Validation of a Free Phenytoin Assay on Cobas c501 Analyzer Using Calibrators From Cobas Integra Free Phenytoin Assay by Comparing its Performance With Fluorescence Polarization Immunoassay for Free Phenytoin on the TDx Analyzer

被引:2
作者
Dasgupta, Amitava [1 ]
Davis, Bonnette
Chow, Loretta
机构
[1] Univ Texas Houston, Sch Med, Dept Pathol & Lab Med, Houston, TX 77030 USA
关键词
free phenytoin; fluorescence polarization; Cobas c501; PROTEIN-BINDING; PLASMA; SERUM;
D O I
10.1002/jcla.21536
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
For many years, fluorescence polarization immunoassay (FPIA) on the TDx analyzer has been used for determination of free phenytoin concentration. Recently Abbott Laboratories decided to discontinue the TDx analyzer and related assays on this analyzer. Free phenytoin assay is also available from Roche Diagnostics for application on the Cobas Integra analyzer (fluorescence polarization assay) but not on Cobas c510 analyzer. Free phenytoin calibrators from the Cobas Integra free phenytoin assay and the reagents from the KIMSphenytoin assay were used for the determination of free phenytoin on the Cobas c501 analyzer. The intra-run and inter-run precisions were both <7.2%. The assay was linear from 0.2 to 4 g/ml. The free phenytoin assay on the Cobas c501 was compared with the FPIAassay on the TDx analyzer using sera from 25 patients receiving phenytoin (phenytoin concentration between 0.3 and 3.7 g/ml). The following regression equation was observed: y = 0.9899 x + 0.0408 (r = 0.98, n = 25). In conclusion, the free phenytoin assay on the Cobas c501 analyzer is a valid alternative to free phenytoin assay on the TDx analyzer. J. Clin. Lab. Anal. 27:1-4, 2013. (c) 2012 Wiley Periodicals, Inc.
引用
收藏
页码:1 / 4
页数:4
相关论文
共 14 条
[1]  
[Anonymous], 2002, EP9A2 CLSI
[2]   Determination of free levels of phenytoin in human plasma by liquid chromatography/tandem mass spectrometry [J].
Bardin, S ;
Ottinger, JC ;
Breau, AP ;
O'Shea, TJ .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2000, 23 (2-3) :573-579
[3]   SERUM CONCENTRATIONS OF FREE DIPHENYLHYDANTOIN AND THEIR RELATIONSHIP TO CLINICAL INTOXICATION [J].
BOOKER, HE ;
DARCEY, B .
EPILEPSIA, 1973, 14 (02) :177-184
[4]  
Burt M, 2000, CLIN CHEM, V46, P1132
[5]   Plasma total phenytoin: A possibly misleading test in developing countries [J].
Fedler, C ;
Stewart, MJ .
THERAPEUTIC DRUG MONITORING, 1999, 21 (02) :155-160
[6]   Differences between the Measured and Calculated Free Serum Phenytoin Concentrations in Epileptic Patients [J].
Hong, Ji-Man ;
Choi, Young-Chul ;
Kim, Won-Joo .
YONSEI MEDICAL JOURNAL, 2009, 50 (04) :517-520
[7]   Clinical evaluation of plasma free phenytoin measurement and factors influencing its protein binding [J].
Iwamoto, T ;
Kagawa, Y ;
Naito, Y ;
Kuzuhara, S ;
Okuda, M .
BIOPHARMACEUTICS & DRUG DISPOSITION, 2006, 27 (02) :77-84
[8]   PLASMA-CONCENTRATIONS OF UNBOUND PHENYTOIN IN THE MANAGEMENT OF EPILEPSY [J].
KILPATRICK, CJ ;
WANWIMOLRUK, S ;
WING, LMH .
BRITISH JOURNAL OF CLINICAL PHARMACOLOGY, 1984, 17 (05) :539-546
[9]   PHENYTOIN TOXICITY ASSOCIATED WITH HYPOALBUMINEMIA IN CRITICALLY ILL PATIENTS [J].
LINDOW, J ;
WIJDICKS, EFM .
CHEST, 1994, 105 (02) :602-604
[10]  
MCNAMARA PJ, 1981, J LAB CLIN MED, V98, P730