Fluid Dynamics of a Novel Micro-Fistula Implant for the Surgical Treatment of Glaucoma

被引:148
作者
Sheybani, Arsham [1 ]
Reitsamer, Herbert [2 ,3 ]
Ahmed, Iqbal Ike K. [4 ,5 ,6 ]
机构
[1] Washington Univ, Sch Med, Dept Ophthalmol & Visual Sci, St Louis, MO 63110 USA
[2] Univ Clin Salzburg SALK, Dept Ophthalmol, Salzburg, Austria
[3] Paracelsus Med Univ, Salzburg, Austria
[4] Univ Toronto, Dept Ophthalmol & Vis Sci, Toronto, ON, Canada
[5] Credit Valley Eye Care, Mississauga, ON, Canada
[6] Trillium Hlth Partners, Mississauga, ON, Canada
关键词
fluidics; glaucoma surgery; MIGS; IN-VITRO FLOW; TRABECULECTOMY TVT; TREATMENT OUTCOMES; AQUEOUS-HUMOR; PRESSURE; DEVICE; TUBE; VALVE;
D O I
10.1167/iovs.15-16625
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. The purpose of this study was to describe the fluidics of a novel non-valved glaucoma implant designed to prevent hypotony and compare the fluidics of this device with two commonly used non-valved glaucoma devices. METHODS. The XEN 45 micro-fistula implant was designed to limit hypotony by virtue of its length and width according to the Hagen-Poiseuille equation. Flow testing was performed using a syringe pump and pressure transducer at multiple flow rates. The pressure differentials across the XEN implant, the Ex-Press implant, and 10 mm of silicone tubing from a Baerveldt implant at a physiologic flow rate (2.5 mu L/min) were extrapolated. RESULTS. The XEN 45 achieved a steady-state pressure calculated at 7.56 mm Hg at 2.5 mu L/min. At the same flow rate, the Ex-Press device and Baerveldt tubing reached steady-state pressures of 0.09 and 0.01 mm Hg, respectively. CONCLUSIONS. Under flow testing, the XEN micro-fistula implant was able to maintain backpressure above numerical hypotony levels without the use of complex valve systems. This is due to the XEN implant's design, derived from the principles that dictate Newtonian fluids.
引用
收藏
页码:4789 / 4795
页数:7
相关论文
共 31 条
[1]   Three-year Treatment Outcomes in the Ahmed Baerveldt Comparison Study [J].
Barton, Keith ;
Feuer, William J. ;
Budenz, Donald L. ;
Schiffman, Joyce ;
Costa, Vital P. ;
Godfrey, David G. ;
Buys, Yvonne M. .
OPHTHALMOLOGY, 2014, 121 (08) :1547-+
[2]   Do scleral flap dimensions influence reliability of intraocular pressure control in experimental trabeculectomy? [J].
Birchall, W. ;
Bedggood, A. ;
Wells, A. P. .
EYE, 2007, 21 (03) :402-407
[3]   Intraoperative testing of opening and closing pressure predicts risk of low intraocular pressure after Ahmed glaucoma valve implantation [J].
Bochmann, F. ;
Kipfer, A. ;
Tarantino, J. ;
Kaufmann, C. ;
Bachmann, L. ;
Thiel, M. .
EYE, 2014, 28 (10) :1184-1189
[4]   Outflow resistance of the Baerveldt glaucoma drainage implant and modifications for early postoperative intraocular pressure control [J].
Breckenridge, RR ;
Bartholomew, LR ;
Crosson, CE ;
Kent, AR .
JOURNAL OF GLAUCOMA, 2004, 13 (05) :396-399
[5]   Priming the Ahmed Glaucoma Valve: Pressure Required and Effect of Overpriming [J].
Cheng, Jason ;
Abolhasani, Milad ;
Beltran-Agullo, Laura ;
Moss, Edward Bram ;
Buys, Yvonne M. ;
Trope, Graham E. .
JOURNAL OF GLAUCOMA, 2015, 24 (04) :E34-E35
[6]   The Ahmed Versus Baerveldt Study [J].
Christakis, Panos G. ;
Tsai, James C. ;
Kalenak, Jeffrey W. ;
Zurakowski, David ;
Cantor, Louis B. ;
Kammer, Jeffrey A. ;
Ahmed, Iqbal I. K. .
OPHTHALMOLOGY, 2013, 120 (11) :2232-2240
[7]   Implantation of a miniature glaucoma device under a scleral flap [J].
Dahan, E ;
Carmichael, TR .
JOURNAL OF GLAUCOMA, 2005, 14 (02) :98-102
[8]   Shunts to Divert Aqueous Humor to Distant Epithelialized Cavities After Keratoprosthesis Surgery [J].
Dohlman, Claes H. ;
Grosskreutz, Cynthia L. ;
Chen, Teresa C. ;
Pasquale, Louis R. ;
Rubin, Peter A. D. ;
Kim, Eva C. ;
Durand, Marlene .
JOURNAL OF GLAUCOMA, 2010, 19 (02) :111-115
[9]  
Eisenberg DL, 1999, OPHTHALMIC SURG LAS, V30, P662
[10]   Comparative In Vitro Flow Study of 3 Different Ex-PRESS Miniature Glaucoma Device Models [J].
Estermann, Stephan ;
Yuttitham, Kanokwan ;
Chen, Julie A. ;
Lee, On-Tat ;
Stamper, Robert L. .
JOURNAL OF GLAUCOMA, 2013, 22 (03) :209-214