A GEM-based dose imaging detector with optical readout for proton radiotherapy

被引:10
作者
Klyachko, A. V. [1 ]
Moskvin, V. [2 ]
Nichiporov, D. F. [1 ]
Solberg, K. A. [1 ]
机构
[1] Indiana Univ Cyclotron Operat, Bloomington, IN 47408 USA
[2] Indiana Univ, Sch Med, Dept Radiat Oncol, Indianapolis, IN 46202 USA
基金
美国国家卫生研究院;
关键词
Proton radiotherapy; Dose imaging; GEM; Optical readout; Monte Carlo simulation; FLUKA; SCINTILLATING GEM; DOSIMETRY; THERAPY; BEAM; CHAMBER; HADRON; FLUKA; GEANT4; AR-CF4; SYSTEM;
D O I
10.1016/j.nima.2012.08.049
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
New techniques in proton radiation therapy and advances in beam delivery systems design such as beam scanning require accurate 2D dosimetry systems to verify the delivered dose distribution. Dose imaging detectors based on gas electron multipliers (GEMS) are capable of providing high sensitivity, improved dose measurement linearity, position resolution, fast response and accurate characterization of depth-dose distributions. In this work, we report on the development of a GEM-based doe imaging detector with optical readout using a CCD camera. A 10 x 10 cm(2) detector has been tested in a 205 MeV proton beam in single- and double-GEM configurations. The detector demonstrates linearity in dose rate up to 100 Gy/min and position resolution (sigma) of 0.42 mm. Transverse non-uniformity of the detector response is <= 10% before correction and the stability of the detector output throughout the day is within +/- 1%, with day-to-day reproducibility of about 10%. The depth-dose response of the detector is close to that of a wide-aperture air-filled ionization chamber and is in good agreement with Monte Carlo simulations. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:271 / 279
页数:9
相关论文
共 40 条
[31]  
Sala P.R., 2005, INFNTC0511
[33]  
SBIG CCDOps Serial and Parallel Port based Cameras Software, SBIG CCDOPS SER PAR
[34]   A scintillating gas detector for 2D dose measurements in clinical carbon beams [J].
Seravalli, E. ;
de Boer, M. ;
Geurink, F. ;
Huizenga, J. ;
Kreuger, R. ;
Schippers, J. M. ;
van Eijk, C. W. E. ;
Voss, B. .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (17) :4651-4665
[35]   2D dosimetry in a proton beam with a scintillating GEM detector [J].
Seravalli, E. ;
de Boer, M. R. ;
Geurink, F. ;
Huizenga, J. ;
Kreuger, R. ;
Schippers, J. M. ;
van Eijk, C. W. E. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (12) :3755-3771
[36]   First results of a scintillating GEM detector for 2-d Dosimetry in an α beam [J].
Seravalli, Enrica ;
Hendrikse, Jeffrey ;
Huizenga, Jan ;
Kreuger, Rob ;
Schippers, J. Marco ;
Simon, Andrea ;
van Eijk, Carel W. E. .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2007, 54 (04) :1271-1278
[37]   A scintillating GEM for 2D-dosimetry in radiation therapy [J].
Timmer, JH ;
van Vuure, TL ;
Bom, V ;
van Eijk, CW ;
de Haas, J ;
Schippers, JM .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2002, 478 (1-2) :98-103
[38]   Inorganic scintillators in medical imaging [J].
van Eijk, CWE .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (08) :R85-R106
[39]   Gafchromic EBT film dosimetry in proton beams [J].
Zhao, Li ;
Das, Indra J. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (10) :N291-N301
[40]  
Ziegler J., 2011, SRIM STOPPING RANGE