Low-cost field test kits for arsenic detection in water

被引:27
作者
Das, Joyati [1 ]
Sarkar, Priyabrata [1 ]
Panda, Jigisha [1 ]
Pal, Priyabrata [2 ]
机构
[1] Univ Calcutta, Dept Polymer Sci & Technol, Biosensor Lab, Kolkata 700009, India
[2] Petr Inst, Abu Dhabi, U Arab Emirates
来源
JOURNAL OF ENVIRONMENTAL SCIENCE AND HEALTH PART A-TOXIC/HAZARDOUS SUBSTANCES & ENVIRONMENTAL ENGINEERING | 2014年 / 49卷 / 01期
关键词
Arsenic; groundwater; field kit; mercuric bromide; silver nitrate;
D O I
10.1080/10934529.2013.824764
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Arsenic, a common contaminant of groundwater, affects human health adversely. According to the World Health Organization (WHO), the maximum recommended contamination level of arsenic in drinking water is 10g/L. The purpose of this research was to develop user-friendly kits for detection of arsenic to measure at least up to 10 g/L in drinking water, so that a preventive measure could be taken. Two different kits for detection of total arsenic in water are reported here. First, the arsenic in drinking water was converted to arsine gas by a strong reducing agent. The arsine produced was then detected by paper strips via generation of color due to reaction with either mercuric bromide (KIT-1) or silver nitrate (KIT-2). These were previously immobilized on the detector strip. The first one gave a yellow color and the second one grey. Both of these kits could detect arsenic contamination within a range of 10g/L-250g/L. The detection time for both the kits was only 7min. The kits exhibited excellent performance compared to other kits available in the market with respect to detection time, ease of operation, cost and could be easily handled by a layman. The field trials with these kits gave very satisfactory results. A study on interference revealed that these kits could be used in the presence of 24 common ions present in the arsenic contaminated water. Though the kits were meant for qualitative assay, the results with unknown concentrations of real samples, when compared with atomic absorption spectrophotometer (AAS) were in good agreement as revealed by the t-test.
引用
收藏
页码:108 / 115
页数:8
相关论文
共 24 条
[1]   Sequential hydride generation/pneumatic nebulisation inductively coupled plasma mass spectrometry for the fractionation of arsenic and selenium species [J].
Anderson, SL ;
Pergantis, SA .
TALANTA, 2003, 60 (04) :821-830
[2]   A rapid field detection method for arsenic in drinking water [J].
Baghel, Anuradha ;
Singh, Beer ;
Pandey, Pratibha ;
Sekhar, K. .
ANALYTICAL SCIENCES, 2007, 23 (02) :135-137
[3]  
BRAMAN RS, 1977, ENVIRON HEALTH PERSP, V19, P1
[4]  
Cherukurii Jyotsna, 2005, Int J Environ Res Public Health, V2, P322, DOI 10.3390/ijerph2005020018
[5]   THE DETERMINATION OF ARSENIC AND ANTIMONY IN BIOLOGIC MATERIAL [J].
DALCORTIVO, LA ;
CEFOLA, M ;
UMBERGER, CJ .
ANALYTICAL BIOCHEMISTRY, 1960, 1 (06) :491-497
[6]   A rapid colorimetric method for measuring arsenic concentrations in groundwater [J].
Dhar, RK ;
Zheng, Y ;
Rubenstone, J ;
van Geen, A .
ANALYTICA CHIMICA ACTA, 2004, 526 (02) :203-209
[7]   Assessment of global industrial-age anthropogenic arsenic contamination [J].
Han, FXX ;
Su, Y ;
Monts, DL ;
Plodinec, MJ ;
Banin, A ;
Triplett, GE .
NATURWISSENSCHAFTEN, 2003, 90 (09) :395-401
[8]   Leaching of chromated copper arsenate wood preservatives: a review [J].
Hingston, JA ;
Collins, CD ;
Murphy, RJ ;
Lester, JN .
ENVIRONMENTAL POLLUTION, 2001, 111 (01) :53-66
[9]   Improving the accuracy and precision of an arsenic field test kit: increased reaction time and digital image analysis [J].
Kearns, James ;
Tyson, Julian .
ANALYTICAL METHODS, 2012, 4 (06) :1693-1698
[10]  
Khim OK., 2010, J DEF SECUR, V2, P251