A lumped-parameter model for kidney pressure during stone removal

被引:2
|
作者
Williams, J. G. [1 ]
Rouse, L. [2 ]
Turney, B. W. [3 ]
Waters, S. L. [1 ]
Moulton, D. E. [1 ]
机构
[1] Math Inst, Andrew Wiles Bldg,Woodstock Rd, Oxford OX2 6GG, England
[2] Urol Associates, Forte Hlth Bldg,132 Peterborough St, Christchurch 8013, New Zealand
[3] Univ Oxford, John Radcliffe Hosp, Nuffield Dept Surg Sci, Oxford OX3 9DU, England
基金
英国工程与自然科学研究理事会;
关键词
urology; pipe network modelling; ex vivo experiments; kidney compliance;
D O I
10.1093/imamat/hxaa020
中图分类号
O29 [应用数学];
学科分类号
070104 ;
摘要
In this paper, we consider a lumped-parameter model to predict renal pressures and flow rate during a minimally invasive surgery for kidney stone removal, ureterorenoscopy. A ureteroscope is an endoscope designed to work within the ureter and the kidney and consists of a long shaft containing a narrow, cylindrical pipe, called the working channel. Fluid flows through the working channel into the kidney. A second pipe, the 'access sheath', surrounds the shaft of the scope, allowing fluid to flow back out of the urinary system. We modify and extend a previously developed model ( ) through the use of an exponential, instead of linear, constitutive law for kidney compliance and by exploring the effects of variable flow resistance, dependent on the presence of auxiliary 'working tools' in the working channel and the cross-sectional shapes of the tools, working channel, scope shaft and access sheath. We motivate the chosen function for kidney compliance and validate the model predictions, with ex vivo experimental data. Although the predicted and measured flow rates agree, we find some disagreement between theory and experiment for kidney pressure. We hypothesize that this is caused by spatial pressure variation in the renal pelvis, i.e. unaccounted for in the lumped-parameter model. We support this hypothesis through numerical simulations of the steady Navier-Stokes equations in a simplified geometry. We also determine the optimal cross-sectional shapes for the scope and access sheath (for fixed areas) to minimize kidney pressure and maximize flow rate.
引用
收藏
页码:703 / 723
页数:21
相关论文
共 50 条
  • [1] A LUMPED-PARAMETER MODEL FOR INFILTRATION
    PINGOUD, K
    JOURNAL OF HYDROLOGY, 1982, 57 (1-2) : 175 - 185
  • [2] SENSITIVITY ANALYSIS OF A LUMPED-PARAMETER MODEL FOR INFILTRATION
    PINGOUD, K
    JOURNAL OF HYDROLOGY, 1984, 67 (1-4) : 97 - 113
  • [3] LUMPED-PARAMETER MODEL FOR INVIVO COCHLEAR STIMULATION
    SUESSERMAN, MF
    SPELMAN, FA
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1993, 40 (03) : 237 - 245
  • [4] Lumped-parameter thermal model for induction machines
    Durán, MJ
    Fernández, J
    IEEE TRANSACTIONS ON ENERGY CONVERSION, 2004, 19 (04) : 791 - 792
  • [5] Lumped-parameter inertia model for flexure hinges
    Lobontiu, N
    Garcia, E
    MECHANICS BASED DESIGN OF STRUCTURES AND MACHINES, 2004, 32 (01) : 73 - 99
  • [6] PARAMETER SENSITIVITY ANALYSIS OF A LUMPED-PARAMETER MODEL OF LYMPHANGIONS IN SERIES
    Jamalian, Samira
    Moore, James E., Jr.
    Bertram, Christopher D.
    Richardson, Will
    PROCEEDINGS OF THE ASME SUMMER BIOENGINEERING CONFERENCE, PTS A AND B, 2012, : 345 - 346
  • [7] LUMPED-PARAMETER MODEL FOR HEMODIALYZER WITH APPLICATION TO SIMULATION OF PATIENT-ARTIFICIAL-KIDNEY SYSTEM
    RAMACHANDRAN, PA
    MASHELKAR, RA
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1980, 18 (02) : 179 - 188
  • [8] Lumped-parameter model for dynamic monolayer graphene sheets
    Wei, Dongming
    Nurakhmetov, Daulet
    Aniyarov, Almir
    Zhang, Dichuan
    Spitas, Christos
    JOURNAL OF SOUND AND VIBRATION, 2022, 534
  • [9] Study on Lumped-parameter Model of Surface Circular Foundation
    Wang, Hai
    Liu, Weiqing
    Wang, Shuguang
    Du, Dongsheng
    Zhou, Ding
    ADVANCES IN BUILDING MATERIALS, PTS 1-3, 2011, 261-263 : 980 - 984
  • [10] A female-specific cardiovascular lumped-parameter model
    Comunale, Giulia
    Susin, Francesca M.
    Mynard, Jonathan P.
    42ND ANNUAL INTERNATIONAL CONFERENCES OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY: ENABLING INNOVATIVE TECHNOLOGIES FOR GLOBAL HEALTHCARE EMBC'20, 2020, : 2654 - 2657