Dynamic Evaluation of a Digital Wireless Intracranial Pressure Sensor for the Assessment of Traumatic Brain Injury in a Swine Model

被引:30
作者
Meng, Xu [1 ]
Browne, Kevin D. [2 ]
Huang, Shi-Min [3 ]
Mietus, Constance [2 ]
Cullen, D. Kacy [2 ]
Tofighi, Mohammad-Reza [4 ]
Rosen, Arye [1 ]
机构
[1] Drexel Univ, Sch Biomed Engn, Philadelphia, PA 19104 USA
[2] Univ Penn, Dept Neurosurg, Ctr Brain Injury & Repair, Philadelphia, PA 19104 USA
[3] Drexel Univ, Dept Elect Engn, Philadelphia, PA 19104 USA
[4] Penn State Univ, Capital Coll, Middletown, PA 17057 USA
基金
美国国家卫生研究院;
关键词
Digital wireless implants; intracranial pressure (ICP); rotational head injury; slot antenna; traumatic brain injury (TBI); CLINICAL-PRACTICE; HEAD-INJURY; DEVICE; TRANSMISSION; IMPLANT; PIG;
D O I
10.1109/TMTT.2012.2224361
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The monitoring and control of increased intracranial pressure (ICP) are the major requirements for the diagnosis and treatment of patients suffering from traumatic brain injury (TBI) or hydrocephalus. Widely used commercially available ICP measuring devices require the use of a tethered fiber optic probe, which can cause complications, such as infection and hemorrhage. The use of a probe also prohibits immediate (at the time of induced injury) and long-term measurements of ICP. Therefore, we have developed and tested a small fully embedded wireless ICP sensor, incorporating a novel antenna and packaging arrangement. Evaluations were performed in-vitro and in-vivo to demonstrate the robustness of this microwave pressure-monitoring system. This is the first report of in-vivo tests in a dynamic study of continuous (every 6 s) wireless ICP measurements in a swine model of closed-head rotational-acceleration induced TBI using a digital device. In particular, our device measured extreme changes in ICP immediately upon rapid head rotation that persisted for hours after the injury. The readings matched well those obtained from a commercially available tethered monitor. Our device can be utilized in the future as a tool to diagnose and track long-term ICP changes.
引用
收藏
页码:316 / 325
页数:10
相关论文
共 29 条
[1]  
[Anonymous], 2009, HYDR STAT
[2]  
[Anonymous], 2011, P 2011 INT S IM DAT
[3]   Wireless Implants [J].
Bashirullah, Rizwan .
IEEE MICROWAVE MAGAZINE, 2010, 11 (07) :S14-S23
[4]   A CLINICAL-EVALUATION OF THE CAMINO SUBDURAL SCREW AND VENTRICULAR MONITORING KITS [J].
CHAMBERS, IR ;
MENDELOW, AD ;
SINAR, EJ ;
MODHA, P .
NEUROSURGERY, 1990, 26 (03) :421-423
[5]  
CHAMBERS IR, 1993, NEUROSURGERY, V33, P866
[6]   Wireless Intraocular Pressure Sensing Using Microfabricated Minimally Invasive Flexible-Coiled LC Sensor Implant [J].
Chen, Po-Jui ;
Saati, Saloomeh ;
Varma, Rohit ;
Humayun, Mark S. ;
Tai, Yu-Chong .
JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2010, 19 (04) :721-734
[7]   Fully Wireless Implantable Cardiovascular Pressure Monitor Integrated with a Medical Stent [J].
Chow, Eric Y. ;
Chlebowski, Arthur L. ;
Chakraborty, Sudipto ;
Chappell, William J. ;
Irazoqui, Pedro P. .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2010, 57 (06) :1487-1496
[8]   Implantable Wireless Telemetry Boards for In Vivo Transocular Transmission [J].
Chow, Eric Y. ;
Yang, Chin-Lung ;
Chlebowski, Arthur ;
Moon, Sungwook ;
Chappell, William J. ;
Irazoqui, Pedro P. .
IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2008, 56 (12) :3200-3208
[9]  
Faul MX., 2010, Centers for Disease Control and Prevention, National Center for Injury Prevention and Control
[10]   Transcranial Doppler pulsatility index is not a reliable indicator of intracranial pressure in children with severe traumatic brain injury [J].
Figaji, Anthony A. ;
Zwane, Eugene ;
Fieggen, A. Graham ;
Slesjo, Peter ;
Peter, Jonathan C. .
SURGICAL NEUROLOGY, 2009, 72 (04) :389-394