Separate density and viscosity measurements of unknown liquid using quartz crystal microbalance

被引:24
作者
Tan, Feng [1 ]
Qiu, Du-Yu [1 ]
Guo, Lian-Ping [1 ]
Ye, Peng [1 ]
Zeng, Hao [1 ]
Jiang, Jun [1 ]
Tang, Yong [2 ]
Zhang, Yi-Cheng [2 ]
机构
[1] Univ Elect Sci & Technol China, Sch Automat Engn, 2006 Xiyuan Ave,West Hitech Zone, Chengdu 611731, Sichuan, Peoples R China
[2] Univ Fribourg, Dept Phys, Chemin Musee 3, CH-1700 Fribourg, Switzerland
关键词
FREQUENCY;
D O I
10.1063/1.4963298
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
Aqueous liquids have a wide range of applications in many fields. Basic physical properties like the density and the viscosity have great impacts on the functionalities of a given ionic liquid. For the millions kinds of existing liquids, only a few have been systematically measured with the density and the viscosity using traditional methods. However, these methods are limited to measure the density and the viscosity of an ionic liquid simultaneously especially in processing micro sample volumes. To meet this challenge, we present a new theoretical model and a novel method to separate density and viscosity measurements with single quartz crystal microbalance (QCM) in this work. The agreement of experimental results and theocratical calculations shows that the QCM is capable to measure the density and the viscosity of ionic liquids. (C) 2016 Author(s). All article content, except where otherwise noted, is licensed under a Creative Commons Attribution (CC BY) license
引用
收藏
页数:6
相关论文
共 17 条
[1]   SOME NEW CONCEPTS ON FACTORS INFLUENCING THE OPERATIONAL FREQUENCY OF LIQUID-IMMERSED QUARTZ MICROBALANCES [J].
BARNES, C .
SENSORS AND ACTUATORS A-PHYSICAL, 1992, 30 (03) :197-202
[2]   Using a deoxyribozyme ligase and rolling circle amplification to detect a non-nucleic acid analyte, ATP [J].
Cho, EJ ;
Yang, LT ;
Levy, M ;
Ellington, AD .
JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 2005, 127 (07) :2022-2023
[3]   SIMULTANEOUS THIN-FILM STRESS AND MASS-CHANGE MEASUREMENTS USING QUARTZ RESONATORS [J].
EERNISSE, EP .
JOURNAL OF APPLIED PHYSICS, 1972, 43 (04) :1330-+
[4]   VISCOSITY OF NACL AND KCL SOLUTIONS IN RANGE 25-50-DEGREES-C [J].
GONCALVES, FA ;
KESTIN, J .
BERICHTE DER BUNSEN-GESELLSCHAFT-PHYSICAL CHEMISTRY CHEMICAL PHYSICS, 1977, 81 (11) :1156-1161
[5]   Ultrasensitive quartz crystal microbalance with porous gold electrodes [J].
Hieda, M ;
Garcia, R ;
Dixon, M ;
Daniel, T ;
Allara, D ;
Chan, MHW .
APPLIED PHYSICS LETTERS, 2004, 84 (04) :628-630
[6]   Separate measurement of the density and viscosity of a liquid using a quartz crystal microbalance based on admittance analysis (QCM-A) [J].
Itoh, Atsushi ;
Ichihashi, Motoko .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2011, 22 (01)
[7]   FREQUENCY OF A QUARTZ MICROBALANCE IN CONTACT WITH LIQUID [J].
KANAZAWA, KK ;
GORDON, JG .
ANALYTICAL CHEMISTRY, 1985, 57 (08) :1770-1771
[8]   Real-time detection of airborne viruses on a mass-sensitive device [J].
Lee, Joonhyung ;
Jang, Jaesung ;
Akin, Dernir ;
Savran, Cagri A. ;
Bashir, Rashid .
APPLIED PHYSICS LETTERS, 2008, 93 (01)
[9]  
Martin S. J., 1993, TECH REP
[10]   VISCOSITIES OF AQUEOUS HYDROCHLORIC-ACID SOLUTIONS, AND DENSITIES AND VISCOSITIES OF AQUEOUS HYDROIODIC ACID-SOLUTIONS [J].
NISHIKATA, E ;
ISHII, T ;
OHTA, T .
JOURNAL OF CHEMICAL AND ENGINEERING DATA, 1981, 26 (03) :254-256