A low-cost solution for the restoration of femoral head centre during total hip arthroplasty

被引:1
作者
Hill, Janet C. [1 ]
Salazar-Torres, Jose J. [2 ]
Orr, John F. [3 ]
Archbold, H. A. Pooler [1 ]
Beverland, David E. [1 ]
机构
[1] Musgrave Pk Hosp, Primary Joint Unit, Belfast BT9 7JB, Antrim, North Ireland
[2] Musgrave Pk Hosp, Gait Anal Lab, Belfast BT9 7JB, Antrim, North Ireland
[3] Queens Univ Belfast, Sch Mech & Aerosp Engn, Belfast, Antrim, North Ireland
关键词
Total hip arthroplasty; calliper; femoral head restoration; LEG-LENGTH DISCREPANCY; TRANSVERSE ACETABULAR LIGAMENT; LIMB-LENGTH; REPLACEMENT; INEQUALITY; COMPONENT; STABILITY; WEAR; METAANALYSIS; ORIENTATION;
D O I
10.1177/0954411913482438
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Restoration of joint centre during total hip arthroplasty is critical. While computer-aided navigation can improve accuracy during total hip arthroplasty, its expense makes it inaccessible to the majority of surgeons. This article evaluates the use, in the laboratory, of a calliper with a simple computer application to measure changes in femoral head centres during total hip arthroplasty. The computer application was designed using Microsoft Excel and used calliper measurements taken pre- and post-femoral head resection to predict the change in head centre in terms of offset and vertical height between the femoral head and newly inserted prosthesis. Its accuracy was assessed using a coordinate measuring machine to compare changes in preoperative and post-operative head centre when simulating stem insertion on 10 saw-bone femurs. A femoral stem with a modular neck was used, which meant nine possible head centre configurations were available for each femur, giving 90 results. The results show that using this technique during a simulated total hip arthroplasty, it was possible to restore femoral head centre to within 6 mm for offset (mean 1.67 +/- 1.16 mm) and vertical height (mean 2.14 +/- 1.51 mm). It is intended that this low-cost technique be extended to inform the surgeon of a best-fit solution in terms of neck length and neck type for a specific prosthesis.
引用
收藏
页码:629 / 635
页数:7
相关论文
共 50 条
[21]   Restoration of Proximal Femoral Anatomy during Total Hip Arthroplasty for High Developmental Dysplasia of the Hip: An Original Technique [J].
Wu, Kerong ;
Zhang, Xianzuo ;
Chen, Min ;
Shang, Xifu .
ORTHOPAEDIC SURGERY, 2020, 12 (01) :343-350
[22]   Digital Fluoroscopic Navigation for Limb Length Restoration During Anterior Total Hip Arthroplasty [J].
O'Leary, Robert ;
Saxena, Anshul ;
Arguelles, William ;
Hernandez, Yvette ;
Osondu, Chukwuemeka U. ;
Suarez, Juan C. .
ARTHROPLASTY TODAY, 2022, 18 :11-15
[23]   Impingement and stability of total hip arthroplasty versus femoral head resurfacing using a cadaveric robotics model [J].
Colbrunn, R. W. ;
Bottros, J. J. ;
Butler, R. S. ;
Klika, A. K. ;
Bonner, T. F. ;
Greeson, C. ;
van den Bogert, A. J. ;
Barsoum, W. K. .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2013, 31 (07) :1108-1115
[24]   Surgical accuracy, function, and quality of life of simultaneous versus staged bilateral Total hip Arthroplasty in patients with Osteonecrosis of the femoral head [J].
Kim, Seung-Chan ;
Lim, Young-Wook ;
Jo, Woo-Lam ;
Park, Dong-Chul ;
Lee, Jin-Woo ;
Kang, Won-Woo ;
Kim, Yong-Sik .
BMC MUSCULOSKELETAL DISORDERS, 2017, 18
[25]   Accuracy of leg length and femoral offset restoration after total hip arthroplasty with the utilisation of an intraoperative calibration gauge [J].
Enke, Oliver ;
Levy, Yadin D. ;
Bruce, Warwick J. M. .
HIP INTERNATIONAL, 2020, 30 (03) :296-302
[26]   Influence of preoperative femoral orientation on radiographic measures of femoral head height in total hip replacement [J].
O'Connor, John D. ;
Hill, Janet C. ;
Beverland, David E. ;
Dunne, Nicholas J. ;
Lennon, Alex B. .
CLINICAL BIOMECHANICS, 2021, 81 (81)
[27]   Femoral Fixation in Revision Total Hip Arthroplasty [J].
Hartman, Curtis W. ;
Garvin, Kevin L. .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2011, 93A (24) :2311-2322
[28]   Delay of total hip arthroplasty to advanced stage worsens post-operative hip motion in patients with femoral head osteonecrosis [J].
Jo, Woo-Lam ;
Lee, Young-Kyun ;
Ha, Yong-Chan ;
Kim, Tae-Young ;
Koo, Kyung-Hoi .
INTERNATIONAL ORTHOPAEDICS, 2018, 42 (07) :1599-1603
[29]   Alumina femoral head fracture in uncemented total hip arthroplasty with a ceramic sandwich cup [J].
Hwang, Deuk-Soo ;
Kim, Young-Mo ;
Lee, Chang-Hwan .
JOURNAL OF ARTHROPLASTY, 2007, 22 (03) :468-471
[30]   Outcome of Total Hip Arthroplasty for Avascular Necrosis of the Femoral Head in Systemic Lupus Erythematosus [J].
Woo, Min Su ;
Kang, Joon Soon ;
Moon, Kyoung Ho .
JOURNAL OF ARTHROPLASTY, 2014, 29 (12) :2267-2270