Dual-Energy Computed Tomography for the Characterization of Intracranial Hemorrhage and Calcification A Systematic Approach in a Phantom System

被引:22
|
作者
Nute, Jessica L. [1 ,2 ,7 ]
Jacobsen, Megan C. [1 ,2 ]
Chandler, Adam [3 ]
Cody, Dianna D. [4 ]
Schellingerhout, Dawid [5 ,6 ]
机构
[1] Univ Texas Houston, Grad Sch Biomed Sci, Houston, TX USA
[2] Univ Texas MD Anderson Canc Ctr, Houston, TX 77030 USA
[3] GE Healthcare, Waukesha, WI USA
[4] Univ Texas MD Anderson Canc Ctr, Dept Imaging Phys, Houston, TX 77030 USA
[5] Univ Texas MD Anderson Canc Ctr, Dept Diagnost Radiol, Houston, TX 77030 USA
[6] Univ Texas MD Anderson Canc Ctr, Dept Canc Syst Imaging, Houston, TX 77030 USA
[7] Cedars Sinai Med Ctr, Dept Environm Hlth & Safety, Los Angeles, CA 90048 USA
关键词
computed tomography; dual energy; calcification; hemorrhage; INTRACEREBRAL CAVERNOUS MALFORMATIONS; IODINATED CONTRAST; URINARY STONES; CT; DIFFERENTIATION; EXPERIENCE; STROKE; ARTIFACTS; NOISE; HEAD;
D O I
10.1097/RLI.0000000000000300
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: The aim of this study was to develop a diagnostic framework for distinguishing calcific from hemorrhagic cerebral lesions using dual-energy computed tomography (DECT) in an anthropomorphic phantom system. Materials and Methods: An anthropomorphic phantom was designed to mimic the CT imaging characteristics of the human head. Cylindrical lesion models containing either calcium or iron, mimicking calcification or hemorrhage, respectively, were developed to exhibit matching, and therefore indistinguishable, single-energy CT (SECT) attenuation values from 40 to 100 HU. These lesion models were fabricated at 0.5, 1, and 1.5 cm in diameter and positioned in simulated cerebrum and skull base locations within the anthropomorphic phantom. All lesion sizes were modeled in the cerebrum, while only 1.5-cm lesions were modeled in the skull base. Images were acquired using a GE 750HD CT scanner and an expansive dual-energy protocol that covered variations in dose (36.7-132.6 mGy CTDIvol, n = 12), image thickness (0.625-5 mm, n = 4), and reconstruction filter (soft, standard, detail, n = 3) for a total of 144 unique technique combinations. Images representing each technique combination were reconstructed into water and calcium material density images, as well as a monoenergetic image chosen to mimic the attenuation of a 120-kVp SECT scan. A true single-energy routine brain protocol was also included for verification of lesion SECT attenuation. Points representing the 3 dual-energy reconstructions were plotted into a 3-dimensional space (water [milligram/milliliter], calcium [milligram/milliliter], monoenergetic Hounsfield unit as x, y, and z axes, respectively), and the distribution of points analyzed using 2 approaches: support vector machines and a simple geometric bisector (GB). Each analysis yielded a plane of optimal differentiation between the calcification and hemorrhage lesion model distributions. By comparing the predicted lesion composition to the known lesion composition, we identified the optimal combination of CTDIvol, image thickness, and reconstruction filter to maximize differentiation between the lesion model types. To validate these results, a new set of hemorrhage and calcification lesion models were created, scanned in a blinded fashion, and prospectively classified using the planes of differentiation derived from support vector machine and GB methods. Results: Accuracy of differentiation improved with increasing dose (CTDIvol) and image thickness. Reconstruction filter had no effect on the accuracy of differentiation. Using an optimized protocol consisting of the maximum CTDIvol of 132.6 mGy, 5-mm-thick images, and a standard filter, hemorrhagic and calcific lesion models with equal SECT attenuation (Hounsfield unit) were differentiated with over 90% accuracy down to 70 HU for skull base lesions of 1.5 cm, and down to 100 HU, 60 HU, and 60 HU for cerebrum lesions of 0.5, 1.0, and 1.5 cm, respectively. The analytic method that yielded the best results was a simple GB plane through the 3-dimensional DECT space. In the validation study, 96% of unknown lesions were correctly classified across all lesion sizes and locations investigated. Conclusions: We define the optimal scan parameters and expected limitations for the accurate classification of hemorrhagic versus calcific cerebral lesions in an anthropomorphic phantom with DECT. Although our proposed DECT protocol represents an increase in dose compared with routine brain CT, this method is intended as a specialized evaluation of potential brain hemorrhage and is thus counterbalanced by increased diagnostic benefit. This work provides justification for the application of this technique in human clinical trials.
