanson, CW ;
Thaler, ER .
OTOLARYNGOLOGY-HEAD AND NECK SURGERY, 2005, 133 (01) :16-19
[4]  
Arshak K., 2004, Sensor Review, V24, P181, DOI 10.1108/02602280410525977
[5]   Noninvasive detection of lung cancer by analysis of exhaled breath [J].
Bajtarevic, Amel ;
Ager, Clemens ;
Pienz, Martin ;
Klieber, Martin ;
Schwarz, Konrad ;
Ligor, Magdalena ;
Ligor, Tomasz ;
Filipiak, Wojciech ;
Denz, Hubert ;
Fiegl, Michael ;
Hilbe, Wolfgang ;
Weiss, Wolfgang ;
Lukas, Peter ;
Jamnig, Herbert ;
Hackl, Martin ;
Haidenberger, Alfred ;
Buszewski, Boguslaw ;
Miekisch, Wolfram ;
Schubert, Jochen ;
Amann, Anton .
BMC CANCER, 2009, 9 :348
[6]   A sensor array and GC study about VOCs and cancer cells [J].
Bartolazzi, A. ;
Santonico, M. ;
Pennazza, G. ;
Martinelli, E. ;
Paolesse, R. ;
D'Amico, A. ;
Di Natale, C. .
SENSORS AND ACTUATORS B-CHEMICAL, 2010, 146 (02) :483-488
[7]   A preliminary study on the possibility to diagnose urinary tract cancers by an electronic nose [J].
Bernabei, M. ;
Pennazza, G. ;
Santortico, M. ;
Corsi, C. ;
Roscioni, C. ;
Paolesse, R. ;
Di Natale, C. ;
D'Amico, A. .
SENSORS AND ACTUATORS B-CHEMICAL, 2008, 131 (01) :1-4
[8]  
Borjesson T, 1996, CEREAL CHEM, V73, P457
[9]   An electronic nose for multimedia applications [J].
Castro, R ;
Mandal, MK ;
Ajemba, P ;
Istihad, MA .
IEEE TRANSACTIONS ON CONSUMER ELECTRONICS, 2003, 49 (04) :1431-1437
[10]   Two-step preconcentration for analysis of exhaled gas of human breath with electronic nose [J].
Cho, Seong M. ;
Kim, Young Jun ;
Heo, Gwi Suk ;
Shin, Sang-Man .
SENSORS AND ACTUATORS B-CHEMICAL, 2006, 117 (01) :50-57