Validation of the Health & Life HL868BA blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol

被引:2
|
作者
Chen, Han-E [1 ]
Cui, Yan [1 ]
Sheng, Chang-Sheng [1 ]
Li, Li-Hua [1 ]
Li, Yan [1 ]
Wang, Ji-Guang [1 ]
机构
[1] Shanghai Jiao Tong Univ, Sch Med, Ruijin Hosp, Ctr Epidemiol Studies & Clin Trials, Shanghai 200030, Peoples R China
关键词
blood pressure measurement; European Society of Hypertension International Protocol; Health & Life HL868BA blood pressure monitor; validation study;
D O I
10.1097/MBP.0b013e32830d4bd0
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
Objective This study aimed to evaluate the accuracy of the automated Health & Life oscillometric upper arm blood pressure monitor HL868BA for home blood pressure monitoring according to the European Society of Hypertension International Protocol for blood pressure measuring devices in adults. Method Systolic and diastolic blood pressures were sequentially measured in 33 adult Chinese using a mercury sphygmomanometer (two observers) and the HL868BA device (one supervisor). Ninety-nine pairs of comparisons were obtained from 15 participants in phase 1 and further 18 participants in phase 2 of the validation study. Data analysis was performed using the ESHIP Analyzer. Results The HL868BA device successfully passed phase 1 of the validation study with a number of absolute differences between device and observers within 5, 10, and 15 mmHg for at least 31/45, 42/45, and 44/45 measurements (required 25, 35, and 40, respectively), respectively. The device also achieved the targets for phase 2.1, with 68/99, 91/99, and 95/99 differences within 5, 10, and 15 mmHg, respectively, for systolic blood pressure, and with 71/99, 88/99, and 94/99 within 5, 10, and 15 mmHg, respectively, for diastolic blood pressure. In phase 2.2, 22 and 28 participants had at least two of the three device-observers differences within 5 mmHg (required >= 22) for systolic and diastolic blood pressure, respectively. Conclusion The Health & Life upper arm blood pressure monitor HL868BA can be recommended for home use in adults.
引用
收藏
页码:305 / 308
页数:4
相关论文
共 50 条
  • [1] Validation of the ANDON KD-5915 blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol
    Huang, Qi-Fang
    Wang, Jie
    Sheng, Chang-Sheng
    Zhang, Na-Na
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2010, 15 (04) : 232 - 234
  • [2] Validation of the HONSUN LD-578 blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol
    Zhang, Yi
    Wang, Jie
    Huang, Qi-Fang
    Sheng, Chang-Sheng
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2009, 14 (03) : 128 - 131
  • [3] Validation of the Health & Life HL168JD wrist blood pressure monitor for home blood pressure monitoring according to the International Protocol
    Sheng, Chang-Sheng
    Huang, Qi-Fang
    Wang, Jie
    Zhang, Yi
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2009, 14 (03) : 125 - 127
  • [4] Validation of the Kingyield BP101H blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol
    Zeng, Wei-Fang
    Huang, Qi-Fang
    Sheng, Chang-Sheng
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2010, 15 (06) : 329 - 331
  • [5] Validation of the Kingyield BP210 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension-International Protocol
    Zeng, Wei-Fang
    Huang, Qi-Fang
    Sheng, Chang-Sheng
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2012, 17 (01) : 42 - 44
  • [6] Validation of the SCIAN LD-735 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Kang, Yuan-Yuan
    Chen, Qi
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2016, 21 (04) : 255 - 258
  • [7] Validation of the Rossmax S150 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Zeng, Wei-Fang
    Huang, Qi-Fang
    Sheng, Chang-Sheng
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2012, 17 (05) : 204 - 206
  • [8] Validation of the AVITA BPM17 wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Kang, Yuan-Yuan
    Chen, Qi
    Liu, Chang-Yuan
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2017, 22 (04) : 230 - 233
  • [9] Validation of the AGILIS ambulatory blood pressure monitor according to the European Society of Hypertension International protocol for validation of blood pressure measuring devices in adults
    Mallion, JJM
    Pierre, H
    Neuder, Y
    Ormezzano, O
    Baguet, JP
    BLOOD PRESSURE MONITORING, 2005, 10 (02) : 97 - 101
  • [10] Validation of the AVITA BPM15S wrist blood pressure monitor for home blood pressure monitoring according to the European Society of Hypertension International Protocol revision 2010
    Kang, Yuan-Yuan
    Zeng, Wei-Fang
    Zhang, Lu
    Li, Yan
    Wang, Ji-Guang
    BLOOD PRESSURE MONITORING, 2014, 19 (03) : 183 - 186