Initial experience with Penumbra Coil 400 versus standard coils in embolization of cerebral aneurysms: a retrospective review

被引:15
作者
Mascitelli, Justin R. [1 ]
Polykarpou, Maritsa F. [1 ]
Patel, Aanand A. [1 ]
Kamath, Ashwin A. [1 ]
Moyle, Henry [1 ]
Patel, Aman B. [1 ]
机构
[1] Mt Sinai Sch Med, Dept Neurosurg, New York, NY 10029 USA
关键词
Aneurysm; Embolic; Brain; Coil; Device; HYDROGEL-COATED COILS; RUPTURED INTRACRANIAL ANEURYSMS; PACKING DENSITY; DETACHABLE COILS; ENDOVASCULAR TREATMENT; PLATINUM COILS; BARE PLATINUM; OCCLUSION; RETREATMENT; MENINGITIS;
D O I
10.1136/neurintsurg-2012-010483
中图分类号
R445 [影像诊断学];
学科分类号
100207 ;
摘要
Background and purpose The Penumbra Coil 400 System (PC 400) is the newly introduced platinum coil system designed specifically to enhance filling efficiency by increasing coil diameter. Our goal was to study the packing and treatment advantage of the PC 400's unique geometric configuration compared with conventional coils (controls). Materials and methods 16 aneurysms embolized with the PC 400 in 2011 were compared with 79 equally matched aneurysms embolized with conventional coils from 2004 to 2011. Primary outcomes assessed were acute packing density, embolization time, and the number of coils required for aneurysm occlusion. Results Aneurysm embolization with the PC 400 achieved a higher packing density (36.8% vs 28.1%; p<0.005) and with fewer coils (an average of 3.9 vs 6.1 coils per aneurysm; p<0.05). In addition, the total procedural time for the subjects treated with the PC 400 was significantly less (45.7 vs 64.1min; p<0.05). There were no procedural complications associated with the PC 400. Conclusions Compared with conventional coils, the PC 400 is more efficient in the embolization of cerebral aneurysms, achieving greater packing density with fewer coils and less time without compromising safety.
引用
收藏
页码:573 / 576
页数:4
相关论文
共 36 条
  • [1] Embolization of intracranial aneurysms with second-generation Matrix-2 detachable coils: mid-term and long-term results
    Ansari, Sameer A.
    Dueweke, Eric J.
    Kanaan, Yassine
    Chaudhary, Neeraj
    Gandhi, Dheeraj
    Thompson, B. Gregory
    Gemmete, Joseph J.
    [J]. JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2011, 3 (04) : 324 - 330
  • [2] Quantitative Effects of Coil Packing Density on Cerebral Aneurysm Fluid Dynamics: An In Vitro Steady Flow Study
    Babiker, M. Haithem
    Gonzalez, L. Fernando
    Albuquerque, Felipe
    Collins, Daniel
    Elvikis, Arius
    Frakes, David H.
    [J]. ANNALS OF BIOMEDICAL ENGINEERING, 2010, 38 (07) : 2293 - 2301
  • [3] Five-year experience in using coil embolization for ruptured intracranial aneurysms: outcomes and incidence of late rebleeding
    Byrne, JV
    Sohn, NJ
    Molyneux, AJ
    [J]. JOURNAL OF NEUROSURGERY, 1999, 90 (04) : 656 - 663
  • [4] Review of 2 Decades of Aneurysm-Recurrence Literature, Part 2: Managing Recurrence after Endovascular Coiling
    Crobeddu, E.
    Lanzino, G.
    Kallmes, D. F.
    Cloft, H. J.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2013, 34 (03) : 481 - 485
  • [5] Endovascular treatment of intracranial aneurysms: review of current practice
    Currie, Stuart
    Mankad, Kshitij
    Goddard, Anthony
    [J]. POSTGRADUATE MEDICAL JOURNAL, 2011, 87 (1023) : 41 - 50
  • [6] Dorfler A, 2011, INTERV NEURORADIO S1, V7, P141
  • [7] Endovascular coils: properties, technical complications and salvage techniques
    Eddleman, Christopher Sean
    Welch, Babu G.
    Vance, Awais Z.
    Rickert, Kim L.
    White, Jonathan A.
    Pride, G. Lee
    Purdy, Phillip D.
    [J]. JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2013, 5 (02) : 104 - 109
  • [8] Coiling of Intracranial Aneurysms A Systematic Review on Initial Occlusion and Reopening and Retreatment Rates
    Ferns, Sandra P.
    Sprengers, Marieke E. S.
    van Rooij, Willem Jan
    Rinkel, Gabriel J. E.
    van Rijn, Jeroen C.
    Bipat, Shandra
    Sluzewski, Menno
    Majoie, Charles B. L. M.
    [J]. STROKE, 2009, 40 (08) : E523 - E529
  • [9] Embolization of intracranial aneurysms with hydrogel-coated coils versus inert platinum coils - Effects on packing density, coil length and quantity, procedure performance, cost, length of hospital stay, and durability of therapy
    Gaba, RC
    Ansari, SA
    Roy, SS
    Marden, FA
    Viana, MAG
    Malisch, TW
    [J]. STROKE, 2006, 37 (06) : 1443 - 1450
  • [10] Influence of Coil Geometry on Intra-Aneurysmal Packing Density: Evaluation of a New Primary Wind Technology
    Grunwald, Iris Quasar
    Molyneux, Andrew
    Kuehn, Anna Luisa
    Watson, Dave
    Byrne, James V.
    [J]. VASCULAR AND ENDOVASCULAR SURGERY, 2010, 44 (04) : 289 - 293