Potentiometric Determination of Ketotifen Fumarate in Pharmaceutical Preparations and Urine Using Carbon Paste and PVC Membrane Selective Electrodes

被引:11
作者
Frag, Eman Y. Z. [1 ]
Mohamed, Gehad G. [1 ]
Khalil, Mohamed M. [2 ]
Hwehy, Mohammad M. A. [2 ]
机构
[1] Cairo Univ, Dept Chem, Fac Sci, Giza 12613, Egypt
[2] Beni Suef Univ, Dept Chem, Fac Sci, Bani Suwayf, Egypt
关键词
HUMAN-PLASMA; SPECTROPHOTOMETRIC ESTIMATION; SPECTROMETRY; COEFFICIENTS; STABILITY;
D O I
10.1155/2011/604741
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This study compares between unmodified carbon paste (CPE; the paste has no ion pair) and polyvinyl chloride (PVC) membrane selective electrodes that were used in potentiometric determination of ketotifen fumarate (KTF), where sodium tetraphenylborate (NaTPB) was used as titrant. The performance characteristics of these sensors were evaluated according to IUPAC recommendations which reveal a fast, stable, and linear response for KTF over the concentration range of 10(-7) to 10(-2) mol L-1. The electrodes show Nernstian slope value of 52.51 +/- 0.20 and 51.51 +/- 0.25 mV decade(-1) for CPE and PVC membrane electrodes at 30 degrees C, respectively. The potential is nearly stable over the pH range 3.0-6.0 and 2.0-7.0 for CPE and PVC membrane electrodes, respectively. Selectivity coefficient values towards different inorganic cations, sugars, and amino acids reflect high selectivity of the prepared electrodes. The electrodes responses at different temperatures were also studied, and long operational lifetime of 12 and 5 weeks for CPE and PVC membrane electrodes, respectively, were found. These are used for determination of ketotifen fumarate using potentiometric titration, calibration, and standard addition methods in pure samples, its pharmaceutical preparations (Zaditen tablets), and biological fluid (urine). The direct potentiometric determination of KTF using the proposed sensors gave recoveries % of 98.97 +/- 0.53 and 98.62 +/- 0.74 with RSD 1.42 and 0.63% for CPE and PVC membrane selective electrodes, respectively. Validation of the method shows suitability of the proposed sensors for use in quality control assessment of KTF. The obtained results were in a good agreement with those obtained using the reported spectrophotometric method.
引用
收藏
页数:7
相关论文
共 28 条
[1]   Stability studies on some benzocycloheptane antihistaminic agents [J].
Abounassif, MA ;
El-Obeid, HA ;
Gadkariem, EA .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2005, 36 (05) :1011-1018
[2]   Determination of ketotifen in human plasma by LC-MS [J].
Alali, FQ ;
Tashtoush, BM ;
Najib, NM .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2004, 34 (01) :87-94
[3]   Determination of Ketotifen fumarate in raw material and pharmaceutical products using ion-pair formation [J].
Amanlou, Massoud ;
Nazlou, Mehdi Hoseinzadeh ;
Azizian, Homa ;
Souri, Effat ;
Farsam, Hassan .
ANALYTICAL LETTERS, 2007, 40 (16-18) :3267-3279
[4]  
[Anonymous], 2009, British Pharmacopoeia
[5]  
Bauer H.H., 1978, Instrumental Analysis
[6]   Determination of ketotifen and its conjugated metabolite in human plasma by liquid chromatography/tandem mass spectrometry: application to a pharmacokinetic study [J].
Chen, XY ;
Zhong, DF ;
Liu, D ;
Wang, YW ;
Han, Y ;
Gu, JK .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2003, 17 (22) :2459-2463
[7]  
Cosofret V.V., 1992, PHARM APPL MEMBRANE
[8]   Application of a Continuous Square-Wave Potential Program for Sub Nano Molar Determination of Ketotifen [J].
Daneshgar, Parandis ;
Norouzi, Parviz ;
Ganjali, Mohammad Reza .
CHEMICAL & PHARMACEUTICAL BULLETIN, 2009, 57 (02) :117-121
[9]   Determination of some antihistaminic drugs by atomic absorption spectrometry and colorimetric methods [J].
El-Kousy, N ;
Bebawy, LI .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1999, 20 (04) :671-679
[10]  