Direct kinetic assay of interactions between small peptides and immobilized antibodies using a surface plasmon resonance biosensor

被引:33
|
作者
Gomes, P
Andreu, D
机构
[1] Univ Barcelona, Dept Organ Chem, E-08028 Barcelona, Spain
[2] Ctr Invest Quim, P-4169007 Oporto, Portugal
关键词
antigen-antibody interactions; real-time biospecific interaction analysis kinetics; surface plasmon resonance analysis of small; analytes;
D O I
10.1016/S0022-1759(01)00503-8
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A surface plasmon resonance (SPR) protocol is described for the direct kinetic analysis of small antigenic peptides interacting with immobilized monoclonal antibodies (mAb). High peptide concentrations (up to 2.5 muM) and medium mAb surface densities (about 1.5 ng/mm(2)) are needed to ensure measurable binding levels, and fast buffer flow rates (60 mul/min) are required to minimize diffusion-controlled kinetics. Good reproducibility levels in the kinetic constants are obtained under these analysis conditions (standard deviations below 10% of the mean values). Application of this protocol to determine the antigenic ranking of viral peptides shows an excellent agreement between SPR and previous competition enzyme-link-ed immunosorbent assays (ELISA) on the same peptide/antibody systems. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:217 / 230
页数:14
相关论文
共 50 条
  • [1] Direct single-step surface plasmon resonance analysis of interactions between small peptides and immobilized monoclonal antibodies
    Gomes, P
    Giralt, E
    Andreu, D
    JOURNAL OF IMMUNOLOGICAL METHODS, 2000, 235 (1-2) : 101 - 111
  • [2] Kinetic Interactions between Cyclolinopeptides and Immobilized Human Serum Albumin by Surface Plasmon Resonance
    Shim, Youn Young
    Reaney, Martin J. T.
    JOURNAL OF AGRICULTURAL AND FOOD CHEMISTRY, 2015, 63 (04) : 1099 - 1106
  • [3] INTERACTIONS OF MHC CLASS-I MOLECULES WITH IMMOBILIZED ANTIGENIC PEPTIDES - DIRECT STUDIES USING SURFACE-PLASMON RESONANCE FOR DETECTION
    KHILKO, S
    CORR, M
    BOYD, LF
    LEES, A
    INMAN, JK
    MARGULIES, DH
    JOURNAL OF IMMUNOLOGY, 1993, 150 (08): : A40 - A40
  • [4] Kinetic analysis of interactions between bispecific monoclonal antibodies and immobilized antigens using a resonant mirror biosensor
    Dmitriev, DA
    Massino, YS
    Segal, OL
    JOURNAL OF IMMUNOLOGICAL METHODS, 2003, 280 (1-2) : 183 - 202
  • [5] Studies on the Interactions of Copper and Zinc Ions with β-Amyloid Peptides by a Surface Plasmon Resonance Biosensor
    Yao, Fujun
    Zhang, Ruiping
    Tian, He
    Li, Xiangjun
    INTERNATIONAL JOURNAL OF MOLECULAR SCIENCES, 2012, 13 (09) : 11832 - 11843
  • [6] Surface plasmon resonance biosensor analysis of RNA-small molecule interactions
    Davis, TM
    Wilson, WD
    DRUG-NUCLEIC ACID INTERACTIONS, 2001, 340 : 22 - 51
  • [7] Kinetic Analysis and Epitope Mapping of Monoclonal Antibodies to Salmonella Typhimurium Flagellin Using a Surface Plasmon Resonance Biosensor
    Bhandari, Devendra
    Chen, Fur-Chi
    Hamal, Shreya
    Bridgman, Roger C.
    ANTIBODIES, 2019, 8 (01)
  • [8] Surface plasmon resonance biosensor for direct detection of antibodies against human growth hormone
    Kausaite-Minkstimiene, Asta
    Ramanaviciene, Almira
    Ramanavicius, Arunas
    ANALYST, 2009, 134 (10) : 2051 - 2057
  • [9] Quantitative assay of hepatitis B surface antigen by using surface plasmon resonance biosensor
    Hwang, SY
    Yoo, CH
    Jeon, JY
    Choi, SC
    Lee, EK
    BIOTECHNOLOGY AND BIOPROCESS ENGINEERING, 2005, 10 (04) : 309 - 314
  • [10] Quantitative assay of hepatitis B surface antigen by using surface plasmon resonance biosensor
    Sang Yoon Hwang
    Chang Hoon Yoon
    Jun Yeoung Jeon
    Sung Chul Choi
    Eun Kyu Lee
    Biotechnology and Bioprocess Engineering, 2005, 10 : 309 - 314