Radiochromic film for verification of superficial x-ray backscatter factors

被引:23
作者
Butson, M. J. [1 ,2 ,3 ]
Cheung, T. [2 ]
Yu, P. K. N. [2 ]
机构
[1] Illawarra Canc Care Ctr, Dept Med Phys, Wollongong, NSW 2500, Australia
[2] City Univ Hong Kong, Dept Phys & Mat Sci, Kowloon, Hong Kong, Peoples R China
[3] Univ Wollongong, Ctr Med Radiat Phys, Gwynneville, Australia
关键词
D O I
10.1007/BF03178436
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Accurate calculation and knowledge of backscatter factors (BSF) in superficial x-ray radiotherapy is required to perfonn accurate absorbed dose determination. These measurements have been performed historically with small thin parallel plate ionisation chambers and Thermoluminescent Dosimeters (TLD's). This note investigates the use of a low energy dependence radiochromic thin film (GAFCHROMIC (TM) EBT) for measurement and verification of backscatter factors. A single layer film and an extrapolation method with multiple films have been investigated. 50kVp to 150kVp beams were analysed and results for BSF were measured and compared to IPEMB (Institution of Physics and Engineering in Medicine and Biology UK) derived results. Agreement within 2% (1 SD) was found using both the single layer and extrapolation techniques with IPEMB derived results at 30cm SSD and equivalent photon energies. A 150kVp beam was found to have BSF of 1.12 +/- 0.02 (2cm circle), 1.24 +/- 0.01 (5cm circle), 1.36 +/- 0.02 (10cm circle) respectively compared to 1.11, 1.23 and 1.36 for IPEMB derived results. In summary a single layer film provided an accurate measurement and verification of BSF and was found to be within 2% of derived IPEMB results in all cases. The extrapolation method in general provided a slightly closer match to IPEMB results (<1%) but with no extra discernable accuracy than the single layer film most likely due to the already small thickness (0.3mm) of one film piece. GAFCHROMIC (TM) EBT. Radiochromic film provides a very simple and easy method for measurement and verification of BSF for x-ray energies commonly used for superficial x-ray therapy.
引用
收藏
页码:269 / 273
页数:5
相关论文
共 15 条
[1]  
Aukett R J, 2005, Phys Med Biol, V50, P2739, DOI 10.1088/0031-9155/50/12/001
[2]  
Butson Martin J, 2004, Australas Phys Eng Sci Med, V27, P230
[3]   Weak energy dependence of EBT gafchromic film dose response in the 50 kVp-10 MVp X-ray range [J].
Butson, MJ ;
Cheung, T ;
Yu, PKN .
APPLIED RADIATION AND ISOTOPES, 2006, 64 (01) :60-62
[4]   Radiochromic film for medical radiation dosimetry [J].
Butson, MJ ;
Yu, PKN ;
Cheung, T ;
Metcalfe, P .
MATERIALS SCIENCE & ENGINEERING R-REPORTS, 2003, 41 (3-5) :61-120
[5]   Skin entrance radiation dose in an interventional radiology procedure [J].
Chu, Robert Y. L. ;
Thomas, George ;
Maqbool, Feroz .
HEALTH PHYSICS, 2006, 91 (01) :41-46
[6]   A SOLID WATER PHANTOM MATERIAL FOR RADIOTHERAPY X-RAY AND GAMMA-RAY BEAM CALIBRATIONS [J].
CONSTANTINOU, C ;
ATTIX, FH ;
PALIWAL, BR .
MEDICAL PHYSICS, 1982, 9 (03) :436-441
[7]   Lithium borate TLD for determining the backscatter factors for low-energy x rays: Comparison with chamber-based and Monte Carlo derived values [J].
Coudin, D ;
Marinello, G .
MEDICAL PHYSICS, 1998, 25 (03) :347-353
[8]   Accurate skin dose measurements using radiochromic film in clinical applications [J].
Devic, S ;
Seuntjens, J ;
Abdel-Rahman, W ;
Evans, M ;
Olivares, M ;
Podgorsak, EB ;
Vuong, T ;
Soares, CG .
MEDICAL PHYSICS, 2006, 33 (04) :1116-1124
[9]   DEPENDENCE OF THE PHOTON BACKSCATTER FACTOR FOR WATER ON SOURCE-TO-PHANTOM DISTANCE AND IRRADIATION FIELD SIZE [J].
GROSSWENDT, B .
PHYSICS IN MEDICINE AND BIOLOGY, 1990, 35 (09) :1233-1245
[10]   A dosimetric evaluation of water equivalent phantoms for kilovoltage x-ray beams [J].
Hill, R ;
Holloway, L ;
Baldock, C .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (21) :N331-N344