Determination of cadmium in sediment by isotope dilution inductively coupled plasma mass spectrometry using a co-precipitation separation technique

被引:28
作者
Inagaki, K [1 ]
Takatsu, A [1 ]
Uchiumi, A [1 ]
Nakama, A [1 ]
Okamoto, K [1 ]
机构
[1] Natl Inst Adv Ind Sci & Technol, Natl Metrol Inst Japan, Environm Stand Sect, Tsukuba, Ibaraki 3058563, Japan
关键词
D O I
10.1039/b106974h
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A co-precipitation separation method is described for the determination of cadmium in sediment by isotope dilution inductively coupled plasma mass spectrometry. A sediment sample is digested with HNO3, HClO4 and HF, and then ammonia solution is added. Zirconium and Sn, which are sources of spectral interference on Cd isotopes in ICP-MS measurements, are co-precipitated with hydrolysis compounds of Al and Fe (major constituents in the digested solution). However, cadmium is masked with ammonia and still remains in the supernatant. The efficiency of the removal of Zr and Sn in the co-precipitation separation are both around 99.5%, and the recovery of Cd is > 80%. The measured n(Cd-111)/n(Cd-112) ratios in the unspiked digested solution of sediment reference materials (PACS-1 and NIES No. 2) after the co-precipitation and separation are in good agreement with those for a standard Cd solution, and the procedural blank of the co-precipitation separation is extremely low (below the detection limit of ICP-MS). The analytical results obtained for Cd in PACS-1 and NIES No. 2 were 2.396 +/- 0.033 mug g(-1) (value +/- expanded uncertainty) and 0.819 +/- 0.011 mug g(-1), respectively, which are in good agreement with the certified values (PACS-1: 2.38 +/- 0.20 mug g(-1), and NIES No. 2: 0.82 +/- 0.06 mug g(-1)). This pretreatment method is simple and reliable for the determination of Cd in sediments by ID-ICP-MS.
引用
收藏
页码:1370 / 1374
页数:5
相关论文
共 16 条
[11]   RECENT PROGRESS AND PERSPECTIVES IN CADMIUM HEALTH-EFFECTS STUDIES [J].
NOMIYAMA, K .
SCIENCE OF THE TOTAL ENVIRONMENT, 1980, 14 (03) :199-232
[12]   SI-traceable values for cadmium and lead concentration in the candidate reference material, MURST-ISS Al Antarctic sediment, by combination of ICP-MS with isotope dilution [J].
Papadakis, I ;
Taylor, PDP ;
DeBievre, P .
ANALYTICA CHIMICA ACTA, 1997, 346 (01) :17-22
[13]   Determination of cadmium in sediment reference materials by isotope dilution inductively coupled plasma mass spectrometry with correction of tin isobaric interference using mass bias equations [J].
Park, CJ ;
Cho, KH ;
Suh, JK ;
Han, MS .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2000, 15 (05) :567-570
[14]   Isotopic compositions of the elements 1997 [J].
Rosman, KJR ;
Taylor, PDP .
PURE AND APPLIED CHEMISTRY, 1998, 70 (01) :217-235
[15]   Liquid-liquid extraction of cadmium by sodium diethyldithiocarbamate from biological matrixes for isotope dilution inductively coupled plasma mass spectrometry [J].
Vanderpool, RA ;
Buckley, WT .
ANALYTICAL CHEMISTRY, 1999, 71 (03) :652-659
[16]   Protocol for isotope dilution using inductively coupled plasma-mass spectrometry (ICP-MS) for the determination of inorganic elements [J].
Watters, RL ;
Eberhardt, KR ;
Beary, ES ;
Fassett, JD .
METROLOGIA, 1997, 34 (01) :87-96