A low-cost solution for the restoration of femoral head centre during total hip arthroplasty

被引:1
作者
Hill, Janet C. [1 ]
Salazar-Torres, Jose J. [2 ]
Orr, John F. [3 ]
Archbold, H. A. Pooler [1 ]
Beverland, David E. [1 ]
机构
[1] Musgrave Pk Hosp, Primary Joint Unit, Belfast BT9 7JB, Antrim, North Ireland
[2] Musgrave Pk Hosp, Gait Anal Lab, Belfast BT9 7JB, Antrim, North Ireland
[3] Queens Univ Belfast, Sch Mech & Aerosp Engn, Belfast, Antrim, North Ireland
关键词
Total hip arthroplasty; calliper; femoral head restoration; LEG-LENGTH DISCREPANCY; TRANSVERSE ACETABULAR LIGAMENT; LIMB-LENGTH; REPLACEMENT; INEQUALITY; COMPONENT; STABILITY; WEAR; METAANALYSIS; ORIENTATION;
D O I
10.1177/0954411913482438
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Restoration of joint centre during total hip arthroplasty is critical. While computer-aided navigation can improve accuracy during total hip arthroplasty, its expense makes it inaccessible to the majority of surgeons. This article evaluates the use, in the laboratory, of a calliper with a simple computer application to measure changes in femoral head centres during total hip arthroplasty. The computer application was designed using Microsoft Excel and used calliper measurements taken pre- and post-femoral head resection to predict the change in head centre in terms of offset and vertical height between the femoral head and newly inserted prosthesis. Its accuracy was assessed using a coordinate measuring machine to compare changes in preoperative and post-operative head centre when simulating stem insertion on 10 saw-bone femurs. A femoral stem with a modular neck was used, which meant nine possible head centre configurations were available for each femur, giving 90 results. The results show that using this technique during a simulated total hip arthroplasty, it was possible to restore femoral head centre to within 6 mm for offset (mean 1.67 +/- 1.16 mm) and vertical height (mean 2.14 +/- 1.51 mm). It is intended that this low-cost technique be extended to inform the surgeon of a best-fit solution in terms of neck length and neck type for a specific prosthesis.
引用
收藏
页码:629 / 635
页数:7
相关论文
共 50 条
[41]   Cost Analysis of Dual-Mobility Versus Large Femoral Head Constructs in Revision Total Hip Arthroplasty [J].
Abdel, Matthew P. ;
Miller, Larry E. ;
Hanssen, Arlen D. ;
Pagnano, Mark W. .
JOURNAL OF ARTHROPLASTY, 2019, 34 (02) :260-264
[42]   Intraoperative Complications and Mid-Term Follow-Up of Large-Diameter Head Metal-on-Metal Total Hip Arthroplasty and Hip Resurfacing Arthroplasty [J].
Miettinen, S. S. A. ;
Makinen, T. J. ;
Makela, K. ;
Huhtala, H. ;
Kettunen, J. S. ;
Remes, V. .
SCANDINAVIAN JOURNAL OF SURGERY, 2018, 107 (02) :180-186
[43]   Total Hip Arthroplasty After Previous Transtrochanteric Anterior Rotational Osteotomy for Femoral Head Osteonecrosis [J].
Lee, Young-Kyun ;
Ha, Yong-Chan ;
Kim, Ki-Choul ;
Yoo, Jeong Joon ;
Koo, Kyung-Hoi .
JOURNAL OF ARTHROPLASTY, 2009, 24 (08) :1205-1209
[44]   Conversion Total Hip Arthroplasty After Previous Transtrochanteric Rotational Osteotomy for Osteonecrosis of the Femoral Head [J].
Park, Kyung-Soon ;
Tumin, Masjudin ;
Peni, Indra ;
Yoon, Taek-Rim .
JOURNAL OF ARTHROPLASTY, 2014, 29 (04) :813-816
[45]   The wear of a polyethylene socket articulating with a zirconia ceramic femoral head in canine total hip arthroplasty [J].
Hamada, Y ;
Horiuchi, T ;
Sano, Y ;
Usui, I .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1999, 48 (03) :301-308
[46]   Plain Radiographs Fail to Reflect Femoral Offset in Total Hip Arthroplasty [J].
Weber, Markus ;
Woerner, Michael L. ;
Springorum, Hans-Robert ;
Hapfelmeier, Alexander ;
Grifka, Joachim ;
Renkawitz, Tobias F. .
JOURNAL OF ARTHROPLASTY, 2014, 29 (08) :1661-1665
[47]   Modular noncemented femoral stem system in revision total hip arthroplasty [J].
Schofer, M. D. ;
Efe, T. ;
Heyse, T. J. ;
Timmesfeld, N. ;
Velte, R. ;
Hinrichs, F. ;
Schmitt, J. .
ORTHOPADE, 2010, 39 (02) :209-+
[48]   Fatal fat embolism following femoral head resection in total hip arthroplasty [J].
Walker, N. M. ;
Bateson, T. ;
Reavley, P. ;
Prakash, D. .
HIP INTERNATIONAL, 2008, 18 (04) :332-334
[49]   Ceramic-on-ceramic total hip arthroplasty in patients with osteonecrosis of the femoral head [J].
Jung, Jae Ho ;
Lee, Tong Joo ;
Kim, Su Yeon ;
Jung, Sae Rom ;
Moon, Kyoung Ho .
CURRENT ORTHOPAEDIC PRACTICE, 2011, 22 (05) :432-436
[50]   Femoral neck osteotomy guide for total hip arthroplasty [J].
Lei Yang ;
Zhanle Zheng ;
Wei Chen ;
Juan Wang ;
Yingze Zhang .
BMC Surgery, 15