Enhancement of objects in photoacoustic tomography using selective filtering

被引:0
|
作者
Shin, Dong-ho [1 ]
Yang, Yoonseok [2 ]
Song, Chul-Gyu [1 ,3 ]
机构
[1] Chonbuk Natl Univ, Dept Elect Engn, Jeonju Si, Jeonbuk, South Korea
[2] Chonbuk Natl Univ, Div Biomed Engn, Jeonju Si, Jeonbuk, South Korea
[3] Chonbuk Natl Univ, Ctr Adv Bioimaging Res, Jeonju Si, Jeonbuk, South Korea
关键词
Photoacoustic image; real-time; back projection algorithm; ultrasonic imaging; selective filtering; ALGORITHM; SYSTEM;
D O I
10.3233/BME-151419
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Here we developed a real-time photoacoustic tomography (PAT) imaging acquisition device based on the linear array transducer utilized on ultrasonic devices. Also, we produced a phantom including diverse contrast media and acquired PAT imaging as the light source wavelength was changing to see if the contrast media reacted. Indocyanine green showed the highest reaction around the 800-nm band, methylene blue demonstrated the same in the 750-nm band, and gold nanoparticle showed the same in the 700-nm band. However, in the case of superparamagnetic iron oxide, we observed not reaction within the wavelength bands used herein to obtain imaging. Moreover, we applied selective filtering to the acquired PAT imaging to remove noise from around and reinforce the object's area. Consequentially, we could see the object area in the imaging was effectively detected and the image noise was removed.
引用
收藏
页码:S1223 / S1230
页数:8
相关论文
共 50 条
  • [1] Photoacoustic tomography setup using LED illumination
    Leskinen, Jarkko
    Pulkkinen, Aki
    Tick, Jenni
    Tarvainen, Tanja
    OPTO-ACOUSTIC METHODS AND APPLICATIONS IN BIOPHOTONICS IV, 2019, 11077
  • [2] Comparison of Deconvolution Filters for Photoacoustic Tomography
    Van de Sompel, Dominique
    Sasportas, Laura S.
    Jokerst, Jesse V.
    Gambhir, Sanjiv S.
    PLOS ONE, 2016, 11 (03):
  • [3] A Hybrid Preprocessing Method Using Geometry Based Diffusion and Selective Enhancement Filtering for Pulmonary Nodule Detection
    Dhara, Ashis Kumar
    Mukhopadhyay, Sudipta
    MEDICAL IMAGING 2012: COMPUTER-AIDED DIAGNOSIS, 2012, 8315
  • [4] Quantitative photoacoustic tomography
    Yuan, Zhen
    Jiang, Huabei
    PHILOSOPHICAL TRANSACTIONS OF THE ROYAL SOCIETY A-MATHEMATICAL PHYSICAL AND ENGINEERING SCIENCES, 2009, 367 (1900): : 3043 - 3054
  • [5] Photoacoustic tomography and fluorescence molecular tomography: A comparative study based on indocyanine green
    Wang, Bo
    Zhao, Qing
    Barkey, Natalie M.
    Morse, David L.
    Jiang, Huabei
    MEDICAL PHYSICS, 2012, 39 (05) : 2512 - 2517
  • [6] Real-time photoacoustic tomograpghy using linear array probe and detection of line structure using Hough transform
    Shin, Seung-Won
    Park, Jaebyung
    Shin, Dong Ho
    Song, Chul-Gyu
    Kim, Kyeong-Seop
    BIO-MEDICAL MATERIALS AND ENGINEERING, 2015, 26 : S1483 - S1490
  • [7] A functional enhancement on scarred fingerprint using sigmoid filtering
    Kolivand, Hoshang
    Hamid, Ainul Azura Binti Abdul
    Asadianfam, Shiva
    Rahim, Mohd Shafry Mohd
    NEURAL COMPUTING & APPLICATIONS, 2022, 34 (22): : 19973 - 19994
  • [8] The influences of finite aperture size in photoacoustic computed tomography
    Luo, Xiaofei
    Jiang, Jinsheng
    Wu, Hualin
    Li, Minhao
    Wang, Bo
    ULTRASONICS, 2023, 133
  • [9] Comprehensive vascular imaging using optical coherence tomography-based angiography and photoacoustic tomography
    Zabihian, Behrooz
    Chen, Zhe
    Rank, Elisabet
    Sinz, Christoph
    Bonesi, Marco
    Sattmann, Harald
    Ensher, Jason
    Minneman, Michael P.
    Hoover, Erich
    Weingast, Jessika
    Ginner, Laurin
    Leitgeb, Rainer
    Kittler, Harald
    Zhang, Edward
    Beard, Paul
    Drexler, Wolfgang
    Liu, Mengyang
    JOURNAL OF BIOMEDICAL OPTICS, 2016, 21 (09)
  • [10] Photoacoustic Tomography: Principles and Advances
    Xia, Jun
    Yao, Junjie
    Wang, Lihong V.
    PROGRESS IN ELECTROMAGNETICS RESEARCH-PIER, 2014, 147 : 1 - 22