A comparison of the wear and physical properties of silane cross-linked polyethylene and ultra-high molecular weight polyethylene

被引:23
作者
Sakoda, H [1 ]
Voice, AM [1 ]
McEwen, HMJ [1 ]
Isaac, GH [1 ]
Hardaker, C [1 ]
Wroblewski, BM [1 ]
Fisher, J [1 ]
机构
[1] Univ Leeds, Sch Mech Engn, Dept Med & Biol Engn, Leeds LS2 9JT, W Yorkshire, England
基金
英国工程与自然科学研究理事会;
关键词
wear; polyethylene; crosslinking oxidation;
D O I
10.1054/arth.2001.27234
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Cross-linked polyethylenes are being introduced widely ill acetabular cups in hip prostheses as a strategy to reduce the incidence of wear debris-induced osteolysis. It will be many years before substantial clinical data can be collected oil the wear of these new materials. Silane cross-linked polyethylene (XLPE) was introduced into clinical practice in a limited series of acetabular cups in 1986 articulating against 22.225-mm alumina ceramic femoral heads and showed reduced wear rates compared with conventionally sterilized (gamma irradiation in air) ultra-high molecular weight polyethylene (UHMWPE). We compared the wear of XLPE manufactured ill 1986 with the weir of UHMWPE manufactured in 1986 in nonirradiated and irradiated forms. in the nonirradiated forms, the wear of XLPE was 3 times less than UHWMPE when articulating against smooth counterfaces. The nonirradiated materials lid not show signs of oxidation. In the irradiated forms, only UHMWPE showed high levels of oxidation, and this caused a substantial increase in wear. Antioxidants added to XLPE during processing gave resistance to oxidative degradation. When sliding against scratched counterfaces, the wear of UHMWPE increased by a factor of 2 to 3 times. Against the same scratched counterfaces, the wear of XLPE increased dramatically by 30 to 200 times. This difference may be attributed to the reduction in toughness of XLPF. Clinically, XLPE has been articulated against damage-resistant ceramic heads, and this probably has been an important factor in contributing to reduced wear. New cross-linked polyethylenes differ considerably front XLPE. This study indicates that it is prudent to examine the wear of new polyethylenes under a range of conditions that may occur in vivo.
引用
收藏
页码:1018 / 1023
页数:6
相关论文
共 10 条
  • [1] MATERIALS FOR INTERNAL PROSTHESES - THE PRESENT POSITION AND POSSIBLE FUTURE-DEVELOPMENTS
    ATKINSON, JR
    DOWLING, JM
    CICEK, RZ
    [J]. BIOMATERIALS, 1980, 1 (02) : 89 - 96
  • [2] A hip joint simulator study using new and physiologically scratched femoral heads with ultra-high molecular weight polyethylene acetabular cups
    Barbour, PSM
    Stone, MH
    Fisher, J
    [J]. PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 2000, 214 (H6) : 569 - 576
  • [3] LANCASTER JK, 1969, P I MECH ENG, V183, P98
  • [4] Comparative wear under four different tribological conditions of acetylene enhanced cross-linked ultra high molecular weight polyethylene
    Marrs, H
    Barton, DC
    Jones, RA
    Ward, IM
    Fisher, J
    Doyle, C
    [J]. JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 1999, 10 (06) : 333 - 342
  • [5] Wear of gamma-crosslinked polyethylene acetabular cups against roughened femoral balls
    McKellop, H
    Shen, FW
    DiMaio, W
    Lancaster, JG
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1999, (369) : 73 - 82
  • [6] Unified wear model for highly crosslinked ultra-high molecular weight polyethylenes (UHMWPE)
    Muratoglu, OK
    Bragdon, CR
    O'Connor, DO
    Jasty, M
    Harris, WH
    Gul, R
    McGarry, F
    [J]. BIOMATERIALS, 1999, 20 (16) : 1463 - 1470
  • [7] Tipper JL, 1999, J MATER SCI-MATER M, V11, P117
  • [8] Wang A, 1996, Proc Inst Mech Eng H, V210, P141, DOI 10.1243/PIME_PROC_1996_210_407_02
  • [9] Charnley low-frictional torque arthroplasty of the hip - 20-to 30-year results
    Wroblewski, BM
    Fleming, PA
    Siney, PD
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1999, 81B (03): : 427 - 430
  • [10] Low-friction arthroplasty of the hip using alumina ceramic and cross-linked polyethylene - A ten-year follow-up report
    Wroblewski, BM
    Siney, PD
    Fleming, PA
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1999, 81B (01): : 54 - 55