引用
收藏
页码:30 / 41
页数:12
相关论文
共 50 条
  • [21] Unenhanced Dual-Energy Computed Tomography Visualization of Brain Edema
    Mohammed, Mohammed F.
    Marais, Olivia
    Min, Adam
    Ferguson, David
    Jalal, Sabeena
    Khosa, Faisal
    O'Keeffe, Michael
    O'Connell, Tim
    Schmiedeskamp, Heiko
    Krauss, Bernhard
    Rohr, Axel
    Nicolaou, Savvas
    INVESTIGATIVE RADIOLOGY, 2018, 53 (02) : 63 - 69
  • [22] Quality evaluation of monoenergetic images generated by dual-energy computed tomography for radiotherapy: A phantom study
    Pawalowski, Bartosz
    Szweda, Hubert
    Dudkowiak, Alina
    Piotrowski, Tomasz
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 63 : 48 - 55
  • [23] In Vivo Analysis of Urinary Stones With Dual-Energy Computed Tomography
    Erdogan, Hasan
    Temizoz, Osman
    Koplay, Mustafa
    Ozturk, Bahadir
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2019, 43 (02) : 214 - 219
  • [24] Feasibility of Single-Source Dual-Energy Computed Tomography for Urinary Stone Characterization and Value of Iterative Reconstructions
    Morsbach, Fabian
    Wurnig, Moritz C.
    Mueller, Daniel
    Krauss, Bernhard
    Korporaal, Johannes Georg
    Alkadhi, Hatem
    INVESTIGATIVE RADIOLOGY, 2014, 49 (03) : 125 - 130
  • [25] Accuracy analysis of intrahepatic fat density measurements using dual-energy computed tomography: Validation using a test phantom
    Arai, Tomohiro
    Misawa, Masaki
    Arai, Miki
    Shinozaki, Masafumi
    Sakamoto, Kayo
    Yajima, Yoshinobu
    Nozaki, Yuichi
    Tajima, Tsuyoshi
    Sato, Masanori
    Hinoshita, Fumihiko
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2017, 25 (03) : 403 - 415
  • [26] Early diagnosis and prediction of intracranial hemorrhage using dual-energy computed tomography after mechanical thrombectomy in patients with acute ischemic stroke
    Wang, Zhenshan
    Chen, Wanqi
    Lin, Haitao
    Luo, Shiwei
    Liu, Yuan
    Lin, Yang
    Tao, Ying
    Huang, Weipeng
    CLINICAL NEUROLOGY AND NEUROSURGERY, 2021, 203
  • [27] Dual-Energy Computed Tomography: Is There a Penalty in Image Quality and Radiation Dose Compared With Single-Energy Computed Tomography?
    Thomas, Christoph
    Ketelsen, Dominik
    Tsiflikas, Ilias
    Reimann, Anja
    Brodoefel, Harald
    Claussen, Claus D.
    Heuschmid, Martin
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2010, 34 (02) : 309 - 315
  • [28] Dual-Energy Computed Tomography Advantages in the Acute Setting
    McLaughlin, Patrick D.
    Mallinson, Paul
    Lourenco, Pedro
    Nicolaou, Sawas
    RADIOLOGIC CLINICS OF NORTH AMERICA, 2015, 53 (04) : 619 - +
  • [29] The Role of Dual-Energy Computed Tomography in Musculoskeletal Imaging
    Fukuda, Takeshi
    Fukuda, Kunihiko
    PET CLINICS, 2018, 13 (04) : 567 - 578
  • [30] Dual-Energy Computed Tomography for Gout Diagnosis and Management
    Dalbeth, Nicola
    Choi, Hyon K.
    CURRENT RHEUMATOLOGY REPORTS, 2013, 15 (01)