MR imaging of implanted depth and subdural electrodes: is it safe?

被引:37
|
作者
Davis, LM
Spencer, DD
Spencer, SS
Bronen, RA
机构
[1] Brown Univ, Rhode Isl Hosp, Sch Med, Dept Diagnost Imaging, Providence, RI 02903 USA
[2] Yale Univ, Sch Med, Dept Neurosurg, New Haven, CT USA
[3] Yale Univ, Sch Med, Dept Neurol, New Haven, CT 06510 USA
[4] Yale Univ, Sch Med, Dept Diagnost Imaging, Neuroradiol Sect, New Haven, CT 06510 USA
关键词
chronic epilepsy; intracranial depth and subdural electrodes; magnetic resonance imaging;
D O I
10.1016/S0920-1211(99)00007-8
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
This study evaluates the safety of imaging chronic epilepsy patients with intracranial depth and subdural electrodes by magnetic resonance (MR). To identify an epileptogenic focus, the precise location of the electrode contacts is necessary, and MR can provide this information. However, many neurosurgeons and neuroradiologists are hesitant to scan patients by MR with these implanted, metallic electrodes for fear of electrode displacement, current induction or heating secondary to the strong magnetic field. In the: present study, the subdural electrodes were made of stainless steel with either stainless steel or platinum contacts. The depth electrodes were made of either platinum or a nickel-chromium alloy (nichrome). We reviewed 98 cases in which patients with implanted depth electrodes, subdural electrodes, or both underwent MR scanning. A total of 143 depth electrodes, 688 subdural strips, and 38 subdural grids were implanted in the 98 procedures. MK scanning was performed on a 1.5-T unit and consisted of T1, T2, and/or spoiled gradient echo pulse sequences. There were no documented complications related to the MR scans. Based on this study and a review of the literature, we feel th;at MR imaging can safely localize intracranial electrodes. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:95 / 98
页数:4
相关论文
共 50 条
  • [1] Risks and benefits of invasive epilepsy surgery workup with implanted subdural and depth electrodes
    Wellmer, Joerg
    von der Groeben, Ferdinand
    Klarmann, Ute
    Weber, Christian
    Elger, Christian E.
    Urbach, Horst
    Clusmann, Hans
    von Lehe, Marec
    EPILEPSIA, 2012, 53 (08) : 1322 - 1332
  • [2] DEPTH VERSUS SUBDURAL ELECTRODES
    MARK, VH
    NEUROLOGY, 1990, 40 (05) : 861 - 861
  • [3] WADA TESTING WITH IMPLANTED SUBDURAL ELECTRODES
    BAKAY, RAE
    EPSTEIN, CM
    BRICKLEY, GP
    EPILEPSIA, 1986, 27 (05) : 623 - 623
  • [4] IMAGING OF MR-COMPATIBLE INTRACEREBRAL DEPTH ELECTRODES
    DUCKWILER, GR
    LEVESQUE, M
    WILSON, CL
    BEHNKE, E
    BABB, TL
    LUFKIN, R
    AMERICAN JOURNAL OF NEURORADIOLOGY, 1990, 11 (02) : 353 - 354
  • [5] Nonhabitual seizures in patients with implanted subdural electrodes
    Fountas, KN
    King, DW
    Jenkins, PD
    Smith, JR
    STEREOTACTIC AND FUNCTIONAL NEUROSURGERY, 2004, 82 (04) : 165 - 168
  • [6] EXTRAOPERATIVE CORTICAL MAPPING VIA IMPLANTED SUBDURAL AND DEPTH ELECTRODES FOR MAXIMIZING ELOQUENT AREA GLIOMA RESECTION
    Fountas, Konstantinos
    Paschalis, Thanasis
    Kapsalaki, Eftychia
    NEURO-ONCOLOGY, 2020, 22 : 205 - 205
  • [7] Implanted Subdural Electrodes: Safety Issues and Complication Avoidance
    Fountas, Kostas N.
    NEUROSURGERY CLINICS OF NORTH AMERICA, 2011, 22 (04) : 519 - +
  • [8] SUBDURAL AND DEPTH ELECTRODES IN THE PRESURGICAL EVALUATION OF EPILEPSY
    BEHRENS, E
    ZENTNER, J
    VANROOST, D
    HUFNAGEL, A
    ELGER, CE
    SCHRAMM, J
    ACTA NEUROCHIRURGICA, 1994, 128 (1-4) : 84 - 87
  • [9] Long-Term Measurement of Impedance in Chronically Implanted Depth and Subdural Electrodes During Responsive Neurostimulation in Humans
    Sillay, Karl A.
    Rutecki, Paul
    Cicora, Kathy
    Worrell, Greg
    Drazkowski, Joseph
    Shih, Jerry J.
    Sharan, Ashwini D.
    Morrell, Martha J.
    Williams, Justin
    Wingeier, Brett
    BRAIN STIMULATION, 2013, 6 (05) : 718 - 726
  • [10] SUBDURAL AND EPIDURAL EMPYEMAS - MR IMAGING
    WEINGARTEN, K
    ZIMMERMAN, RD
    BECKER, RD
    HEIER, LA
    HAIMES, AB
    DECK, MDF
    AMERICAN JOURNAL OF ROENTGENOLOGY, 1989, 152 (03) : 615 - 621