Comparison of different registration methods and landmarks for image-guided radiation therapy of pulmonary tumors

被引:5
作者
Cao, Xiaohui [1 ]
Liu, Ming [1 ]
Zhai, Fushan [1 ]
Li, Nan [1 ]
Bao, Chaoen [1 ]
Liu, Yinliang [1 ]
Chen, Gang [2 ]
机构
[1] Hebei Med Univ, Dept Radiotherapy & Oncol, Hosp 3, Shijiazhuang 050051, Hebei, Peoples R China
[2] Hebei Med Univ, Dept Resp Med, Hosp 3, 139 Ziqiang Rd, Shijiazhuang 050051, Hebei, Peoples R China
关键词
Cone beam computed tomography; Image-guided; Image registration; Lung cancer; BEAM COMPUTED-TOMOGRAPHY; LUNG-CANCER; RADIOTHERAPY; ACCURACY; CT; EPIDEMIOLOGY; MOTION; CHINA;
D O I
10.1186/s12880-019-0343-3
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: To compare the accuracy, advantages and disadvantages of automatic registration methods at different anatomical-sites for thoracic image-guided radiation therapy (IGRT). Methods: The Varian-IX IGRT system was used to perform a manual registration of the images collected on the first fraction of 60 patients with lung cancer (42 cases central location and 18 cases of peripheral). The registered images were used as reference images. Offline registration was performed for computed tomography-CBCT images using four methods: whole image registration, ipsilateral registration, soft tissue tumor registration, and vertebral body registration. Time taken to complete and deviation value were analyzed between the different methods. Results: There were significant differences in absolute deviation value of all the three directions (P < 0.001) and the time consumption (P < 0.001) between 4 methods. The Z direction had significant differences in deviation value of 4 methods (0.023 +/- 0.128 mm, - 0.030 +/- 0.175 mm, - 0.010 +/- 0.238 mm, - 0.075 +/- 0.137 mm, P = 0.011). The difference was significant in the X direction of the ipsilateral registration method between central and peripheral lung cancer (0.033 +/- 0.053 mm vs. 0.067 +/- 0.067 mm, P = 0.045). Conclusions: The whole lung or affected side registration methods could be recommended to be used in the automatic registration function of the Varian-IX's On-Board Imaging (OBI) system.
引用
收藏
页数:6
相关论文
共 24 条
[21]   Applications of linac-mounted kilovoltage Cone-beam Computed Tomography in modern radiation therapy: A review [J].
Srinivasan, Kavitha ;
Mohammadi, Mohammad ;
Shepherd, Justin .
POLISH JOURNAL OF RADIOLOGY, 2014, 79 :181-193
[22]  
Wang Yong-chuan, 2012, Cancer Biology Medicine, V9, P128, DOI 10.3969/j.issn.2095-3941.2012.02.009
[23]   Lung cancer molecular epidemiology in China: recent trends [J].
Zhou, Caicun .
TRANSLATIONAL LUNG CANCER RESEARCH, 2014, 3 (05) :270-279
[24]   Stereoscopic X-ray imaging, cone beam CT, and couch positioning in stereotactic radiotherapy of intracranial tumors: preliminary results from a cross-modality pilot installation [J].
Zollner, Barbara ;
Heinz, Christian ;
Pitzler, Sabrina ;
Manapov, Farkhad ;
Kantz, Steffi ;
Rottler, Maya Christine ;
Niyazi, Maximilian ;
Ganswindt, Ute ;
Belka, Claus ;
Ballhausen, Hendrik .
RADIATION ONCOLOGY, 2016, 11