Standardization of Techniques for Using Planar (2D) Imaging for Aerosol Deposition Assessment of Orally Inhaled Products

被引:81
作者
Newman, Stephen
Bennett, William D. [1 ]
Biddiscombe, Martyn [2 ]
Devadason, Sunalene G. [3 ]
Dolovich, Myrna B. [4 ]
Fleming, John [5 ]
Haeussermann, Sabine [6 ]
Kietzig, Claudius [6 ]
Kuehl, Philip J. [7 ]
Laube, Beth L. [8 ]
Sommerer, Knut [6 ]
Taylor, Glyn [9 ]
Usmani, Omar S. [10 ]
Zeman, Kirby L. [11 ]
机构
[1] Univ N Carolina, Dept Med, Dept Pulm Dis, Chapel Hill, NC USA
[2] Royal Brompton Hosp, Dept Nucl Med, London SW3 6LY, England
[3] Univ Western Australia, Sch Paediat & Child Hlth, Perth, WA 6009, Australia
[4] McMaster Univ, Fac Sci, Michael de Groote Sch Med, Hamilton, ON, Canada
[5] Univ Hosp Southampton NHS Fdn Trust, Southampton, Hants, England
[6] Inamed GmbH, Gauting, Germany
[7] Lovelace Resp Res Inst, Albuquerque, NM USA
[8] Johns Hopkins Univ, Sch Med, Baltimore, MD USA
[9] Cardiff Scintig & Welsh Sch Pharm, Cardiff, S Glam, Wales
[10] Univ London Imperial Coll Sci Technol & Med, Natl Heart & Lung Inst, Airway Dis Sect, London, England
[11] Univ N Carolina, Ctr Environm Med Asthma & Lung Biol, Chapel Hill, NC USA
基金
美国国家卫生研究院;
关键词
lung deposition; planar imaging; gamma scintigraphy; orally inhaled products; DRY POWDER INHALER; LUNG DEPOSITION; GAMMA-CAMERA; MUCOCILIARY CLEARANCE; REGIONAL DEPOSITION; COUGH CLEARANCE; PARTICLES; DELIVERY; HEALTHY; QUANTIFICATION;
D O I
10.1089/jamp.2012.1Su4
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Two-dimensional (2D or planar) imaging with Tc-99m radiolabels enables quantification of whole-lung and regional lung depositions for orally inhaled drug products. This article recommends standardized methodology for 2D imaging studies. Simultaneous anterior and posterior imaging with a dual-headed gamma camera is preferred, but imaging with a single-headed gamma camera is also acceptable. Correction of raw data for the effects of gamma ray attenuation is considered essential for accurate quantification, for instance, using transmission scanning with a flood-field source of Tc-99m or Co-57. Evidence should be provided of the accuracy of the quantification method, for instance, by determining "mass balance.'' Lung deposition may be expressed as a percentage of ex-valve or ex-device dose, but should also be given as mass of drug when possible. Assessment of regional lung deposition requires delineation of the lung borders, using X-ray computed tomography, radioactive gas scans (Xe-133 or Kr-81m), or transmission scans. When quantifying regional lung deposition, the lung should be divided into outer (O) and inner (I) zones. A penetration index should be calculated, as the O/I ratio for aerosol, normalized to that for a radioactive gas or transmission scan. A variety of methods can be used to assess lung deposition and distribution. Methodology and results should be documented in detail, so that data from different centers may be compared. The use of appropriate methodology will provide greater confidence in the results of 2D imaging studies, and should allay concerns that such studies are qualitative or semiquantitative in nature.
引用
收藏
页码:S10 / S28
页数:19
相关论文
共 73 条
[1]  
Administration of Radioactive Substances Advisory Committee, NOT GUID CLIN ADM RA
[2]  
AGNEW JE, 1984, J NUCL MED, V25, P170
[3]  
AGNEW JE, 1981, EUR J RESPIR DIS, V62, P239
[4]  
[Anonymous], RAD MED PROC RAD DOS
[5]   REGIONAL DEPOSITION OF AEROSOLIZED PENTAMIDINE - EFFECTS OF BODY POSITION AND BREATHING PATTERN [J].
BASKIN, MI ;
ABD, AG ;
ILOWITE, JS .
ANNALS OF INTERNAL MEDICINE, 1990, 113 (09) :677-683
[6]   Effect of salmeterol on mucociliary and cough clearance in chronic bronchitis [J].
Bennett, WD ;
Almond, MA ;
Zeman, KL ;
Johnson, JG ;
Donohue, JF .
PULMONARY PHARMACOLOGY & THERAPEUTICS, 2006, 19 (02) :96-100
[7]   Targeting delivery of aerosols to different lung regions [J].
Bennett, WD ;
Brown, JS ;
Zeman, KL ;
Hu, SC ;
Scheuch, G ;
Sommerer, K .
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 2002, 15 (02) :179-188
[8]   Mucociliary and Cough Clearance as a Biomarker for Therapeutic Development [J].
Bennett, William D. ;
Daviskas, Evangelia ;
Hasani, Amir ;
Mortensen, Jann ;
Fleming, John ;
Scheuch, Gerhard .
JOURNAL OF AEROSOL MEDICINE AND PULMONARY DRUG DELIVERY, 2010, 23 (05) :261-272
[9]   Generating monodisperse pharmacological aerosols using the spinning-top aerosol generator [J].
Biddiscombe, Martyn F. ;
Barnes, Peter J. ;
Usmani, Omar S. .
JOURNAL OF AEROSOL MEDICINE-DEPOSITION CLEARANCE AND EFFECTS IN THE LUNG, 2006, 19 (03) :245-253
[10]   Comparing Lung Regions of Interest in Gamma Scintigraphy for Assessing Inhaled Therapeutic Aerosol Deposition [J].
Biddiscombe, Martyn F. ;
Meah, Sally N. ;
Underwood, S. Richard ;
Usmani, Omar S. .
JOURNAL OF AEROSOL MEDICINE AND PULMONARY DRUG DELIVERY, 2011, 24 (03) :165-173