Retinal Nerve Fiber Layer Thickness Reproducibility Using Seven Different OCT Instruments

被引:113
作者
Pierro, Luisa [1 ]
Gagliardi, Marco [1 ]
Iuliano, Lorenzo [1 ]
Ambrosi, Alessandro [2 ]
Bandello, Francesco [1 ]
机构
[1] Univ Vita Salute San Raffaele, Dept Ophthalmol, San Raffaele Sci Inst, I-20132 Milan, Italy
[2] Univ Vita Salute San Raffaele, Milan, Italy
关键词
OPTICAL COHERENCE TOMOGRAPHY; SPECTRAL-DOMAIN OCT; CIRRUS-HD-OCT; TIME-DOMAIN; NORMAL EYES; HEALTHY EYES; COMPARABILITY; OCT/SLO; RTVUE;
D O I
10.1167/iovs.11-8644
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. The clinical utility of new optical coherence tomography (OCT) instruments strongly depends on measurements reproducibility. The aim of this study was to assess retinal nerve fiber layer (RNFL) thickness reproducibility using six different spectral-domain OCTs (SD-OCTs) and one time-domain OCT. METHODS. RNFL thickness (average and four quadrant) from six SD-OCTs (Spectral OCT/SLO OPKO/OTI, 3D-OCT 2000 Topcon, RS-3000 NIDEK, Cirrus HD-OCT Zeiss, RTVue-100 Optovue, and Spectralis Heidelberg) and one time-domain OCT (Stratus OCT Zeiss) was measured twice in 38 right eyes of 38 randomly chosen healthy volunteers by two masked operators. Inter-and intraoperator reproducibility was evaluated by the intraclass correlation coefficient (ICC), coefficient of variation (CV), and Bland-Altman test analysis. Instrument-to-instrument reproducibility was determined by ANOVA for repeated measures. We also tested how the devices disagree in terms of systemic bias and random error using a structural equation model. RESULTS. Mean RNFL average thickness ranged from 90.08 mu m to 106.51 mu m. Cirrus and Heidelberg showed the thinnest RNFL values in all measurements, Topcon the highest. ICC, CV, and Bland-Altman plots showed variable inter- and intraoperator agreement depending on the instrument. Heidelberg demonstrated the best interoperator (ICC, 0.92; CV, 1.56%) and intraoperator (ICC, 0.94 and 0.95; CV, 1.28% and 1.26%, respectively, for operator A and operator B) agreement for average RNFL thickness. CONCLUSIONS. Heidelberg demonstrated the higher agreement in inter-and intraoperator reproducibility, Optovue the worst. In light of our error analysis results, we found that a scale bias among instruments could interfere with a thorough RNFL monitoring, suggesting that best monitoring is obtained with the same operator and the same device. (Invest Ophthalmol Vis Sci. 2012; 53: 5912-5920) DOI:10.1167/iovs.11-8644
引用
收藏
页码:5912 / 5920
页数:9
相关论文
共 50 条
  • [41] Residual and Dynamic Range of Retinal Nerve Fiber Layer Thickness in Glaucoma: Comparison of Three OCT Platforms
    Mwanza, Jean-Claude
    Kim, Hanna Y.
    Budenz, Donald L.
    Warren, Joshua L.
    Margolis, Michael
    Lawrence, Scott D.
    Jani, Pooja D.
    Thompson, Garrett S.
    Lee, Richard K.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2015, 56 (11) : 6344 - 6351
  • [42] The Optimal Diameter for Circumpapillary Retinal Nerve Fiber Layer Thickness Measurement by SD-OCT in Glaucoma
    Heindl, Ludwig M.
    Adler, Werner
    El-Malahi, Ouahiba
    Schaub, Friederike
    Hermann, Manuel M.
    Dietlein, Thomas S.
    Cursiefen, Claus
    Enders, Philip
    JOURNAL OF GLAUCOMA, 2018, 27 (12) : 1086 - 1093
  • [43] Effects of Axial Length and Age on Circumpapillary Retinal Nerve Fiber Layer and Inner Macular Parameters Measured by 3 Types of SD-OCT Instruments
    Ueda, Kaori
    Kanamori, Akiyasu
    Akashi, Azusa
    Tomioka, Mari
    Kawaka, Yuki
    Nakamura, Makoto
    JOURNAL OF GLAUCOMA, 2016, 25 (04) : 383 - 389
  • [44] Correlation and Agreement Between Cirrus HD-OCT "RNFL Thickness Map" and Scan Circle Retinal Nerve Fiber Layer Thickness Measurements
    Taibbi, Giovanni
    Kim, James D.
    Bakir, Belal H.
    Shenoy, Sudhir R.
    Pearce, William A.
    Taroyan, Gregory
    Birdsong, Orry C.
    Loucks, Emma K.
    Vizzeri, Gianmarco
    JOURNAL OF GLAUCOMA, 2016, 25 (02) : 208 - 216
  • [45] Macular Inner Plexiform and Retinal Nerve Fiber Layer Thickness in Glaucoma
    Jung, Hyun-Ho
    Sung, Mi-Sun
    Heo, Hwan
    Park, Sang-Woo
    OPTOMETRY AND VISION SCIENCE, 2014, 91 (11) : 1320 - 1327
  • [46] Retinal Nerve Fiber Layer Thickness in Vigabatrin-Exposed Patients
    Clayton, Lisa M.
    Devile, Marita
    Punte, Trusjen
    Kallis, Constantinos
    de Haan, Gerrit-Jan
    Sander, Josemir W.
    Acheson, James
    Sisodiya, Sanjay M.
    ANNALS OF NEUROLOGY, 2011, 69 (05) : 845 - 854
  • [47] Age, ocular magnification, and circumpapillary retinal nerve fiber layer thickness
    Wang, Mengyu
    Elze, Tobias
    Li, Dian
    Baniasadi, Neda
    Wirkner, Kerstin
    Kirsten, Toralf
    Thiery, Joachim
    Loeffler, Markus
    Engel, Christoph
    Rauscher, Franziska G.
    JOURNAL OF BIOMEDICAL OPTICS, 2017, 22 (12)
  • [48] Comparison of a commercial spectral-domain OCT and swept-source OCT based on an angiography scan for measuring circumpapillary retinal nerve fibre layer thickness
    Tan, Bingyao
    Chua, Jacqueline
    Harish, Thiyagrajan
    Lau, Amanda
    Gan, Alfred Tau Liang
    Tan, Yar Li
    Wong, Damon W. K.
    Chong, Rachel Shujuan
    Ang, Marcus
    Husain, Rahat
    Schmetterer, Leopold
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2020, 104 (07) : 974 - 979
  • [49] Peripapillary retinal nerve fiber layer thickness in different glaucoma stages measured by optical coherence tomography
    Zivkovic, Maja
    Jaksic, Vesna
    Jovanovic, Predrag
    Zlatanovic, Marko
    Zlatanovic, Gordana
    Djordjevic-Jocic, Jasmina
    VOJNOSANITETSKI PREGLED, 2017, 74 (02) : 121 - 126
  • [50] Glaucoma Diagnostic Capability of Circumpapillary Retinal Nerve Fiber Layer Thickness in Circle Scans With Different Diameters
    Ghassibi, Mark P.
    Chien, Jason L.
    Patthanathamrongkasem, Thipnapa
    Abumasmah, Ramiz K.
    Rosman, Michael S.
    Skaat, Alon
    Tello, Celso
    Liebmann, Jeffrey M.
    Ritch, Robert
    Park, Sung Chul
    JOURNAL OF GLAUCOMA, 2017, 26 (04) : 335 - 342