Metal Release of Multilayer Coatings by Physical Vapour Deposition (PVD)

被引:19
作者
Bin Abdullah, Mohd Zakuan [1 ]
Ahmad, Mohamad Azmirrudin [1 ]
Abdullah, Ahmad Nizam [1 ]
Othman, Mohamad Hazri [1 ]
Hussain, Patthi [2 ]
Zainuddin, Azman [2 ]
机构
[1] Adv Mat Res Ctr AMREC SIRIM Berhad, Lot 34,Jalan HiTech 2-3,Kulim HiTech Pk, Kedah 09000, Malaysia
[2] Univ Teknol PETRONAS UTP Bandar Seri Iskandar, Dept Mech Engn, Tronoh 31750, Perak, Malaysia
来源
PROCEEDING OF 4TH INTERNATIONAL CONFERENCE ON PROCESS ENGINEERING AND ADVANCED MATERIALS (ICPEAM 2016) | 2016年 / 148卷
关键词
Multilayer coating; Physical Vapour Deposition (PVD); Metal Release; IMPLANTS; ALLOY;
D O I
10.1016/j.proeng.2016.06.612
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Orthopaedic implants are mostly fabricated by Stainless Steel, Titanium, and Ni-Cr Alloys that consist of Chromium (Cr), Cobalt (Co), Nickel (Ni) and Molybdenum (Mo). 10-15% of population are affected by metals (Ni, Co, Cr, Mo) hypersensitivity reaction. Multilayer coating by Physical Vapour Deposition (PVD) has been applied onto orthopedic implants to prevent metallic ions leaching into human body. This paper aims to study the metal ions leaching of Cr, Ni, Mo, and Co from multilayer coatings of Chromium (Cr), Chromium Nitride (CrN), Chromium Carbonitride (CrCN) and Zirconium Nitride (ZrN) by Physical Vapour Deposition (PVD). Cr, CrN, CrCN and ZrN have been successfully deposited onto stainless steel substrates by CAPVD. XRD analysis detected major peak in preferred orientation of (200) and other peaks with (111), (220), (311) for CrN cubic phase. For CrCN, XRD analysis detected only low intensity peaks of Cr7C3 and ZrN peaks with preferred orientation of (111), (200) with other peak (220), (311) and (222). A seven days metal released test by ICP-MS showed that generally all ions concentration for Ni, Co, Mo decreased from uncoated substrate to multilayer coatings except for Cr. Metal released showed higher concentration for coatings deposition with longer deposition time at 10 minutes. (C) 2016 Published by Elsevier Ltd.
引用
收藏
页码:254 / 260
页数:7
相关论文
共 13 条
[1]   Metal ion levels in patients with stainless steel spinal instrumentation [J].
Bruce, I. ;
Swanson, Cheryl E. .
SPINE, 2007, 32 (18) :1963-1968
[2]   Gradient CrCN cathodic arc PVD coatings [J].
Fuentes, G. G. ;
de Cerio, M. J. Diaz ;
Garcia, J. A. ;
Martinez, R. ;
Bueno, R. ;
Rodriguez, R. J. ;
Rico, M. ;
Montala, F. ;
Qin, Yi .
SURFACE & COATINGS TECHNOLOGY, 2008, 203 (5-7) :670-674
[3]   Metal sensitivity in patients with orthopaedic implants [J].
Hallab, N ;
Merritt, K ;
Jacobs, JJ .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2001, 83A (03) :428-436
[4]   Mechanical and tribological properties evaluation of cathodic arc deposited CrN/ZrN multilayer coatings [J].
Huang, Sung-Hsiu ;
Chen, Siao-Fan ;
Kuo, Yu-Chu ;
Wang, Chaur-Jeng ;
Lee, Jyh-Wei ;
Chan, Yu-Chen ;
Chen, Hsien-Wei ;
Duh, Jenq-Gong ;
Hsieh, Tsung-Eong .
SURFACE & COATINGS TECHNOLOGY, 2011, 206 (07) :1744-1752
[5]   Metal release in patients who have has a primary total hip arthroplasty - A prospective, controlled, longitudinal study [J].
Jacobs, JJ ;
Skipor, AK ;
Patteson, LM ;
Hallab, NJ ;
Paprosky, WG ;
Black, J ;
Galante, JO .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1998, 80A (10) :1447-1458
[6]  
Jacobs JJ, 1999, CLIN ORTHOP RELAT R, P173
[7]   Metal release from stainless steel, Co-Cr-Mo-Ni-Fe and Ni-Ti alloys in vascular implants [J].
Okazaki, Yoshimitsu ;
Goth, Emiko .
CORROSION SCIENCE, 2008, 50 (12) :3429-3438
[8]  
Reich J, 2010, ORTHOPADE, V39, P495, DOI 10.1007/s00132-009-1581-9
[9]   Nanometric chromium nitride/chromium carbide multilayers by r.f. magnetron sputtering [J].
Romero, J ;
Martínez, E ;
Esteve, J ;
Lousa, A .
SURFACE & COATINGS TECHNOLOGY, 2004, 180 :335-340
[10]   Relating hardness-curve shapes with deformation mechanisms in TiAlN thin films enduring indentation [J].
Santana, AE ;
Karimi, A ;
Derflinger, VH ;
Schütze, A .
MATERIALS SCIENCE AND ENGINEERING A-STRUCTURAL MATERIALS PROPERTIES MICROSTRUCTURE AND PROCESSING, 2005, 406 (1-2) :11-18