Clinical Impact of an Open-Irrigated Radiofrequency Catheter with Direct Force Measurement on Atrial Fibrillation Ablation

被引:65
作者
Martinek, Martin
Lemes, Christine
Sigmund, Elisabeth
Derndorfer, Michael
Aichinger, Josef
Winter, Siegmund
Nesser, Hans-Joachim
Puererfellner, Helmut
机构
[1] Univ Innsbruck, Elisabethinen Univ Teaching Hosp, Dept Cardiol, Linz, Austria
[2] Graz Univ, Linz, Austria
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2012年 / 35卷 / 11期
关键词
Atrial fibrillation; radiofrequency catheter ablation; contact force measurement; ELECTRODE-TISSUE CONTACT; LESION SIZE; IMPEDANCE; MODEL; AREA;
D O I
10.1111/j.1540-8159.2012.03503.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: Electrode-tissue contact is crucial for adequate lesion formation in radiofrequency catheter ablation (RFCA). Objective: We assessed the impact of direct catheter force measurement on acute procedural parameters during RFCA of atrial fibrillation (AF). Methods: Fifty consecutive patients (28 male) with paroxysmal AF who underwent their first procedure of circumferential pulmonary vein (PV) isolation (PVI) were assigned to either RFCA using (1) a standard 3.5-mm open-irrigated-tip catheter or (2) a catheter with contact force measurement capabilities. Using the endpoint of PVI with entry and exit block, acute procedural parameters were assessed. Results: Procedural data showed a remarkable decline in ablation time (radiofrequency time needed for PVI) from 50.5 +/- 15.9 to 39.0 +/- 11.0 minutes (P = 0.007) with a reduction in overall procedure duration from 185 +/- 46 to 154 +/- 39 minutes (P = 0.022). In parallel, the total energy delivered could be significantly reduced from 70,926 +/- 19,470 to 58,511 +/- 14,655 Ws (P = 0.019). The number of acute PV reconnections declined from 36% to 12% (P = 0.095). Conclusions: The use of contact force sensing technology is able to significantly reduce ablation and procedure times in PVI. In addition, energy delivery is substantially reduced by avoiding radiofrequency ablation in positions with insufficient surface contact. Procedural efficacy and safety of this new feature have to be evaluated in larger cohorts. (PACE 2012; 35:13121318)
引用
收藏
页码:1312 / 1318
页数:7
相关论文
共 15 条
[1]   The effects of electrode-tissue contact on radiofrequency lesion generation [J].
Avitall, B ;
Mughal, K ;
Hare, J ;
Helms, R ;
Krum, D .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (12) :2899-2910
[2]   The LETR-Principle: A novel method to assess electrode-tissue contact in radiofrequency ablation [J].
Eick, OJ ;
Wittkampf, FH ;
Bronneberg, T ;
Schumacher, B .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1998, 9 (11) :1180-1185
[3]  
HAINES D E, 1991, Journal of Cardiovascular Electrophysiology, V2, P509, DOI 10.1111/j.1540-8167.1991.tb01353.x
[4]   Contact force-controlled zero-fluoroscopy catheter ablation of right-sided and left atrial arrhythmia substrates [J].
Kerst, Gunter ;
Weig, Hans-Joerg ;
Weretka, Slawomir ;
Seizer, Peter ;
Hofbeck, Michael ;
Gawaz, Meinrad ;
Schreieck, Juergen .
HEART RHYTHM, 2012, 9 (05) :709-714
[5]   A novel radiofrequency ablation catheter using contact force sensing: Toccata study [J].
Kuck, Karl-Heinz ;
Reddy, Vivek Y. ;
Schmidt, Boris ;
Natale, Andrea ;
Neuzil, Petr ;
Saoudi, Nadir ;
Kautzner, Josef ;
Herrera, Claudia ;
Hindricks, Gerhard ;
Jais, Pierre ;
Nakagawa, Hiroshi ;
Lambert, Hendrik ;
Shah, Dipen C. .
HEART RHYTHM, 2012, 9 (01) :18-23
[6]   Comparison between anatomical and integrated approaches to atrial fibrillation ablation: Adjunctive role of electrical pulmonary vein disconnection [J].
Mantovan, R ;
Verlato, R ;
Calzolari, V ;
Baccillieri, S ;
De Leo, A ;
Turrini, P ;
Pastore, G ;
Crosato, M ;
Ramondo, A ;
Stritoni, P .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2005, 16 (12) :1293-1297
[7]   Identification of a high-risk population for esophageal injury during radiofrequency catheter ablation of atrial fibrillation: Procedural and anatomical considerations [J].
Martinek, Martin ;
Meyer, Christian ;
Hassanein, Said ;
Aichinger, Josef ;
Bencsik, Gabor ;
Schoefl, Rainer ;
Boehm, Gernot ;
Nesser, Hans-Joachim ;
Purerfellner, Helmut .
HEART RHYTHM, 2010, 7 (09) :1224-1230
[8]   Effects of dispersive electrode position and surface area on electrical parameters and temperature during radiofrequency catheter ablation [J].
Nath, S ;
DiMarco, JP ;
Gallop, RG ;
McRury, ID ;
Haines, DE .
AMERICAN JOURNAL OF CARDIOLOGY, 1996, 77 (09) :765-&
[9]   Area Under the Real-Time Contact Force Curve (Force-Time Integral) Predicts Radiofrequency Lesion Size in an In Vitro Contractile Model [J].
Shah, Dipen C. ;
Lambert, Hendrik ;
Nakagawa, Hiroshi ;
Langenkamp, Arne ;
Aeby, Nicolas ;
Leo, Giovanni .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2010, 21 (09) :1038-1043
[10]   Incidence of Asymptomatic Intracranial Embolic Events After Pulmonary Vein Isolation Comparison of Different Atrial Fibrillation Ablation Technologies in a Multicenter Study [J].
Siklody, Claudia Herrera ;
Deneke, Thomas ;
Hocini, Meleze ;
Lehrmann, Heiko ;
Shin, Dong-In ;
Miyazaki, Shinsuke ;
Henschke, Susanne ;
Fluegel, Peter ;
Schiebeling-Roemer, Jochen ;
Bansmann, Paul M. ;
Bourdias, Thomas ;
Dousset, Vincent ;
Haissaguerre, Michel ;
Arentz, Thomas .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2011, 58 (07) :681-688