共 50 条
MR imaging of implanted depth and subdural electrodes: is it safe?
被引:37
|作者:
Davis, LM
Spencer, DD
Spencer, SS
Bronen, RA
机构:
[1] Brown Univ, Rhode Isl Hosp, Sch Med, Dept Diagnost Imaging, Providence, RI 02903 USA
[2] Yale Univ, Sch Med, Dept Neurosurg, New Haven, CT USA
[3] Yale Univ, Sch Med, Dept Neurol, New Haven, CT 06510 USA
[4] Yale Univ, Sch Med, Dept Diagnost Imaging, Neuroradiol Sect, New Haven, CT 06510 USA
关键词:
chronic epilepsy;
intracranial depth and subdural electrodes;
magnetic resonance imaging;
D O I:
10.1016/S0920-1211(99)00007-8
中图分类号:
R74 [神经病学与精神病学];
学科分类号:
摘要:
This study evaluates the safety of imaging chronic epilepsy patients with intracranial depth and subdural electrodes by magnetic resonance (MR). To identify an epileptogenic focus, the precise location of the electrode contacts is necessary, and MR can provide this information. However, many neurosurgeons and neuroradiologists are hesitant to scan patients by MR with these implanted, metallic electrodes for fear of electrode displacement, current induction or heating secondary to the strong magnetic field. In the: present study, the subdural electrodes were made of stainless steel with either stainless steel or platinum contacts. The depth electrodes were made of either platinum or a nickel-chromium alloy (nichrome). We reviewed 98 cases in which patients with implanted depth electrodes, subdural electrodes, or both underwent MR scanning. A total of 143 depth electrodes, 688 subdural strips, and 38 subdural grids were implanted in the 98 procedures. MK scanning was performed on a 1.5-T unit and consisted of T1, T2, and/or spoiled gradient echo pulse sequences. There were no documented complications related to the MR scans. Based on this study and a review of the literature, we feel th;at MR imaging can safely localize intracranial electrodes. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:95 / 98
页数:4