Comparison of precision and accuracy of linear measurements performed by two different imaging software programs and obtained from 3D-CBCT images for Le Fort I osteotomy

被引:9
|
作者
Gaia, B. F. [1 ]
Pinheiro, L. R. [1 ]
Umetsubo, O. S. [1 ]
Costa, F. F. [1 ]
Cavalcanti, M. G. P. [1 ]
机构
[1] Univ Sao Paulo, Sch Dent, Dept Stomatol, Sao Paulo, Brazil
关键词
cone beam computed tomography; 3D imaging; X-ray computed tomography; osteotomy Le Fort; ORTHOGNATHIC SURGERY; COMPUTERIZED-TOMOGRAPHY; ANGULAR MEASUREMENTS; COMPLICATIONS; SURFACE;
D O I
10.1259/dmfr.20120178
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective: The purpose of this study was to compare the precision and accuracy of linear measurements for Le Fort I osteotomy performed by two different imaging software programs and obtained from three-dimensional cone beam CT (3D-CBCT) images. Methods: The study population consisted of 11 dried skulls submitted to CBCT, which generated 3D images. Linear measurements were based on craniometric anatomical landmarks pre-defined by the authors as specifically used for Le Fort I osteotomy and were identified by two radiologists twice each, independently, using Vitrea 3.8.1 (Vital Images Inc., Plymouth, MN) and open-source digital imaging communication in medicine viewer OsiriX 1.2 64-bit (Pixmeo, Geneva, Switzerland). Subsequently, a third examiner made physical measurements using a digital caliper (167 series; Mitutoyo Sul Americana Ltd, Suzano, SP, Brazil). Results: The results demonstrated a statistically significant difference between OsiriX and the gold standard, especially in the pterygoid process (TPtg L = 0.019, LLpPtg R = 0.016 and LLpPtg L = 0.012). Vitrea showed no statistical difference in comparison with the gold standard, and showed a high level of accuracy in all the measurements performed. The major difference found was 0.42 mm (LLpPtg R). Interexaminer analysis ranged from 0.90 to 0.97 using Vitrea and from 0.8 to 0.97 using OsiriX. Intraexaminer correlation coefficient ranged from 0.90 to 0.98 and from 0.84 to 0.98 for Examiners 1 and 2, respectively, using Vitrea and from 0.93 to 0.99 for Examiner 1 and from 0.64 to 0.96 for Examiner 2 using OsiriX. Conclusion: Vitrea may be considered as precise and accurate, insofar as it was able to perform all the 3D linear measurements. On the other hand, linear measurements performed using OsiriX were not successful in producing accurate linear measurements for Le Fort I osteotomy.
引用
收藏
页数:8
相关论文
共 5 条
  • [1] Accuracy and reliability of linear measurements using 3-dimensional computed tomographic imaging software for Le Fort I Osteotomy
    Gaia, Bruno Felipe
    Pinheiro, Lucas Rodrigues
    Umetsubo, Otavio Shoite
    Santos, Oseas, Jr.
    Costa, Felipe Ferreira
    Paraiso Cavalcanti, Marcelo Gusmao
    BRITISH JOURNAL OF ORAL & MAXILLOFACIAL SURGERY, 2014, 52 (03): : 258 - 263
  • [2] Accuracy and reliability of linear measurements using 3-dimensional computed tomographic imaging software for Le Fort I Osteotomy: common mistakes
    Sabour, Siamak
    BRITISH JOURNAL OF ORAL & MAXILLOFACIAL SURGERY, 2014, 52 (09): : 872 - 872
  • [3] Accuracy of three-dimensional soft tissue prediction for Le Fort I osteotomy using Dolphin 3D software: a pilot study
    Resnick, C. M.
    Dang, R. R.
    Glick, S. J.
    Padwa, B. L.
    INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2017, 46 (03) : 289 - 295
  • [4] Analysis of linear measurements on 3D surface models using CBCT data segmentation obtained by automatic standard pre-set thresholds in two segmentation software programs: an in vitro study
    Poleti, Marcelo Lupion
    Freire Fernandes, Thais Maria
    Pagin, Otavio
    Moretti, Marcela Rodrigues
    Fischer Rubira-Bullen, Izabel Regina
    CLINICAL ORAL INVESTIGATIONS, 2016, 20 (01) : 179 - 185
  • [5] Analysis of linear measurements on 3D surface models using CBCT data segmentation obtained by automatic standard pre-set thresholds in two segmentation software programs: an in vitro study
    Marcelo Lupion Poleti
    Thais Maria Freire Fernandes
    Otávio Pagin
    Marcela Rodrigues Moretti
    Izabel Regina Fischer Rubira-Bullen
    Clinical Oral Investigations, 2016, 20 : 179 - 185