Refill Port Identification of Intrathecal Drug Delivery System Devices With a Raised Fill Port

被引:4
作者
Maino, Paolo [1 ]
van Kuijk, Sander M. J. [2 ]
Koetsier, Eva [1 ]
机构
[1] Reg Hosp Lugano, Neuroctr Southern Switzerland, Pain Management Ctr, Lugano, Switzerland
[2] Maastricht Univ, Med Ctr, Clin Epidemiol & Med Technol Assessment, Maastricht, Netherlands
关键词
intrathecal drug delivery; intrathecal pumps; chronic pain; spasticity; CONFERENCE PACC RECOMMENDATIONS; CLONIDINE OVERDOSE; MANAGEMENT; PREVENTION; THERAPY;
D O I
10.1111/papr.12759
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Objectives The aim of this study was to assess the accuracy of the manual identification of the reservoir fill port (RFP) for refill of intrathecal drug delivery systems (IDDSs) with a raised RFP on the pump surface (raised-RFP-IDDSs), and compare this to previously reported data of patients with IDDSs with a recessed RFP (recessed-RFP-IDDSs). Methods Nineteen patients underwent 2 IDDS refills for the treatment of noncancer pain or spasticity. The primary endpoint of this prospective observational study was the deviation between the needle insertion point and the RFP center, quantified by fluoroscopic visualization. A distance surpassing that between the center and the margin of the RFP of 4 mm was considered a clinically relevant deviation. The results were compared with previously reported data of a patient cohort with recessed-RFP-IDDSs, and the differences were tested using Student's t-test. Results The mean deviation from the RFP center was 4.9 mm (standard deviation = 3.7). The RFP identification accuracy deviated more than the clinically relevant difference in 17 out of 35 instances (48.6%). The number of attempts and median procedural time was significantly correlated to the needle deviation. The mean deviations in the raised-RFP-IDDS cohort were consistently lower compared to the recessed-RFP-IDDS cohort (first refill procedure 4.0 vs. 8.5, P < 0.001; second procedure 5.9 vs. 8.1, P = 0.074). Conclusion The results of this study suggest that the manual localization of the RFP for raised-RFP-IDDSs is moderately accurate, and more accurate if compared to previously published accuracy of the template-guided technique for recessed-RFP-IDDSs.
引用
收藏
页码:418 / 425
页数:8
相关论文
共 48 条
  • [41] Lumbar Subarachnoid Hematoma Following an Epidural Blood Patch for Meningeal Puncture Headache Related to the Implantation of an Intrathecal Drug Delivery System
    Hustak, Erik C.
    Engle, Mitchell P.
    Viswanathan, Ashwin
    Koyyalagunta, Dhanalakshmi
    PAIN PHYSICIAN, 2014, 17 (03) : E405 - E411
  • [42] Pre- and Post-Interventional Changes in Physiological Profiles in a Patient Presenting With Opioid Withdrawal After Intrathecal Drug Delivery System Failure Related to Assumed Catheter Microfracture
    Ehsanian, Reza
    Koshkin, Eugene
    Goins, Aleyah E.
    Montera, Marena A.
    Alles, Sascha
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2021, 13 (05)
  • [43] Preventing Post Dural Puncture Headache after Intrathecal Drug Delivery System Implantation Through Preventive Fibrin Glue Application: A Retrospective Study
    Dupoiron, Denis
    Narang, Sanjeet
    Seegers, Valerie
    Lebrec, Nathalie
    Bore, Francois
    Jaoul, Virginie
    Pechard, Marie
    Hamon, Sabrina Jubier
    Delorme, Thierry
    Douillard, Thomas
    PAIN PHYSICIAN, 2021, 24 (02) : E211 - E220
  • [44] Ranibizumab Port Delivery System in Neovascular Age-Related Macular Degeneration: Where Do We Stand? Overview of Pharmacokinetics, Clinical Results, and Future Directions
    Carla, Matteo Mario
    Savastano, Maria Cristina
    Boselli, Francesco
    Giannuzzi, Federico
    Rizzo, Stanislao
    PHARMACEUTICS, 2024, 16 (03)
  • [45] Underutilized treatments for patients with refractory cancer pain: a qualitative study assessing the use of intrathecal drug delivery devices in the United Kingdom compared to alternative treatments in cancer pain management
    Barnosky, Victoria
    Brown, Matthew
    Bagchi, Somnath
    Blain, Remy
    Husejnovic, Demir
    Johnson, Sandra
    Mackworth-Praed, Meredith
    FRONTIERS IN PAIN RESEARCH, 2025, 6
  • [46] Risk Factors Associated with Postoperative Cerebrospinal Fluid Leaks After Intrathecal Drug Delivery System and an External Pump Implantation in Cancer Patients: A Retrospective Study
    Li, Yuan
    Zhao, Shuwu
    Candiotti, Keith
    Su, Chen
    PAIN AND THERAPY, 2024, : 637 - 650
  • [47] Patient Preference and Treatment Satisfaction With a Port Delivery System for Ranibizumab vs Intravitreal Injections in Patients With Neovascular Age-Related Macular Degeneration A Randomized Clinical Trial
    Chang, Margaret A.
    Kapre, Audrey
    Kaufman, Derrick
    Kardatzke, David R.
    Rabena, Melvin
    Patel, Shienal
    Bobbala, Ashwini
    Gune, Shamika
    Fung, Anne
    Wallenstein, Gene
    JAMA OPHTHALMOLOGY, 2022, 140 (08) : 771 - 778
  • [48] Rationale and Design of VOYAGER: Long-term Outcomes of Faricimab and Port Delivery System with Ranibizumab for Neovascular Age-Related Macular Degeneration and Diabetic Macular Edema in Clinical Practice
    Guymer, Robyn
    Bailey, Clare
    Chaikitmongkol, Voraporn
    Chakravarthy, Usha
    Chaudhary, Varun
    Finger, Robert P.
    Gallego-Pinazo, Roberto
    Chuan, Adrian Koh Hock
    Ishida, Susumu
    Loevestam-Adrian, Monica
    Parravano, Mariacristina
    Pinto, Jose D. Luna
    Schmitz-Valckenberg, Steffen
    Sheth, Veeral
    Souied, Eric H.
    Chi, Gloria C.
    Gilberg, Frank
    Glittenberg, Carl
    Scheidl, Stefan
    Bengus, Monica
    OPHTHALMOLOGY SCIENCE, 2024, 4 (03):