Evaluation of polycrystalline thallium (I) bromide dosimeter for radiotherapy quality assurance

被引:5
|
作者
Han, Moo-Jae [1 ]
Shin, Yo-Han [1 ]
Yang, Seung-Woo [1 ]
Oh, Kyung-Min [2 ]
Cho, Heung-Lae [3 ]
Park, Sung-Kwang [3 ]
机构
[1] Inje Univ, Coll Med, Dept Radiat Oncol, Bokji Ro 75, Busan, South Korea
[2] Korea Atom Energy Res Inst, Radiat Equipment Res Div, Geumgu Gil 29, Jeongup, South Korea
[3] Inje Univ, Busan Paik Hosp, Dept Radiat Oncol, Bokji Ro 75, Busan, South Korea
关键词
Radiotherapy concepts; Dosimetry concepts and apparatus; Radiation damage to detector materials (solid state);
D O I
10.1088/1748-0221/15/06/P06014
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
In radiotherapy, dose assessment is performed as part of quality assurance (QA) to verify the dose accuracy of the treatment administered to patients. Various dosimeters are used depending on the part of the body and the desired purpose, and tools such as the thermoluminescent dosimeter (TLD) and semiconductor dosimeter are used as in vivo dosimeters. However, TLDs have demonstrated measurement errors of approximately 11.8%, and the diodes used in semiconductor dosimeters fluctuate in performance owing to radiation damage. Consequently, various photoconductor materials are being researched to replace diodes. In particular, Thallium (I) bromide (TlBr) is being used as a substitute material for semiconductors because of its low cost and high performance. Therefore, this study determined the optimised performance of TlBr, which is a potential photoconductor material to replace other materials commercially, using the particle-in-binder deposition method. Upon evaluating the reproducibility and linearity, all sensors demonstrated remarkable performance at 6 MV and 15 MV as the results demonstrated a relative standard deviation (RSD) less than 1.5% and R-2 value above 0.9998, which were the evaluation criteria. A 3:1 material ratio was chosen, as the performance demonstrated in this case was superior to that of other material ratios. The results of the monitor unit rate dependence and percent depth dose (PDD) were compared with those of silicon diode and thimble chambers, which are widely used for dose measurement. According to the monitor unit rate dependence results, the RSD was 0.66% at 6 MV and 0.27% at 15 MV with the standard set at 100 MU, indicating that the diode was outperformed by less than 1%. Similar results were obtained for the PDD with the ion chamber, and the surface dose required for in vivo dose measurement indicated an error within 8.67%. Therefore, this study provides basic data for polycrystalline TlBr dosimeters, which can replace the existing semiconductor dosimeters for QA.
引用
收藏
页数:10
相关论文
共 17 条
  • [1] Development and evaluation of a thallium (I) bromide dosimeter for intracavitary radiotherapy quality assurance
    Han, M-J
    Yang, S-W
    Jung, J-H
    Lee, D-H
    Kim, J-Y
    Cho, S-J
    Kim, K-H
    Mun, C-W
    Cho, H-L
    Park, S-K
    JOURNAL OF INSTRUMENTATION, 2022, 17 (02)
  • [2] Evaluation of two-dimensional dosimeter based on thallium (I) bromide for measurement of surface dose in radiotherapy
    Han, M. J.
    Yang, S. W.
    Lee, D. H.
    Kim, J. Y.
    Mun, C. W.
    Park, S. G.
    Cho, H. L.
    Park, S. K.
    JOURNAL OF INSTRUMENTATION, 2021, 16 (09)
  • [3] Evaluation of mercury (II) iodide dosimeter for digital quality assurance in intracavitary radiotherapy
    Jung, Jae-Hoon
    Han, Moo-Jae
    Yang, Seung-Woo
    Shin, Yong-Il
    Park, Sung-Kwang
    JOURNAL OF INSTRUMENTATION, 2023, 18 (01)
  • [4] Feasibility study of a photoconductor based dosimeter for quality assurance in radiotherapy
    Lee, Y. K.
    Kim, S. W.
    Kim, J. N.
    Kang, Y. N.
    Kim, J. Y.
    Lee, D. S.
    Kim, K. T.
    Han, M. J.
    Ahn, K. J.
    Park, S. K.
    JOURNAL OF INSTRUMENTATION, 2017, 12
  • [5] Development and evaluation of lead(II) iodide-based polycrystalline digital dosimeter for quality assurance in brachytherapy
    Yang, S. W.
    Han, M. J.
    Park, S. K.
    JOURNAL OF INSTRUMENTATION, 2021, 16 (11):
  • [6] Feasibility study of a lead monoxide-based dosimeter for quality assurance in radiotherapy
    Kim, K. T.
    Han, M. J.
    Heo, Y. J.
    Park, J. E.
    Lee, Y. K.
    Kim, J. N.
    Oh, K. M.
    Cho, H. L.
    Choi, Y. S.
    Kim, J. Y.
    Nam, S. H.
    Park, S. K.
    JOURNAL OF INSTRUMENTATION, 2016, 11
  • [7] Development of quality assurance dosimeter for brachytherapy based on CsPbBr3 and evaluation of applicability
    Yang, S. W.
    Han, M. . J.
    Shin, Y. H.
    Jung, J. H.
    Cho, H. . L.
    Park, S. K.
    JOURNAL OF INSTRUMENTATION, 2023, 18 (06)
  • [8] Performance evaluation of lead-monoxide dosimeter with parylene coating for quality assurance of brachytherapy devices
    Yang, S. W.
    Han, M. J.
    Jung, J. H.
    Mun, C. W.
    Cho, H. L.
    Park, S. K.
    JOURNAL OF INSTRUMENTATION, 2021, 16 (11)
  • [9] Development and evaluation of multi-energy PbO dosimeter for quality assurance of image-guide radiation therapy devices
    Kim, Kyo-Tae
    Heo, Ye-Ji
    Han, Moo-Jae
    Oh, Kyung-Min
    Lee, Young-Kyu
    Kim, Shin-Wook
    Park, Sung-Kwang
    JOURNAL OF INSTRUMENTATION, 2017, 12
  • [10] Development and evaluation of polycrystalline cadmium telluride dosimeters for accurate quality assurance in radiation therapy
    Oh, K.
    Han, M.
    Kim, K.
    Heo, Y.
    Moon, C.
    Park, S.
    Nam, S.
    JOURNAL OF INSTRUMENTATION, 2016, 11