Parallel-opposed dual-electrode detector with recycling amperometric enhancement for capillary electrophoresis

被引:29
作者
Chen, DC [1 ]
Chang, SS [1 ]
Chen, CH [1 ]
机构
[1] Natl Sun Yat Sen Univ, Dept Chem, Kaohsiung 80424, Taiwan
关键词
D O I
10.1021/ac990069t
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The assembly and characterization of dual-electrode amperometric detection for capillary electrophoresis are described. The detector consists of a disk electrode and an integrated on-capillary electrode fabricated by depositing a gold film onto the end of the separation capillary. The two electrodes are brought together, aligned, and fixed in position using a pair of acrylic plates with a straight groove on one of the plates, the same design as that of a conventional end-column detector. A portion of the on-capillary electrode is parallel-opposed to the disk electrode in a thin-layer geometry. In this region, the redox cycling established between these two electrodes significantly enhances the amperometric signals of electrochemically reversible analytes. For measurements of dopamine in pH 6.9 phosphate electrolyte with a 12.5-mu m-i.d. capillary, such a configuration is 10-fold more sensitive than conventional end-column detection. The linear range exceeds 4 orders of magnitude (1.2 mM-50 nM) and the detection limit is 12 nM (4.2 amol, S/N = 3). Various modes of potential settings for the dual-electrode detection are also discussed.
引用
收藏
页码:3200 / 3205
页数:6
相关论文
共 31 条
[1]  
ALLEN JP, 1982, J ENDOCRINOL INVEST, V5, P1
[2]  
Beale SC, 1998, ANAL CHEM, V70, p279R
[3]   Integration of separation capillary with a Au film electrode and an etched decoupler in amperometric capillary electrophoresis [J].
Chen, DC ;
Hsieh, RMR ;
Chen, CH .
JOURNAL OF THE CHINESE CHEMICAL SOCIETY, 1998, 45 (02) :257-262
[4]   CAPILLARY ELECTROPHORESIS WITH AMPEROMETRIC DETECTION USING A POROUS CELLULOSE-ACETATE JOINT [J].
CHEN, IC ;
WHANG, CW .
JOURNAL OF CHROMATOGRAPHY, 1993, 644 (01) :208-212
[5]   AN ELECTROCHEMICAL COIL FOR END-COLUMN AMPEROMETRIC DETECTION IN CAPILLARY ELECTROPHORESIS [J].
CHEN, MC ;
HUANG, HJ .
ANALYTICAL CHEMISTRY, 1995, 67 (21) :4010-4014
[6]   LIQUID-CHROMATOGRAPHY DETECTOR BASED ON SINGLE AND TWIN ELECTRODE THIN-LAYER ELECTROCHEMISTRY - APPLICATION TO DETERMINATION OF CATECHOLAMINES IN BLOOD-PLASMA [J].
FENN, RJ ;
SIGGIA, S ;
CURRAN, DJ .
ANALYTICAL CHEMISTRY, 1978, 50 (08) :1067-1073
[7]   Rudimentary capillary-electrode alignment for capillary electrophoresis with electrochemical detection [J].
Fermier, AM ;
Gostkowski, ML ;
Colon, LA .
ANALYTICAL CHEMISTRY, 1996, 68 (09) :1661-1664
[8]   Postcolumn reaction detection with dual-electrode capillary electrophoresis-electrochemistry and electrogenerated bromine [J].
Holland, LA ;
Lunte, SM .
ANALYTICAL CHEMISTRY, 1999, 71 (02) :407-412
[9]   Amperometric detection in capillary electrophoresis with on etched joint [J].
Hu, S ;
Wang, ZL ;
Li, PB ;
Cheng, JK .
ANALYTICAL CHEMISTRY, 1997, 69 (02) :264-267
[10]   END-COLUMN DETECTION FOR CAPILLARY ZONE ELECTROPHORESIS [J].
HUANG, XH ;
ZARE, RN ;
SLOSS, S ;
EWING, AG .
ANALYTICAL CHEMISTRY, 1991, 63 (02) :189-192