Retinal Nerve Fiber Layer Thickness Reproducibility Using Seven Different OCT Instruments

被引:113
作者
Pierro, Luisa [1 ]
Gagliardi, Marco [1 ]
Iuliano, Lorenzo [1 ]
Ambrosi, Alessandro [2 ]
Bandello, Francesco [1 ]
机构
[1] Univ Vita Salute San Raffaele, Dept Ophthalmol, San Raffaele Sci Inst, I-20132 Milan, Italy
[2] Univ Vita Salute San Raffaele, Milan, Italy
关键词
OPTICAL COHERENCE TOMOGRAPHY; SPECTRAL-DOMAIN OCT; CIRRUS-HD-OCT; TIME-DOMAIN; NORMAL EYES; HEALTHY EYES; COMPARABILITY; OCT/SLO; RTVUE;
D O I
10.1167/iovs.11-8644
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. The clinical utility of new optical coherence tomography (OCT) instruments strongly depends on measurements reproducibility. The aim of this study was to assess retinal nerve fiber layer (RNFL) thickness reproducibility using six different spectral-domain OCTs (SD-OCTs) and one time-domain OCT. METHODS. RNFL thickness (average and four quadrant) from six SD-OCTs (Spectral OCT/SLO OPKO/OTI, 3D-OCT 2000 Topcon, RS-3000 NIDEK, Cirrus HD-OCT Zeiss, RTVue-100 Optovue, and Spectralis Heidelberg) and one time-domain OCT (Stratus OCT Zeiss) was measured twice in 38 right eyes of 38 randomly chosen healthy volunteers by two masked operators. Inter-and intraoperator reproducibility was evaluated by the intraclass correlation coefficient (ICC), coefficient of variation (CV), and Bland-Altman test analysis. Instrument-to-instrument reproducibility was determined by ANOVA for repeated measures. We also tested how the devices disagree in terms of systemic bias and random error using a structural equation model. RESULTS. Mean RNFL average thickness ranged from 90.08 mu m to 106.51 mu m. Cirrus and Heidelberg showed the thinnest RNFL values in all measurements, Topcon the highest. ICC, CV, and Bland-Altman plots showed variable inter- and intraoperator agreement depending on the instrument. Heidelberg demonstrated the best interoperator (ICC, 0.92; CV, 1.56%) and intraoperator (ICC, 0.94 and 0.95; CV, 1.28% and 1.26%, respectively, for operator A and operator B) agreement for average RNFL thickness. CONCLUSIONS. Heidelberg demonstrated the higher agreement in inter-and intraoperator reproducibility, Optovue the worst. In light of our error analysis results, we found that a scale bias among instruments could interfere with a thorough RNFL monitoring, suggesting that best monitoring is obtained with the same operator and the same device. (Invest Ophthalmol Vis Sci. 2012; 53: 5912-5920) DOI:10.1167/iovs.11-8644
引用
收藏
页码:5912 / 5920
页数:9
相关论文
共 50 条
  • [21] Effect of Weiss ring on peripapillary retinal nerve fiber layer thickness measurements using SD-OCT
    Kim, Min-Su
    Nam, Ki-Yup
    Hwang, Young Hoon
    Lee, Min-Woo
    Lee, Woo-Hyuk
    Lim, Hyung-Bin
    Kim, Jung-Yeul
    SCIENTIFIC REPORTS, 2022, 12 (01)
  • [22] Relationship Between Position of Peak Retinal Nerve Fiber Layer Thickness and Retinal Arteries on Sectoral Retinal Nerve Fiber Layer Thickness
    Yamashita, Takehiro
    Asaoka, Ryo
    Tanaka, Minoru
    Kii, Yuya
    Yamashita, Toshifumi
    Nakao, Kumiko
    Sakamoto, Taiji
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2013, 54 (08) : 5481 - 5488
  • [23] Retinal nerve fiber layer thickness map
    Mujat, Mircea
    Chan, Raymond C.
    Cense, Barry
    Park, Hyle
    Joo, Chulmin
    Chen, Teresa C.
    de Boer, Johannes F.
    OPHTHALMIC TECHNOLOGIES XVI, 2006, 6138
  • [24] Retinal nerve fiber layer thickness in amblyopia
    Parikh, Rajul
    Sachdeval, Virender
    Kekunnaya, Ramesh
    Rao, B. Venkateshwar
    Parikh, Shefali
    Thomas, Ravi
    INDIAN JOURNAL OF OPHTHALMOLOGY, 2022, 70 (08) : 3065 - 3072
  • [25] Reproducibility of Retinal Nerve Fiber Layer Thickness Measurements Using the Eye Tracker and the Retest Function of Spectralis SD-OCT in Glaucomatous and Healthy Control Eyes
    Langenegger, Stefan J.
    Funk, Jens
    Toeteberg-Harms, Marc
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2011, 52 (06) : 3338 - 3344
  • [26] Influence of pupil size and cataract on retinal nerve fiber layer thickness measurements by stratus OCT
    Savini, Giacomo
    Zanini, Maurizio
    Barboni, Piero
    JOURNAL OF GLAUCOMA, 2006, 15 (04) : 336 - 340
  • [27] Reproducibility of RTVue retinal nerve fiber layer thickness and optic nerve head measurements in normal and glaucoma eyes
    Li Ji-peng
    Wang Xiao-zhen
    Fu Jing
    Li Shu-ning
    Wang Ning-li
    CHINESE MEDICAL JOURNAL, 2010, 123 (14) : 1898 - 1903
  • [28] Reproducibility of macular retinal nerve fiber layer and ganglion cell layer thickness measurements in a healthy pediatric population
    Jimenez Santos, Maria
    Acebal Montero, Alejandra
    Saenz-Frances San Baldomero, Federico
    Valverde-Megias, Alcia
    Gomez de Liano, Rosario
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2021, 31 (04) : 2087 - 2094
  • [29] Retinal Nerve Fiber Layer Thickness in Various Retinal Diseases
    Lim, Hyung-Bin
    Sung, Jae-Yun
    Ahn, Seung-Il
    Jo, Young-Joon
    Kim, Jung-Yeul
    OPTOMETRY AND VISION SCIENCE, 2018, 95 (03) : 247 - 255
  • [30] Analysis of peripapilary retinal nerve fiber layer thickness of healthy Chinese from northwestern Shanghai using Cirrus HD-OCT
    Qu, Shen
    Sun, Xiao-Ting
    Xu, Wei
    Rong, Ao
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2014, 7 (04) : 654 - 658