Treatment vault shielding for a flattening filter-free medical linear accelerator

被引:29
|
作者
Kry, Stephen F. [1 ]
Howell, Rebecca M. [1 ]
Polf, Jerimy [1 ]
Mohan, Radhe [1 ]
Vassiliev, Oleg N. [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2009年 / 54卷 / 05期
关键词
UNFLATTENED PHOTON BEAMS; DOSIMETRIC PROPERTIES; MONTE-CARLO;
D O I
10.1088/0031-9155/54/5/011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The requirements for shielding a treatment vault with a Varian Clinac 2100 medical linear accelerator operated both with and without the flattening filter were assessed. Basic shielding parameters, such as primary beam tenth-value layers (TVLs), patient scatter fractions, and wall scatter fractions, were calculated using Monte Carlo simulations of 6, 10 and 18 MV beams. Relative integral target current requirements were determined from treatment planning studies of several disease sites with, and without, the flattening filter. The flattened beam shielding data were compared to data published in NCRP Report No. 151, and the unflattened beam shielding data were presented relative to the NCRP data. Finally, the shielding requirements for a typical treatment vault were determined for a single-energy ( 6 MV) linac and a dual-energy ( 6 MV/18 MV) linac. With the exception of large-angle patient scatter fractions and wall scatter fractions, the vault shielding parameters were reduced when the flattening filter was removed. Much of this reduction was consistent with the reduced average energy of the FFF beams. Primary beam TVLs were reduced by 12%, on average, and small-angle scatter fractions were reduced by up to 30%. Head leakage was markedly reduced because less integral target current was required to deliver the target dose. For the treatment vault examined in the current study, removal of the flattening filter reduced the required thickness of the primary and secondary barriers by 10-20%, corresponding to 18 m(3) less concrete to shield the single-energy linac and 36 m3 less concrete to shield the dual-energy linac. Thus, a shielding advantage was found when the linac was operated without the flattening filter. This translates into a reduction in occupational exposure and/or the cost and space of shielding.
引用
收藏
页码:1265 / 1273
页数:9
相关论文
共 33 条
  • [11] Evaluation of beam modeling for small fields using a flattening filter-free beam
    Kawahara D.
    Ozawa S.
    Nakashima T.
    Aita M.
    Tsuda S.
    Ochi Y.
    Okumura T.
    Masuda H.
    Ohno Y.
    Murakami Y.
    Nagata Y.
    Radiological Physics and Technology, 2017, 10 (1) : 33 - 40
  • [12] DNA strand breaks based on Monte Carlo simulation in and around the Lipiodol with flattening filter and flattening filter-free photon beams
    Kawahara, Daisuke
    Saito, Akito
    Nakano, Hisashi
    Nagata, Yasushi
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2022, 27 (03) : 392 - 400
  • [13] Flattening filter-free accelerators: a report from the AAPM Therapy Emerging Technology Assessment Work Group
    Xiao, Ying
    Kry, Stephen F.
    Popple, Richard
    Yorke, Ellen
    Papanikolaou, Niko
    Stathakis, Sotirios
    Xia, Ping
    Huq, Saiful
    Bayouth, John
    Galvin, James
    Yin, Fang-Fang
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2015, 16 (03): : 12 - 29
  • [14] Stereotactic body radiation therapy (SBRT) for lung malignancies: preliminary toxicity results using a flattening filter-free linear accelerator operating at 2400 monitor units per minute
    Prendergast, Brendan M.
    Dobelbower, Michael C.
    Bonner, James A.
    Popple, Richard A.
    Baden, Craig J.
    Minnich, Douglas J.
    Cerfolio, Robert J.
    Spencer, Sharon A.
    Fiveash, John B.
    RADIATION ONCOLOGY, 2013, 8
  • [15] PERIPHERAL SURFACE DOSE FROM A LINEAR ACCELERATOR: RADIOCHROMIC FILM EXPERIMENTAL MEASUREMENTS OF FLATTENING FILTER FREE VERSUS FLATTENED BEAMS
    Garcia-Hernandez, Trinitat
    Vicedo-Gonzalez, Aurora
    Sanchez-Nieto, Beatriz
    Romero-Exposito, Maite
    Rosello-Ferrando, Joan
    RADIATION PROTECTION DOSIMETRY, 2020, 188 (03) : 285 - 298
  • [16] A 4 MV flattening filter-free beam: commissioning and application to conformal therapy and volumetric modulated arc therapy
    Stevens, S. W.
    Rosser, K. E.
    Bedford, J. L.
    PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (13): : 3809 - 3824
  • [17] Small-field Dosimetry of 6-and 10-MV Flattening Filter-free and Flattening Filter Photon Beams for Therapeutic Use
    Shukla, Jooli
    Dwivedi, Shekhar
    Kansal, Sandeep
    Dangwal, Vinod Kumar
    Tiwari, K. P.
    Singla, Amit Kumar
    Rani, Supriya
    INDIAN JOURNAL OF PURE & APPLIED PHYSICS, 2023, 61 (10) : 846 - 850
  • [18] Dosimetry of a 6 MV flattening filter-free small photon beam using various detectors
    Dwivedi, Shekhar
    Kansal, Sandeep
    Dangwal, Vinod Kumar
    Bharati, Avinav
    Shukla, Jooli
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2021, 7 (04):
  • [19] Commissioning of Elekta InfinityTM 6 MV flattening filter-free using Monte Carlo simulation
    Maulana, Iswan Fadhil
    Yani, Sitti
    Sumaryada, Tony
    Rhani, Mohamad Fahdillah
    Haryanto, Freddy
    RADIATION PHYSICS AND CHEMISTRY, 2023, 210
  • [20] Flattening Filter-Free Beams in Intensity-Modulated Radiotherapy and Volumetric Modulated Arc Therapy for Sinonasal Cancer
    Lu, Jia-Yang
    Zheng, Jing
    Zhang, Wu-Zhe
    Huang, Bao-Tian
    PLOS ONE, 2016, 11 (01):