Direct Detection of Acetonitrile in Human Breath by Extractive Electrospray Ionization Mass Spectrometry

被引:4
作者
Zhang Yan [1 ]
Pan Su-Su [1 ]
Zhu Zhi-Qiang [1 ]
Zhang Xing-Lei [1 ]
Xu Gao-Si [2 ]
Wei Yi-Ping [2 ]
Chen Huan-Wen [1 ]
Ding Jian-Hua [1 ]
机构
[1] E China Inst Technol, Jiangxi Key Lab Mass Spectrometry & Instrumentat, Nanchang 330013, Peoples R China
[2] Nanchang Univ, Affiliated Hosp 2, Nanchang 330006, Peoples R China
关键词
Extractive electrospray ionization-mass spectrometry; Acetonitrile; Exhaled breath; VOLATILE ORGANIC-COMPOUNDS; EXHALED BREATH; PTR-MS; QUANTIFICATION; NONSMOKERS; SMOKERS; PPTV;
D O I
10.3724/SP.J.1096.2013.20989
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Breath analysis by virtue of its potential medical research value, and as a non-invasive sample, simple sampling and continuous available is of interest for modern clinical diagnosis and metabonomics. The concentration levels of acetonitrile, an important biomarker in human exhaled gas, are correlated with human health. Herein we constructed a novel platform of extractive electrospray ionization-mass spectrometry (EESI-MS) for direct, rapid detection of acetonitrile in human breath without any sample pretreatment. Under the optimized working conditions, the quantitative information of acetonitrile in human breath was acquired effectively. For acetonitrile, a limit of detection was found to be 5. 71 pL/L; the linear range was from 76. 4 pL/L to 1910 pL/L. The concentration of acetonitrile in the exhaled breath of two healthy volunteers was in vivo determined to be 134. 49 and 104. 40 pL/L, respectively, the relative standard deviations were 6. 3% and 7. 2%. The results show that EESI-Ion Trap MS is a powerful analytical platform to provide high sensitivity, high precision, rapid analysis and high specificity for semi-quantitative analysis of acetonitrile in human breath, and for detection of other bio-molecules in exhaled breath potentially.
引用
收藏
页码:1220 / 1225
页数:6
相关论文
共 25 条
[1]   Quantification of acetonitrile in exhaled breath and urinary headspace using selected ion flow tube mass spectrometry [J].
Abbott, SM ;
Elder, JB ;
Spanel, P ;
Smith, D .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2003, 228 (2-3) :655-665
[2]   Review article: breath testing for human liver function assessment [J].
Armuzzi, A ;
Candelli, M ;
Zocco, MA ;
Andreoli, A ;
De Lorenzo, A ;
Nista, EC ;
Miele, L ;
Cremonini, F ;
Cazzato, IA ;
Grieco, A ;
Gasbarrini, G ;
Gasbarrini, A .
ALIMENTARY PHARMACOLOGY & THERAPEUTICS, 2002, 16 (12) :1977-1996
[3]   Noninvasive detection of lung cancer by analysis of exhaled breath [J].
Bajtarevic, Amel ;
Ager, Clemens ;
Pienz, Martin ;
Klieber, Martin ;
Schwarz, Konrad ;
Ligor, Magdalena ;
Ligor, Tomasz ;
Filipiak, Wojciech ;
Denz, Hubert ;
Fiegl, Michael ;
Hilbe, Wolfgang ;
Weiss, Wolfgang ;
Lukas, Peter ;
Jamnig, Herbert ;
Hackl, Martin ;
Haidenberger, Alfred ;
Buszewski, Boguslaw ;
Miekisch, Wolfram ;
Schubert, Jochen ;
Amann, Anton .
BMC CANCER, 2009, 9 :348
[4]   Amphetamines Detected in Exhaled Breath from Drug Addicts: A New Possible Method for Drugs-of-Abuse Testing [J].
Beck, Olof ;
Leine, Kathinka ;
Palmskog, Goran ;
Franck, Johan .
JOURNAL OF ANALYTICAL TOXICOLOGY, 2010, 34 (05) :233-237
[5]   HENRYS LAW COEFFICIENTS FOR AQUEOUS-SOLUTIONS OF ACETONE, ACETALDEHYDE AND ACETONITRILE, EQUILIBRIUM-CONSTANTS FOR THE ADDITION-COMPOUNDS OF ACETONE AND ACETALDEHYDE WITH BISULFITE [J].
BENKELBERG, HJ ;
HAMM, S ;
WARNECK, P .
JOURNAL OF ATMOSPHERIC CHEMISTRY, 1995, 20 (01) :17-34
[6]   Analysis of exhaled breath from smokers, passive smokers and non-smokers by solid-phase microextraction gas chromatography/mass spectrometry [J].
Buszewski, Boguslaw ;
Ulanowska, Agnieszka ;
Ligor, Tomasz ;
Denderz, Natalia ;
Amann, Anton .
BIOMEDICAL CHROMATOGRAPHY, 2009, 23 (05) :551-556
[7]  
Chen H., 2007, Angewandte Chemie, V119, P586, DOI [10.1002/ange.200602942, DOI 10.1002/ANGE.200602942]
[8]   Fundamental Principles and Practical Applications of Ambient Ionization Mass Spectrometry for Direct Analysis of Complex Samples [J].
Chen Huan-Wen ;
Hu Bin ;
Zhang Xie .
CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2010, 38 (08) :1069-1088
[9]   Selective Detection of Diethylene Glycol in Toothpaste Products Using Neutral Desorption Reactive Extractive Electrospray Ionization Tandem Mass Spectrometry [J].
Ding, Jianhua ;
Gu, Haiwei ;
Yang, Shuiping ;
Li, Ming ;
Li, Jianqiang ;
Chen, Huanwen .
ANALYTICAL CHEMISTRY, 2009, 81 (20) :8632-8638
[10]   Development of extractive electrospray ionization ion trap mass spectrometry for in vivo breath analysis [J].
Ding, Jianhua ;
Yang, Shuiping ;
Liang, Dapeng ;
Chen, Huanwen ;
Wu, Zhuanzhang ;
Zhang, Lili ;
Ren, Yulin .
ANALYST, 2009, 134 (10) :2040-2050