Comparison between a flash glucose monitoring system and a portable blood glucose meter for monitoring dogs with diabetes mellitus

被引:29
|
作者
Del Baldo, Francesca [1 ]
Canton, Claudia [1 ]
Testa, Silvia [1 ]
Swales, Harry [2 ]
Drudi, Ignazio [3 ]
Golinelli, Stefania [1 ]
Fracassi, Federico [1 ]
机构
[1] Univ Bologna, Dept Vet Med Sci, Via Tolara di Sopra 50, I-40064 Bologna, Italy
[2] Univ Liverpool, Small Anim Teaching Hosp, Leahurst Campus, Wirral, Merseyside, England
[3] Univ Bologna, Dept Stat Sci, Bologna, Italy
关键词
canine; diabetes mellitus; FreeStyle Libre; interstitial glucose; TO-DAY VARIATION; GLYCEMIC VARIABILITY; INTERSTITIAL GLUCOSE; SENSING TECHNOLOGY; GROWTH-HORMONE; REAL-LIFE; INSULIN; TYPE-1; HYPOGLYCEMIA; MULTICENTER;
D O I
10.1111/jvim.15930
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Background Flash glucose monitoring system (FGMS; FreeStyle Libre) was recently validated for use in diabetic dogs (DD). It is not known if this system is clinically useful in monitoring DD. Objective To compare the clinical utility of FGMS against blood glucose curves (BGCs) obtained with a portable blood glucose meter (PBGM) in monitoring DD. Animals Twenty dogs with diabetes mellitus. Methods Prospective study. Dogs with diabetes mellitus on insulin treatment for at least 1 month were included. Comparisons of insulin dose recommendations based on the in-hospital GCs acquired using FGMS and a PBGM, consecutive-day interstitial GCs (IGCs) acquired at home using the FGMS, and consecutive-day, home vs hospital IGCs acquired using the FGMS were made using concordance analysis. Results There was good concordance between insulin dose recommendations based on FGMS and PBGM generated GCs and IGCs obtained in the 2 different environments on 2 consecutive days, but almost absent concordance between IGCs obtained on 2 consecutive days at home. Glucose nadirs were detected in 34/43 (79%) of Ambulatory Glucose Profile (AGP) reports of the FGMS. In comparison, concordant glucose nadirs were identified in 14/34 (41%) BGCs using PBGM. The individual FGMS scans and PBGM identified 60% and 9% of low IG/hypoglycemic episodes, respectively. Conclusions and Clinical Importance Insulin dose adjustments based on BGCs can be suboptimal. The FGMS allows a more accurate identification of the glucose nadirs and hypoglycemic episodes compared to the use of a PBGM and assessment of day-to-day variations in glycemic control.
引用
收藏
页码:2296 / 2305
页数:10
相关论文
共 50 条
  • [31] Effects of flash glucose monitoring on glycemic control in participants with diabetes mellitus: A meta-analysis of randomized controlled trials
    Gao, Yuan
    Zhou, Min
    Xu, Xuan
    Chen, Wen Yu
    JOURNAL OF DIABETES AND ITS COMPLICATIONS, 2022, 36 (11)
  • [32] A review of flash glucose monitoring in type 2 diabetes
    Krakauer, Marcio
    Botero, Jose Fernando
    Lavalle-Gonzalez, Fernando J.
    Proietti, Adrian
    Barbieri, Douglas Eugenio
    DIABETOLOGY & METABOLIC SYNDROME, 2021, 13 (01)
  • [33] Association between scanning frequency of flash glucose monitoring and continuous glucose monitoring-derived glycemic makers in children and adolescents with type 1 diabetes
    Suzuki, Junichi
    Urakami, Tatsuhiko
    Yoshida, Kei
    Kuwabara, Remi
    Mine, Yusuke
    Aoki, Masako
    Morioka, Ichiro
    PEDIATRICS INTERNATIONAL, 2021, 63 (02) : 154 - 159
  • [34] HOME BLOOD GLUCOSE MONITORING IN TYPE 1 DIABETES MELLITUS
    Alves Grossi, Sonia Aurora
    Lottenberg, Simao Augusto
    Lottenberg, Ana Maria
    Della Manna, Thais
    Kuperman, Hilton
    REVISTA LATINO-AMERICANA DE ENFERMAGEM, 2009, 17 (02): : 194 - 200
  • [35] Accuracy of a portable glucose meter and of a Continuous Glucose Monitoring device used at home by patients with type 1 diabetes
    Francescato, M. P.
    Geat, M.
    Stel, G.
    Cauci, S.
    CLINICA CHIMICA ACTA, 2012, 413 (1-2) : 312 - 318
  • [36] Flash glucose monitoring and glycemic control in type 1 diabetes with subcutaneous insulin infusion flash glucose monitoring and glycemic control
    Pinheiro, Sara Lomelino
    Bastos, Margarida
    Barros, Luisa
    Melo, Miguel
    Paiva, Isabel
    ACTA DIABETOLOGICA, 2022, 59 (04) : 509 - 515
  • [37] Efficacy of flash glucose monitoring in pregnant women with poorly controlled pregestational diabetes (FlashMom): A randomized pilot
    Tumminia, Andrea
    Milluzzo, Agostino
    Festa, Camilla
    Fresa, Raffaella
    Pintaudi, Basilio
    Scavini, Marina
    Vitacolonna, Ester
    Napoli, Angela
    Sciacca, Laura
    NUTRITION METABOLISM AND CARDIOVASCULAR DISEASES, 2021, 31 (06) : 1851 - 1859
  • [38] Flash glucose monitoring in gestational diabetes mellitus (FLAMINGO): a randomised controlled trial
    Majewska, Agata
    Stanirowski, Pawel Jan
    Tatur, Jacek
    Wojda, Barbara
    Radosz, Iwona
    Wielgos, Miroslaw
    Bomba-Opon, Dorota Agata
    ACTA DIABETOLOGICA, 2023, 60 (09) : 1171 - 1177
  • [39] Ambulatory glucose profile: Flash glucose monitoring
    Kalra, Sanjay
    Gupta, Yashdeep
    JOURNAL OF THE PAKISTAN MEDICAL ASSOCIATION, 2015, 65 (12) : 1360 - 1362
  • [40] The Effect of Novel Glucose Monitoring System (Flash Glucose Monitoring) on Mental Well-being and Treatment Satisfaction in Japanese People with Diabetes