Correlation between gamma passing rate and complexity of IMRT plan due to MLC position errors

被引:28
|
作者
Wang, Yewei [1 ]
Pang, Xueying [2 ]
Feng, Lina [1 ]
Wang, Helong [1 ]
Bai, Yanling [1 ]
机构
[1] Harbin Med Univ, Affiliated Tumor Hosp, Dept Radiat Phys, Baojian Rd 6, Harbin 150081, Heilongjiang, Peoples R China
[2] Heilongjiang Univ Chinese Med, Affiliated Hosp 1, Dept Oncol, Harbin, Heilongjiang, Peoples R China
基金
中国国家自然科学基金;
关键词
ArcCHECK; Gamma passing rate; MLC; Complexity; DELIVERY ERRORS; DYNAMIC IMRT; QA SYSTEMS; SENSITIVITY; ACCURACY; IMPACT; VMAT; VERIFICATION; RADIOTHERAPY; QUALITY;
D O I
10.1016/j.ejmp.2018.03.003
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: This study evaluates the correlation between the susceptibility of the gamma passing rate of IMRT plans to the multi-leaf collimator (MLC) position errors and a quantitative plan complexity metric. Methods: Twenty patients were selected for this study. For each patient, two IMRT plans were generated using sliding window and step-&-shoot techniques, respectively. Modulation complexity score (MCS) was calculated for all IMRT plans, and symmetric MLC leaf bank errors, ranging from 0.3mm to 1 mm, were introduced. Original and modified plans were delivered using Varian's Clinac iX. The obtained dose distribution using ArcCHECK was then compared with the TPS calculated dose distribution of the original plans. 3D gamma analysis was performed for each verification with passing criteria of 2%/2 mm. The gamma passing rate decreasing gradient were calculated to evaluate relationship between variation of gamma passing rate due to MLC errors and complexity. Results: A linear regression analysis was applied between gamma gradient and complexity, and the results showed a linear correlation (R-2=0.81 and 0.82 for open and closed MLC error types, respectively) indicating the more complex plans are more susceptible to MLC leaf bank errors. Meanwhile, correlation of re-normalized gamma passing rate and complexity for all errors scenarios also presented a strong correlation (r > 0.75). Conclusion: The statistics results revealed variation relationship of dosimetry robust of plans with various complexities to MLC errors. Our results also suggested that the observed susceptibility is independent of the delivery techniques.
引用
收藏
页码:112 / 120
页数:9
相关论文
共 50 条
  • [1] Investigating a Correlation Between MLC Positional Errors and IMRT QA Passing Rate
    Yousif, Ahmed
    Abdelgawad, Mahmoud
    Eldib, Ahmed
    El-Sayed, Somia
    Talaat, M.
    MEDICAL PHYSICS, 2022, 49 (06) : E892 - E893
  • [2] Investigating a correlation between MLC positional errors and IMRT QA passing rate
    Yousif, Ahmed E.
    Abdelgawad, Mahmoud H.
    Eldib, Ahmed A.
    El-Sayed, Somia M.
    Talaat, M. S.
    RADIATION PHYSICS AND CHEMISTRY, 2023, 209
  • [3] Correlation Between Average Segment Width and Gamma Passing Rate as a Function of MLC Position Error in Volumetric Modulated Arc Therapy
    Moon, Young Min
    Bae, Sang Il
    Han, Moo Jae
    Jeon, Wan
    Yu, Tosol
    Choi, Chul Won
    Kim, Jin Young
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2021, 20
  • [4] On the Sensitivity of Patient-Specific IMRT QA to MLC Position Errors
    Yan, G.
    Simon, T.
    Peng, L.
    Fox, C.
    Liu, C.
    Li, J.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [5] Correlation between VMAT plans complexity indices and gamma passing rate by using three QA phantoms
    Vacchieri, I.
    De Fatis, P. Tabarelli
    Liotta, M.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S962 - S963
  • [6] Impact of Plan Quality and Complexity Metrics on Gamma Passing Rate SIMT SRS Technique
    Salari, E.
    Sperling, N. N.
    Hadziahmetovic, M.
    Parsai, E. I.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2022, 114 (03): : E547 - E547
  • [7] Improving IMRT-plan quality with MLC leaf position refinement post plan optimization
    Niu, Ying
    Zhang, Guowei
    Berman, Barry L.
    Parke, William C.
    Yi, Byongyong
    Yu, Cedric X.
    MEDICAL PHYSICS, 2012, 39 (08) : 5118 - 5126
  • [8] Improving IMRT-Plan Quality with MLC Leaf Position Refinement Post Plan Optimization
    Niu, Y.
    Zhang, G.
    Berman, B.
    Yu, C.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [9] Are simple IMRT beams more robust against MLC error? Exploring the impact of MLC errors on planar quality assurance and plan quality for different complexity beams
    Wang, Jiazhou
    Jin, Xiance
    Peng, Jiayuan
    Xie, Jiang
    Chen, Junchao
    Hu, Weigang
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2016, 17 (03): : 147 - 157
  • [10] Improvement of prediction and classification performance for gamma passing rate by using plan complexity and dosiomics features
    Hirashima, Hideaki
    Ono, Tomohiro
    Nakamura, Mitsuhiro
    Miyabe, Yuki
    Mukumoto, Nobutaka
    Iramina, Hiraku
    Mizowaki, Takashi
    RADIOTHERAPY AND ONCOLOGY, 2020, 153 : 250 - 257